CLAS Certificate Number 94-03

Company name
Miller Instruments Ltd.
Company address
#1 – 3871 North Fraser Way 
Burnaby BC  V5J 5G6
Contact
Laurie Dubovik, President 
Telephone: (604) 431-8882 
Email: laurie@miller.bc.ca
Clients served
  • All interested parties
  • Some calibration services are available on-site. See Remarks for details.
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 15208
  • First issued 
  • Issue 13.0e

This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body.

The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS.

See supplementary notes.

On this page

Electrical

In this section

Current, AC

For the calibration of ac current sources and measurement devices. Both the measure and generate capabilities are monitored using a digital multimeter. 
See Annex A for details.

Current, AC: Table 1
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
100 μA to 1000 μA45 Hz to 10 kHz0.26 % to 0.05 %II
1 mA to 10 mA45 Hz to 10 kHz0.26 % to 0.05 %II
10 mA to 100 mA45 Hz to 10 kHz0.26 % to 0.05 %II
100 mA to 1 A45 Hz to 5 kHz0.3 % to 0.1 %II

Both the measure and generate capabilities are monitored with a digital multimeter using resistance shunts.

Current, AC: Table 2
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
1 A to 2 A45 Hz to 100 Hz0.09 %II
2 A to 3 A45 Hz to 100 Hz0.06 %II
3 A to 5 A45 Hz to 100 Hz0.05 %II
5 A to 100 A45 Hz to 100 Hz0.04 %II

Measure: For the calibration of AC current source devices using a reference current shunt and a digital multimeter. 
See Annex E for details.

Current, AC: Table 3
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
100 mA to 1.1 A45 Hz to 100 kHz0.011 % to 0.12 %II
1 A to 11 A45 Hz to 100 kHz0.011 % to 0.13 %II

Source: For the calibration of AC current measurement devices using a reference current shunt and a digital multimeter. 
See Annex F for details.

Current, AC: Table 4
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
100 mA to 0.33 A45 Hz to 30 kHz0.017 % to 0.064 %II
0.33 A to 1.1 A45 Hz to 10 kHz0.011 % to 0.025 %II
1 A to 3 A45 Hz to 10 kHz0.018 % to 0.046 %II
3 A to 11 A45 Hz to 5 kHz0.011 % to 0.026 %II

Measure: For the calibration of AC current source devices using a reference current transducer and a digital multimeter. 
See Annex G for details.

Current, AC: Table 5
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
100 A to 2000 A60 Hz0.1 % to 0.3 %II

Source: For the calibration of AC current measurement devices using a reference current transducer and a digital multimeter. 
See Annex G for details.

Current, AC: Table 6
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
100 A to 1500 A60 Hz0.1 % to 0.3 %II

Current, DC

Source: For the calibration of DC current measuring devices using a DC voltage source, resistors and an electrometer. 
See Annex J for details.

Current, DC: Table 1
Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
10 pA to 100 μA0.13 % to 0.018 %II

Measure: For the calibration of DC current source devices using a DC voltage source, resistors and an electrometer. 
See Annex K for details.

Current, DC: Table 2
Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
10 pA to 100 μA0.18 % to 0.022 %II

For the calibration of dc current sources and measurement devices. Both the measure and generate capabilities are monitored using a digital multimeter.

Current, DC: Table 3
Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
1 μA to 10 μA0.002 % + 0.1 nAII
10 μA to 100 μA0.002 % + 0.8 nAII
100 μA to 1 mA0.002 % + 5 nAII
1 mA to 10 mA0.002 % + 50 nAII
10 mA to 100 mA0.0035 % + 500 nAII
100 mA to 1 A0.011 % + 10 μAII

For the calibration of dc current sources and measurement devices. Both the measure and generate capabilities are monitored with a digital multimeter using resistance shunts.

Current, DC: Table 4
Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
1 A to 100 A0.03 %II

Measure and Source: For the calibration of DC current source devices and DC current measurement devices using a reference current shunt and a digital multimeter. 
See Annex H for details.

Current, DC: Table 5
Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
100 mA to 1.1 A0.0026 % to 0.0089 %II
1 A to 11 A0.004 % to 0.09 %II

Measure and Source: For the calibration of DC current source devices and DC current measurement devices using a reference current transducer and a digital multimeter. 
See Annex H for details.

Current, DC: Table 6
Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
100 A to 1000 A0.15 % to 0.31 %II

Power, AC 
At power factor ranging from 1 to 0.5 leading and lagging.

For the calibration of sinusoidal power measurement devices and the ac power function of multifunction calibrators by comparison to a reference wattmeter. This service applies to both leading and lagging power factor (PF).

12 W to 10.08 kW

Measure and generate.

Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
Voltage range: 120 V to 480 V 
Current range: 0.2 A to 21 A
60 Hz0.018 % of readingII

48 W to 12.6 kW

Measure and generate. The uncertainty increases linearly with voltage.

Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
Voltage range: 480 V to 600 V 
Current range: 0.2 A to 21 A
60 Hz0.018 % to 0.024 % of readingII

Power, DC

Generate for the calibration of DC power measurement devices using a multi-function calibrator for the independent generation of current and voltage. 
See Annex D for details.

mW to 330 W

Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
Current range: 0.33 mA to 330 mA 
Voltage range: 0.33 V to 1000 V
0.017 % to 0.028 % of readingII

0.1 W to 3000 W

Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
Current range: 330 mA to 3 A 
Voltage range: 0.33 V to 1000 V
0.026 % to 0.041 % of readingII

1 W to 11 kW

Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
Current range: 3 A to 11 A 
Voltage range: 0.33 V to 1000 V
0.055 % to 0.067 % of readingII

3.6 W to 20.5 kW

Measured Quantity & Range or Instrument FrequencyExpanded Measurement UncertaintyType
Current range: 11 A to 20.5 A 
Voltage range: 0.33 V to 1000 V
0.11 % of readingII

Resistance, DC

For the calibration of resistors and the resistance of temperature probes such as SPRTs and PRTs at the triple point of water using a resistance bridge.

Resistance, DC: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
10 Ω to 500 Ω1 μΩ/ΩII

For the calibration of resistors and resistance measuring devices in air at 23 °C using standard resistors, a nanovolt meter, and a constant current source.

Resistance, DC: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
10 μΩ to 100 μΩ390 μΩ/Ω to 200 μΩ/ΩII
0.1 mΩ to 1 mΩ210 μΩ/Ω to 190 μΩ/ΩII
1 mΩ to 10 mΩ210 μΩ/Ω to 150 μΩ/ΩII
10 mΩ to 100 mΩ350 μΩ/Ω to 98 μΩ/ΩII
0.1 Ω to 1 Ω96 μΩ/Ω to 87 μΩ/ΩII

For the calibration of resistors and resistance measuring devices in air at 23 °C using a multifunction calibrator that is dynamically monitored by a digital multimeter.

Resistance, DC: Table 3
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
1 Ω to 10 Ω(19 μΩ/Ω + 50 μΩ)II
10 Ω to 100 Ω(16 μΩ/Ω + 0.5 mΩ)II
100 Ω to 1 kΩ(14 μΩ/Ω + 0.5 mΩ)II
1 kΩ to 10 kΩ(14 μΩ/Ω + 5 mΩ)II
10 kΩ to 100 kΩ(14 μΩ/Ω + 50 mΩ)II
100 kΩ to 1 MΩ(19 μΩ/Ω + 2 Ω)II
1 MΩ to 10 MΩ(54 μΩ/Ω + 100 Ω)II
10 MΩ to 100 MΩ(500 μΩ/Ω + 1 kΩ)II
100 MΩ to 330 MΩ(0.5 % + 10 kΩ)II

For the measurement of resistance in air at 23 °C using a digital multimeter.

Resistance, DC: Table 4
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
330 MΩ to 1GΩ(0.5 % + 10 kΩ)II

For the calibration of resistors and resistance measuring devices using reference standard resistors and a high value resistance bridge that consists of two programmable DC voltage sources and an electrometer.

Resistance, DC: Table 5
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
100 MΩ to 10 GΩ0.015 %II
10 GΩ 4to 100 GΩ0.015 % to 0.14 %II
100 GΩ to 1 TΩ0.14 %II
1 TΩ to 10 TΩ0.14 % to 0.5 %II

Resistance, AC

Measure and Source 
For the calibration of AC resistors or AC resistance measurement devices using standard AC resistors and an LCR meter. 
See Annex C for details.

Resistance, AC: Table 1
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
1 Ω to 100 kΩ50 Hz to 10 kHz0.04 % to 0.2 % of valueII

Measure and Source 
For the calibration of 1 Ω, 10 Ω, 100 Ω, 1 kΩ and 10 kΩ AC resistors by substitution method using an LCR meter. GenRad 1689 LCR meter and standard resistors (1 Ω, 10 Ω, 100 Ω, 1 kΩ and 10 kΩ). 
See Annex C for details.

Resistance, AC: Table 2
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
1 Ω60 Hz, 1 kHz & 10 kHz0.09 % to 0.2 %II
10 Ω60 Hz, 1 kHz & 10 kHz0.04 % to 0.08 %II
100 Ω60 Hz, 1 kHz & 10 kHz0.04 % to 0.07 %II
1 KΩ60 Hz, 1 kHz & 10 kHz0.04 % to 0.07 %II
10 KΩ60 Hz, 1 kHz & 10 kHz0.04 % to 0.07 %II

Voltage, AC

For the calibration of ac voltage sources. The measure capability is monitored with a digital multimeter. 
See Annex B.

Voltage, AC: Table 1
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
1 mV to 10 mV10 Hz to 1 MHz2.4 % to 0.04 %II
10 mV to 100 mV10 Hz to 1 MHz2.0 % to 0.008 %II
0.1 V to 1 V10 Hz to 1 MHz1.5 % to 0.007 %II
1 V to 10 V10 Hz to 1 MHz1.5 % to 0.007 %II
10 V to 100 V10 Hz to 1 MHz3.5 % to 0.008 %II
100 V to 1050 V10 Hz to 100 kHz0.13 % to 0.012 %II

For the calibration of ac measurement devices. The generate capability is monitored with a digital multimeter. 
See Annex L.

Voltage, AC: Table 2
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
1 mV to 10 mV10 Hz to 500 kHz2.4 % to 0.04 %II
10 mV to 100 mV10 Hz to 500 kHz2.0 % to 0.008 %II
0.1 V to 1 V10 Hz to 500 kHz1.5 % to 0.007 %II
1 V to 3.3 V10 Hz to 500 kHz1.8 % to 0.008 %II
3.3 V to 10 V10 Hz to 100 kHz1.3 % to 0.007 %II
10 V to 33 V10 Hz to 100 kHz0.1 % to 0.009 %II
33 V to 100 V45 Hz to 100 kHz0.07 % to 0.008 %II
100 V to 330 V45 Hz to 100 kHz0.13 % to 0.017 %II
330 V to 1020 V45 Hz to 10 kHz0.08 % to 0.012 %II

Measure: For the calibration of voltage sources and voltage measuring devices using high voltage divider and digital multimeter. 
This service is also available on-site.

Voltage, AC: Table 3
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
1 kV to 10 kV50 Hz to 60 Hz0.38 %II

Source: For the calibration of voltage sources and voltage measuring devices using high voltage divider and digital multimeter.

Voltage, AC: Table 4
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
1 kV to 10 kV50 Hz to 60 Hz0.38 %II

Measure: For the calibration of voltage sources using high voltage divider and digital multimeter. 
This service is also available on-site.

Voltage, AC: Table 5
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
10 kV to 85 kV60 Hz0.38 %II

Voltage, DC

For the calibration of voltage sources and voltage measuring devices using a multifunction calibrator that is monitored with a digital multimeter.

Voltage, DC: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
1 mV to 100 mV(9 μV/V + 1 μV)II
0.1 V to 1 V(8 μV/V + 1 μV)II
1 V to 10 V(8 μV/V + 2 μV)II
10 V to 100 V(10 μV/V + 30 μV)II
100 V to 1000 V(10 μV/V + 100 μV)II

Measure: For the calibration of voltage sources using high voltage divider and digital multimeter. 
This service is also available on-site.

Voltage, DC: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
1 kV to 120 kV0.02 %II

Source: For the calibration of voltage measuring devices using high voltage divider and digital multimeter.

Voltage, DC: Table 3
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
1 kV to 70 kV0.02 %II

Capacitance

For the calibration of capacitors and capacitance measuring devices of the same nominal value by comparison to reference standard capacitors using a substitution method.

Capacitance: Table 1
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
10 pF1 kHz0.025 %II
100 pF1 kHz0.006 %II
1000 pF1 kHz0.006 %II
0.01 μF1 kHz0.015 %II
0.1 μF1 kHz0.015 %II
1 μF1 kHz0.017 %II

Measurement using a constant current and DMM. Suitable for the calibration of capacitors, capacitance measuring devices, and the capacitance function of multifunction calibrators such as a Fluke 5520A.

Capacitance: Table 2
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
100 μF to 110 mFDC0.025 % of valueII

For the calibration of capacitors and capacitance measuring devices using an LCR meter. See Annex I for details.

Capacitance: Table 3
Measured Quantity & Range or InstrumentFrequencyExpanded Measurement UncertaintyType
10 pF to 65 μF50 Hz to 10 kHz0.015 % to 13 %II
65 μF to 100 μF50 Hz to 5 kHz0.072 % to 2.3 %II

Electrical Calibration by Simulation

Conductivity

By electrical simulation using resistance reference standards without realization or measurement of conductivity

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
1 μS/cm to 100 mS/cm0.07 %II

pH meters

For the calibration of pH meters by electrical simulation without realization or measurement of pH

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0 pH to 14 pH0.01 pHII

pH simulators

For the calibration of pH simulators by electrical simulation without realization or measurement of pH

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0 mV to 100 mV0.01 mVII
100 mV to 415 mV0.02 mVII

Electrical Calibration of Temperature Indicators and Sources

By electrical simulation using a multifunction calibrator without realisation or measurement of physical temperatures. Temperature sensors are not calibrated. 
The International Temperature Scale of 1990 (ITS-90) is used to express temperature values. 
Standard reference functions published in NIST Monograph 175:1993 and ASTM E230-98 are used to convert voltage or resistance signals to units of simulated temperature. Results are reported in simulated degrees Celsius or the equivalent on the Fahrenheit scale.

Type E

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-250 °C to -100 °C0.44 °CII
-100 °C to -25 °C0.14 °CII
-25 °C to 350 °C0.12 °CII
350 °C to 650 °C0.14 °CII
650 °C to 1000 °C0.19 °CII

Type J

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-210 °C to -100 °C0.25 °CII
-100 °C to -30 °C0.14 °CII
-30 °C to 150 °C0.12 °CII
150 °C to 760 °C0.15 °CII
760 °C to 1200 °C0.21 °CII

Type K

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-200 °C to -100 °C0.31 °CII
-100 °C to -25 °C0.18 °CII
-25 °C to 120 °C0.15 °CII
120 °C to 1000 °C0.23 °CII
1000 °C to 1372 °C0.35 °CII

Type N

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-200 °C to -100 °C0.35 °CII
-100 °C to -25 °C0.20 °CII
-25 °C to 120 °C0.18 °CII
120 °C to 410 °C0.17 °CII
410 °C to 1300 °C0.25 °CII

Type R

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0 °C to 250 °C0.58 °CII
250 °C to 400 °C0.37 °CII
400 °C to 1000 °C0.35 °CII
1000 °C to 1760 °C0.38 °CII

Type S

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0 °C to 250 °C0.56 °CII
250 °C to 1000 °C0.38 °CII
1000 °C to 1400 °C0.37 °CII
1400 °C to 1760 °C0.42 °CII

Type T

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-250 °C to -150 °C0.55 °CII
-150 °C to 0 °C0.21 °CII
0 °C to 120 °C0.14 °CII
120 °C to 400 °C0.12 °CII

Resistance Temperature Devices

Indicators only. By electrical simulation using a multifunction calibrator without realisation or measurement of physical temperatures. Temperature sensors are not calibrated. 
The International Temperature Scale of 1990 (ITS-90) is used to express temperature values. 
Standard reference functions published in NIST Monograph 175:1993 and ASTM E230-98 are used to convert voltage or resistance signals to units of simulated temperature. Results are reported in simulated degrees Celsius or the equivalent on the Fahrenheit scale.

Pt 385 (100 Ω)

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-200 °C to 800 °C0.06 °C to 0.27 °CII

Pt 385 (200 Ω)

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-190 °C to 630 °C0.05 °C to 0.18 °CII

Pt 385 (500 Ω)

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-190 °C to 630 °C0.05 °C to 0.13 °CII

Pt 385 (1000 Ω)

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-190 °C to 630 °C0.036 °C to 0.27 °CII

Time and Frequency

In this section

Frequency

For the calibration of frequency standards, frequency counters, and frequency measuring devices using a Rubidium frequency standard and a frequency counter.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
10 MHz3.3 x 10-10I
1 Hz to 550 MHz3.3 x 10-10I

Timers

For the calibration of manually-activated time interval measurement devices, including stopwatches and timers using a frequency counter and a quartz reference standard that is referenced to a Rubidium frequency standard.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Timebase error0.16 sII

Thermometry

For the realisation of a reference ice point in accordance with ASTM E563.

Thermometry: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.000 °C6 mKII

For the calibration of RTD sensors using a reference ice point in accordance with ASTM E563. This service is available on-site.

Thermometry: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.000 °C7 mKII

For the calibration of long-stem SPRTs. The uncertainty associated with temperatures between the fixed points is obtained by considering the effect of the error propagation associated with the ITS-90 equations.

Thermometry: Table 3
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Hg Triple Point2.0 mKII
H2O Triple Point0.5 mKII
Ga Melting Point2.5 mKII
In Freezing Point2.0 mKII
Sn Freezing Point2.3 mKII
Zn Freezing Point3.7 mKII

For the calibration of RTD sensors and direct reading thermometers, using a liquid bath or dry block and reference SPRT thermometer.

Thermometry: Table 4
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-95 °C to -40 °C0.065 °CII
-40 °C to 0 °C0.010 °CII
0 °C to 40 °C0.010 °CII
40 °C to 100 °C0.007 °CII
100 °C to 200 °C0.010 °CII
200 °C to 300 °C0.017 °CII
300 °C to 425 °C0.29 °CII
425 °C to 650 °C0.46 °CII

For the calibration of Type E, J, K, N, R, S, and T thermocouples using a liquid bath or dry block and reference PRT thermometer.

Thermometry: Table 5
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-95 °C to 0 °C0.35 °C to 0.20 °CII
0 °C to 400 °C0.2 °C to 0.6 °CII
400 °C to 600 °C0.6 °C to 1.4 °CII

For the calibration of liquid-in-glass thermometers using a liquid bath and reference PRT thermometer.

Thermometry: Table 6
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-78 °C to 0 °C0.10 °CII
0 °C to 300 °C0.05 °CII

For on-site service: calibration of direct reading thermometers using a dry block calibrator and reference PRT thermometer.

Thermometry: Table 7
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
35 °C to 75 °C0.10 °CII
75 °C to 150 °C0.20 °CII
150 °C to 250 °C0.40 °CII
250 °C to 400 °C0.60 °CII

For the calibration of temperature block (i.e., dryblock or drywell) calibrators using a standard PRT thermometer following the guidelines of EA 10/13.

Thermometry: Table 8
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-95 °C to -50 °C0.052 °CII
-50 °C to 155 °C0.018 °CII
155 °C to 425 °C0.06 °CII
425 °C to 600 °C0.23 °CII

Infrared radiation thermometers

35 °C to 500 °C

For the calibration of infrared thermometers using an infrared calibrator.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Emissivity 0.9 to 1.0 with spectral band of 8 to 14 μm(0.37 °C + 0.0037T) T in°CII

95 °F to 932 °F

For the calibration of infrared thermometers using an infrared calibrator.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Emissivity 0.9 to 1.0 with spectral band of 8 to 14 μm(0.55 °F + 0.0037T) 
T in°F
II

Temperature chambers

For the characterization of temperature chambers. 
This service is also available on-site.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-40 °C to 0 °C0.6 °CII
0 °C to 200 °C1.1 °CII
200 °C to 250 °C1.4 °CII
250 °C to 400 °C2.1 °CII
400 °C to 600 °C3.0 °CII

Humidity

For the calibration of thermo-hygrometers by comparison to reference relative humidity and temperature sensors.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
5 % RH to 15 % RH at 25 °C1.5 % RHII
15 % RH to 95 % RH at 25 °C(0.833 + 0.0111RH) % RH 
(Note: RH in %)
II
5 °C to 60 °C at 50 % RH0.32 °CII

Pressure

In this section

Pneumatic deadweight tester

Pneumatic deadweight tester by direct comparison against a high precision reference transducer.

Pneumatic deadweight tester: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
25 psi to 100 psi90 parts per 106 or 0.0028 psi 
(whichever is greater)
II

Pneumatic deadweight tester by cross floating against a reference pneumatic deadweight tester.

Pneumatic deadweight tester: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
100 psi to 1000 psi87 parts per 106 or 0.05 psi 
(whichever is greater)
II

Hydraulic deadweight tester

Hydraulic deadweight tester by cross floating against a reference hydraulic deadweight tester.

Hydraulic deadweight tester: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
87 psi to 810 psi92 parts per 106 or 0.025 psi 
(whichever is greater)
II

Hydraulic deadweight tester by cross floating against a reference hydraulic deadweight tester.

Hydraulic deadweight tester: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
810 psi to 16000 psi92 parts per 106 or 0.5 psi 
(whichever is greater)
II

Pressure & Vacuum

Absolute mode

For the calibration of pressure measuring devices with absolute reference. 
Pneumatic pressure by direct comparison against a high precision reference transducer.

Absolute mode: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.02 psi to 15 psi0.012 % of reading or 0.001 psi (whichever is greater)II
15 psi to 45 psi0.013 % of reading or 0.0025 psi (whichever is greater)II
45 psi to 115 psi0.012 % of reading or 0.0062 psi (whichever is greater)II
115 psi to 315 psi0.012 % of reading or 0.019 psi (whichever is greater)II
315 psi to 1015 psi0.012 % of reading or 0.06 psi (whichever is greater)II

Hydraulic pressure by direct comparison against a high precision reference transducer.

Absolute mode: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
2000 psi to 20000 psi2.3 psiII

Gauge mode

For the calibration of pressure measuring devices including differential manometers with atmospheric reference. 
Pneumatic pressure by direct comparison against a high precision reference transducer.

Gauge mode: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-14.5 psi to 0 psi0.0023 psiII
0 psi to 30 psi0.01 % of reading or 0.0018 psi (whichever is greater)II
30 psi to 100 psi0.01 % of reading or 0.0059 psi (whichever is greater)II
100 psi to 300 psi0.01 % of reading or 0.018 psi (whichever is greater)II
300 psi to 1000 psi0.01 % of reading or 0.059 psi (whichever is greater)II

Pneumatic pressure by direct comparison with a dead weight tester.

Gauge mode: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
100 psi to 1015 psi83 parts per 106II

Hydraulic pressure by direct comparison with a dead weight tester.

Gauge mode: Table 3
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
30 psi to 16 000 psi88 parts per 106II

Pneumatic pressure measured on-site by direct comparison against a reference pressure manometer. 
Above 400 psi, the customer must provide the pressure generator.

Gauge mode: Table 4
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
-14 psi to 30 psi0.018 psiII
30 psi to 1000 psi0.12 % of reading or 0.24 psi (whichever is greater)II

Hydraulic pressure by direct comparison against a reference pressure manometer. 
This service is also available on-site.

Gauge mode: Table 5
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0 psi to 300 psi0.12 % of reading or 0.07 psi (whichever is greater)II
300 psi to 10 000 psi0.12 % of reading or 0.47 psi (whichever is greater)II
2000 psi to 20 000 psi2.3 psiII

Dimensional

Gauge block, length

Steel and ceramic, rectangular and square

Comparison to reference gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, 0.01 inches up to 1 inches4.0 μinchI
Metric, 0.254 mm to 25.4 mm0.10 μmI
Inch, 1 inch up to 4 inches(3.52 + 1.66L) μinch 
(Note: L in 'inches')
I
Metric, 25.4 mm up to 101.6 mm(0.0865 + 0.0017L) μm 
(Note: L in 'mm')
I

Steel, rectangular and square

Comparison to reference gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, 4 inches up to 12 inches(2.0 + 2.0L) μinch 
(Note: L in 'inches')
I
Metric, 101.6 mm to 304.8 mm(0.051 + 0.0020L) μm 
(Note: L in 'mm')
I

Micrometer, outside

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, up to 6 inches(45 + 2L) μinch 
(L in inches)
II
Inch, > 6 inches to 12 inches(50+4.2L) μinch 
(L in inches)
II

Caliper, inside

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, up to 12 inches320 μinchII

Caliper, outside

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, up to 8 inches320 μinchII
8 inches to 12inches330 μinchII
12 inches to 18 inches610 μinchII
18 inches to 24 inches650 μinchII
24 inches to 40 inches860 μinchII

Caliper, depth

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, up to 12 inches320 μinchII

Dial indicator

Up to 2 inch travel. Comparison to gauge blocks.

Dial indicator: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.0001 inch graduation22 μinchII
0.0005 inch graduation180 μinchII

Up to 58 mm travel. Comparison to gauge blocks.

Dial indicator: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.001 mm graduation0.38 μmII
0.01 mm graduation1.8 μmII

Digital indicator

Up to 2 inch travel. Comparison to gauge blocks.

Digital indicator: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.0001 inch resolution70 μinchII
0.0005 inch resolution350 μinchII
0.00005 inch resolution37 μinchII

Up to 58 mm travel. Comparison to gauge blocks.

Digital indicator: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.001 mm resolution0.76 μmII
0.01 mm resolution6.9 μmII

Height gauge

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, up to 12 inches300 μinchII
Metric, up to 300 mm0.0076 mmII

Height gauge - Inch, 12 to 24 inches

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.0005 inch resolution350 μinchII
0.001 inch resolution0.0006 inchII

Height gauge - Metric, 300 to 600 mm

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.01 mm resolution0.009 mmII
0.02 mm resolution0.015 mmII

Cylindrical ring gauge, diameter

Comparison to master ring gauge.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, 1 inch to 4 inches(17.0 + 3.0D) μinch, D in inchI
Inch, > 4 inches to 6 inches(17.0 + 2.9D) μinch, D in inchI
Metric, 25.4 mm to 101.6 mm(0.44 + 0.003D) μm, D in mmI
Metric, > 101.6 mm to 152.4 mm(0.42 + 0.003D) μm, D in mmI

Micrometer Setting Standard

Comparison to gauge blocks.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Inch, 1 inch to 11 inches(11 + 3.5L) μin, L in inchI
Metric, 25.4 mm to 279.4 mm(0.27 + 0.0035L) μm, L in mmI

Mass

In this section

Mass, conventional

For the calibration of masses with traceability to the International System of Units (SI) using a mass comparator and an analytical balance.

Calibration results are reported in terms of conventional mass as defined by the Organisation internationale de métrologie légale (OIML):

"The conventional value of the value of the result of weighing a body in air is equal to mass of a standard, of conventionally chosen density (8000 kg/m3), at a conventionally chosen temperature (20 °C), which balances this body at this reference temperature in air of conventionally chosen density (1.2 kg/m3)."

Mass, conventional: Table 1
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
1 mg0.0020 mgI
2 mg0.0021 mgI
3 mg0.0022 mgI
5 mg0.0020 mgI
10 mg0.0020 mgI
20 mg0.0020 mgI
30 mg0.0040 mgI
50 mg0.0021 mgI
100 mg0.0023 mgI
200 mg0.0025 mgI
300 mg0.0048 mgI
500 mg0.0068 mgI
1 g0.0070 mgI
2 g0.0070 mgI
3 g0.014 mgI
5 g0.010 mgI
10 g0.02 mgI
20 g0.04 mgI
30 g0.06 mgI
50 g0.06 mgI
100 g0.09 mgI
200 g0.14 mgI
220 g0.17 mgI
300 g0.22 mgI
500 g0.27 mgI
1000 g0.56 mgI
2000 g1.1 mgI
3000 g1.7 mgI
5000 g2.9 mgI

For the calibration of masses with traceability to the International System of Units (SI).

Calibration results are reported in terms of conventional mass as defined by the Organisation internationale de métrologie légale (OIML):

"The conventional value of the value of the result of weighing a body in air is equal to mass of a standard, of conventionally chosen density (8000 kg/m3), at a conventionally chosen temperature (20 °C), which balances this body at this reference temperature in air of conventionally chosen density (1.2 kg/m3)."

Mass, conventional: Table 2
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
5 g to 1000 g0.015 gI
1000 g to 3000 g0.018 gI
3000 g to 6000 g0.029 gI
6000 g to 10000 g0.09 gI

For the calibration of masses with traceability to the International System of Units (SI).

Calibration results are reported in terms of conventional mass as defined by the Organisation internationale de métrologie légale (OIML):

"The conventional value of the value of the result of weighing a body in air is equal to mass of a standard, of conventionally chosen density (8000 kg/m3), at a conventionally chosen temperature (20 °C), which balances this body at this reference temperature in air of conventionally chosen density (1.2 kg/m3)."

Mass, conventional: Table 3
Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
20 Kg0.09 gI

Scales

Calibration of scales using standard weights.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
9 kg to 12 kg0.2 gII
> 12 kg to 22 kg0.4 gII
> 22 kg to 45 kg1.0 gII
> 45 kg to 90 kg1.7 gII
> 90 kg to 135 kg2.7 gII
> 135 kg to 177 kg3.2 gII
> 177 kg to 220 kg3.9 gII
> 220 kg to 227 kg5.3 gII

Balances (Electronic, non-automatic)

The calibration uses ASTM Class 1 or OIML E2 reference weights. Calibrations are also available on-site.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
Up to 200 mg0.0024 mgII
> 200 mg to 500 mg0.0047 mgII
> 500 mg to 2 g0.0070 mgII
> 2 g to 5 g0.011 mgII
> 5 g to 10 g0.018 mgII
> 10 g to 20 g0.034 mgII
> 20 g to 50 g0.052 mgII
> 50 g to 100 g0.11 mgII
> 100 g to 200 g0.13 mgII
> 200 g to 500 g0.25 mgII
> 500 g to 1 kg0.5 mgII
> 1 kg to 2 kg1.0 mgII
> 2 kg to 5 kg3.2 mgII
> 5 kg to 9 kg5.3 mgII

Force

Calibration of force measuring instruments in tension and compression using standard weights. This service is also available on-site.

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0 pound force to 1 pound force0.0002 pound forceII
1 pound force to 5 pound force0.001 pound forceII
5 pound force to 20 pound force0.024 % of readingII
20 pound force to 250 pound force0.02 % of readingII

Torque. Clockwise and Counter clockwise

For the calibration of torque tools by comparison with a reference torque transducer

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyType
0.4 lbf•in to 2.0 lbf•in1.5 % of readingII
2.0 lbf•in to 12.5 lbf•in0.70 % of readingII
12.5 lbf•in to 250 lbf•in0.50 % of readingII
12.5 lbf•ft to 250 lbf•ft0.80 % of readingII

Annex A

Calibration Measurement Capability in the Measurement of AC Current and in the Calibration of Digital Multimeters and AC Current Measuring Devices, expressed as ± { (% of reading) + residual }

Current, ACFrequency
45 Hz to 100 Hz100 Hz to 5 kHz5 kHz to 10 kHz
100 μA to 1 mA0.06 % + 0.2 μA0.03 % + 0.2 μA0.07 % + 0.2 μA
1 mA to 10 mA0.06 % + 2 μA0.03 % + 2 μA0.07 % + 2 μA
10 mA to 100 mA0.06 % + 20 μA0.03 % + 20 μA0.07 % + 20 μA
0.1 A to 1 A0.08 % + 0.2 mA0.1 % + 0.2 mAn/a

Annex B

Calibration Measurement Capability in the Measurement of AC Voltage, expressed as ± {(% of reading) + residual}

Voltage, ACFrequency
10 Hz to 2 kHz2 kHz to 10 kHz10 kHz to 30 kHz30 kHz to 100 kHz100 kHz to 300 kHz300 kHz to 1 MHz
1 mV to 10 mV0.025 % +1.1 μV0.033 % +1.1 μV0.034 % +1.1 μV0.3 % +1.1 μV1 % + 4 μV2 % + 4 μV
10 mV to 100 mV0.0068 % + 0.5 μV0.011 % + 0.5 μV0.021 % + 1 μV0.051 % +5 μV0.2 % +30 μV1 % + 100 μV
0.1 V to 1 V0.0064 % +5 μV0.011 % + 5 μV0.021 % + 10 μV0.051 % + 50 μV0.2 % + 300 μV1 % + 1 mV
1 V to 10 V0.0064 % +50 μV0.011 % + 50 μV0.021 % + 100 μV0.051 % + 500 μV0.2 % + 3 mV1 % + 10 mV
10 V to 100 V0.007 % +500 μV0.009 % + 500 μV0.021 % + 1 mV0.051 % + 5 mV0.35 % + 50 mV1 % + 500 mV
100 V to 1050 V0.009 % +25 mV0.009 % + 25 mV0.021 % + 25 mV0.051 % + 100 mVn/an/a

Annex C

Uncertainty in the Measurement of AC Resistance Expressed as % of Value

Resistance (Ω)Frequency
0.05 kHz0.10 kHz0.20 kHz0.40 kHz0.60 kHz0.80 kHz1.00 kHz2.00 kHz4.00 kHz6.00 kHz8.00 kHz10.00 kHz
10.200.200.180.160.130.100.080.090.120.140.170.20
20.200.200.180.160.130.100.080.090.120.140.170.20
30.200.200.180.160.130.100.080.090.120.140.170.20
40.200.200.180.160.130.100.080.090.120.140.170.20
50.200.200.180.160.130.100.080.090.120.140.170.20
60.200.200.180.160.130.100.080.090.120.140.170.20
70.200.200.180.160.130.100.080.090.120.140.170.20
80.200.200.180.160.130.100.080.090.120.140.170.20
90.200.200.180.160.130.100.080.090.120.140.170.20
100.080.070.070.060.050.050.040.040.050.060.070.08
200.080.070.070.060.050.050.040.040.050.060.070.08
300.080.070.070.060.050.050.040.040.050.060.070.08
400.080.070.070.060.050.050.040.040.050.060.070.08
500.080.070.070.060.050.050.040.040.050.060.070.08
600.080.070.070.060.050.050.040.040.050.060.070.08
700.080.070.070.060.050.050.040.040.050.060.070.08
800.080.070.070.060.050.050.040.040.050.060.070.08
900.080.070.070.060.050.050.040.040.050.060.070.08
1000.070.060.060.050.050.040.040.040.050.050.060.07
2000.070.060.060.050.050.040.040.040.050.050.060.07
3000.070.060.060.050.050.040.040.040.050.050.060.07
4000.070.060.060.050.050.040.040.040.050.050.060.07
5000.070.060.060.050.050.040.040.040.050.050.060.07
6000.070.060.060.050.050.040.040.040.050.050.060.07
7000.070.060.060.050.050.040.040.040.050.050.060.07
8000.070.060.060.050.050.040.040.040.050.050.060.07
9000.070.060.060.050.050.040.040.040.050.050.060.07
10000.070.060.060.060.050.050.040.050.060.060.070.07
20000.070.060.060.060.050.050.040.050.060.060.070.07
30000.070.060.060.060.050.050.040.050.060.060.070.07
40000.070.060.060.060.050.050.040.050.060.060.070.07
50000.070.060.060.060.050.050.040.050.060.060.070.07
60000.070.060.060.060.050.050.040.050.060.060.070.07
70000.070.060.060.060.050.050.040.050.060.060.070.07
80000.070.060.060.060.050.050.040.050.060.060.070.07
90000.070.060.060.060.050.050.040.050.060.060.070.07
100000.070.070.070.060.060.060.050.140.140.140.140.14
200000.070.070.070.060.060.060.050.140.140.140.140.14
300000.070.070.070.060.060.060.050.140.140.140.140.14
400000.070.070.070.060.060.060.050.140.140.140.140.14
500000.070.070.070.060.060.060.050.140.140.140.140.14
600000.070.070.070.060.060.060.050.140.140.140.140.14
700000.070.070.070.060.060.060.050.140.140.140.140.14
800000.070.070.070.060.060.060.050.140.140.140.140.14
900000.070.070.070.060.060.060.050.140.140.140.140.14
1000000.070.070.070.060.060.060.050.140.140.140.140.14

Annex D

Uncertainty in the Measurement of DC Power Expressed as % of Value

CurrentVoltage
0.33 V0.5 V1 V3 V5 V10 V30 V50 V100 V300 V500 V1000 V
0.33 mA0.0280.0280.0280.0270.0280.0270.0270.0280.0280.0270.0280.028
0.5 mA0.0230.0230.0230.0230.0230.0230.0230.0230.0230.0230.0230.023
1.0 mA0.0200.0190.0190.0190.0190.0190.0190.0190.0190.0190.0190.019
3.0 mA0.0180.0170.0170.0170.0170.0170.0170.0170.0170.0170.0170.017
5.0 mA0.0200.0190.0190.0190.0190.0190.0190.0190.0190.0190.0190.019
10 mA0.0180.0180.0180.0180.0180.0180.0180.0180.0180.0180.0180.018
30 mA0.0170.0170.0170.0170.0170.0170.0170.0170.0170.0170.0170.017
50 mA0.0200.0190.0190.0190.0190.0190.0190.0190.0190.0190.0190.019
100 mA0.0180.0180.0180.0180.0180.0180.0180.0180.0180.0180.0180.018
320 mA0.0170.0170.0170.0170.0170.0170.0170.0170.0170.0170.0170.017
0.33 A0.0340.0340.0340.0340.0340.0340.0340.0340.0340.0340.0340.034
0.5 A0.0300.0300.0300.0300.0300.0300.0300.0300.0300.0300.0300.030
1.0 A0.0270.0270.0270.0260.0270.0260.0260.0270.0260.0260.0270.026
2.0 A0.0410.0410.0410.0410.0410.0410.0410.0410.0410.0410.0410.041
2.9 A0.0410.0410.0410.0410.0410.0410.0410.0410.0410.0410.0410.041
3.0 A0.0670.0670.0670.0670.0670.0670.0670.0670.0670.0670.0670.067
5.0 A0.0600.0600.0600.0600.0600.0600.0600.0600.0600.0600.0600.060
10.9 A0.0550.0550.0550.0550.0550.0550.0550.0550.0550.0550.0550.055
11.0 A0.1130.1130.1130.1130.1130.1130.1130.1130.1130.1130.1130.113
15.0 A0.1090.1090.1090.1090.1090.1090.1090.1090.1090.1090.1090.109
20.0 A0.1070.1070.1070.1070.1070.1070.1070.1070.1070.1070.1070.107

Annex E

Uncertainty in the Measurement of AC Current Expressed as % of Value

Annex E: Table 1
Current (A)Frequency
0.045 kHz0.1 kHz0.2 kHz0.5 kHz1 kHz2 kHz5 kHz10 kHz20 kHz30 kHz50 kHz100 kHz
0.10.0370.0370.0370.0370.0450.0450.0450.0450.0640.0640.120.12
0.20.0230.0230.0230.0230.0310.0310.0310.0310.0490.0490.110.11
0.30.0180.0180.0180.0180.0260.0260.0260.0260.0440.0440.100.10
0.40.0160.0160.0160.0160.0240.0240.0240.0240.0420.0420.100.10
0.50.0140.0140.0140.0140.0220.0220.0220.0220.0410.0410.0980.098
0.60.0130.0130.0130.0130.0210.0210.0210.0210.0400.0400.0970.097
0.70.0120.0120.0120.0120.0210.0210.0210.0210.0390.0390.0960.096
0.80.0120.0120.0120.0120.0200.0200.0200.0200.0380.0380.0960.096
0.90.0120.0120.0120.0120.0200.0200.0200.0200.0380.0380.0960.096
10.0110.0110.0110.0110.0190.0190.0190.0190.0380.0380.0950.095
1.10.0110.0110.0110.0110.0190.0190.0190.0190.0370.0370.0950.095

Uncertainty in the Measurement of AC Current Expressed as % of Value

Annex E: Table 2
Current (A)Frequency
0.045 kHz0.1 kHz0.2 kHz0.5 kHz1 kHz2 kHz5 kHz10 kHz20 kHz30 kHz50 kHz100 kHz
10.0370.0370.0370.0370.0450.0450.0460.0460.0640.0640.120.12
20.0230.0230.0230.0230.0310.0310.0310.0310.0500.0500.110.11
30.0180.0180.0180.0180.0260.0260.0260.0270.0450.0450.100.10
40.0160.0160.0160.0160.0240.0240.0240.0240.0430.0430.100.10
50.0140.0140.0140.0140.0220.0220.0230.0230.0410.0410.0990.099
60.0130.0130.0130.0130.0210.0210.0220.0220.0400.0400.0980.098
70.0130.0130.0130.0130.0210.0210.0210.0210.0400.0400.0970.097
80.0120.0120.0120.0120.0200.0200.0200.0210.0390.0390.0960.097
90.0120.0120.0120.0120.0200.0200.0200.0200.0390.0390.0960.096
100.0120.0120.0120.0120.0190.0190.0190.0200.0380.0380.0960.096
110.0110.0110.0110.0110.0190.0190.0190.0200.0380.0380.0950.096

Annex F

Uncertainty in the Measurement of AC Current Expressed as % of Value

Annex F: Table 1
Current (A)Frequency
0.045 kHz0.1 kHz0.2 kHz0.5 kHz1 kHz2 kHz5 kHz10 kHz20 kHz30 kHz
0.10.0370.0370.0370.0370.0450.0450.0450.0450.0640.064
0.20.0230.0230.0230.0230.0310.0310.0310.0310.0490.049
0.30.0180.0180.0180.0180.0260.0260.0260.0260.0440.044
0.330.0170.0170.0170.0170.0250.0250.0250.0250.0430.043
0.40.0160.0160.0160.0160.0240.0240.0240.024n/an/a
0.50.0140.0140.0140.0140.0220.0220.0220.022n/an/a
0.60.0130.0130.0130.0130.0210.0210.0210.021n/an/a
0.70.0120.0120.0120.0120.0210.0210.0210.021n/an/a
0.80.0120.0120.0120.0120.0200.0200.0200.020n/an/a
0.90.0120.0120.0120.0120.0200.0200.0200.020n/an/a
10.0110.0110.0110.0110.0190.0190.0190.019n/an/a
1.10.0110.0110.0110.0110.0190.0190.0190.019n/an/a

Uncertainty in the Measurement of AC Current Expressed as % of Value

Annex F: Table 2
Current (A)Frequency
0.045 kHz0.1 kHz0.2 kHz0.5 kHz1 kHz2 kHz5 kHz10 kHz
10.0370.0370.0370.0370.0450.0450.0460.046
20.0230.0230.0230.0230.0310.0310.0310.031
30.0180.0180.0180.0180.0260.0260.0260.027
40.0160.0160.0160.0160.0240.0240.024n/a
50.0140.0140.0140.0140.0220.0220.023n/a
60.0130.0130.0130.0130.0210.0210.022n/a
70.0130.0130.0130.0130.0210.0210.021n/a
80.0120.0120.0120.0120.0200.0200.020n/a
90.0120.0120.0120.0120.0200.0200.020n/a
100.0120.0120.0120.0120.0190.0190.019n/a
110.0110.0110.0110.0110.0190.0190.019n/a

Annex G

Uncertainty in the Measurement of AC Current Expressed as % of Value. Current from 100 A to 1500 A applies to both source and measure. Current from 1500 A to 2000 A applies to measure only.

Current (A)Frequency 60 Hz
1000.30
1200.26
1400.22
1600.20
1800.18
2000.17
2200.16
2400.15
2600.14
2800.13
3000.25
3200.24
3400.23
3600.22
3800.21
4000.21
4200.20
4400.19
4600.19
4800.18
5000.18
5200.17
5400.17
5600.16
5800.16
6000.16
6200.16
6400.15
6600.15
6800.15
7000.14
7200.14
7400.14
7600.14
7800.14
8000.13
8200.13
8400.13
8600.13
8800.13
9000.13
9200.13
9400.12
9600.12
9800.12
10000.12
10200.12
10400.12
10600.12
10800.12
11000.12
11200.11
11400.11
11600.11
11800.11
12000.11
12200.11
12400.11
12600.11
12800.11
13000.11
13200.11
13400.11
13600.11
13800.10
14000.10
14200.10
14400.10
14600.10
14800.10
15000.10
15200.10
15400.10
15600.10
15800.10
16000.10
16200.10
16400.10
16600.10
16800.10
17000.10
17200.10
17400.10
17600.10
17800.10
18000.10
18200.10
18400.10
18600.10
18800.10
19000.10
19200.10
19400.10
19600.10
19800.10
20000.11

Annex H

Uncertainty in the Measurement of DC Current

Annex H: Table 1
Current (A)Uncertainty % of Value
0.10.0089
0.20.0049
0.30.0037
0.40.0033
0.50.0030
0.60.0029
0.70.0028
0.80.0027
0.90.0027
1.00.0027
1.10.0026

Uncertainty in the Measurement of DC Current

Annex H: Table 2
Current (A)Uncertainty % of Value
10.009
20.006
30.005
40.004
50.004
60.004
70.004
80.004
90.004
100.004
110.004

Uncertainty in the Measurement of DC Current

Annex H: Table 3
Current (A)Uncertainty % of Value
1000.31
1100.31
1200.30
1300.30
1400.30
1500.29
1600.29
1700.29
1800.29
1900.28
2000.28
2100.28
2200.27
2300.27
2400.27
2500.26
2600.26
2700.26
2800.25
2900.25
3000.25
3100.24
3200.24
3300.24
3400.24
3500.23
3600.23
3700.23
3800.22
3900.22
4000.22
4100.22
4200.21
4300.21
4400.21
4500.20
4600.20
4700.20
4800.20
4900.19
5000.19
5100.19
5200.19
5300.19
5400.19
5500.19
5600.18
5700.18
5800.18
5900.18
6000.18
6100.18
6200.18
6300.18
6400.18
6500.18
6600.18
6700.18
6800.18
6900.18
7000.17
7100.17
7200.17
7300.17
7400.17
7500.17
7600.17
7700.17
7800.17
7900.17
8000.17
8100.16
8200.16
8300.16
8400.16
8500.16
8600.16
8700.16
8800.16
8900.16
9000.16
9100.16
9200.16
9300.16
9400.15
9500.15
9600.15
9700.15
9800.15
9900.15
10000.15

Annex I

Uncertainty in the Measurement of Capacitance Expressed as % of Value

CapacitanceFrequency
0.05 kHz0.1 kHz0.2 kHz0.4 kHz0.6 kHz0.8 kHz1 kHz2 kHz4 kHz6 kHz8 kHz10 kHz
10 pF0.0700.0270.0250.0200.0170.0150.0150.0150.0210.0310.0420.053
20 pF136.43.21.10.710.540.210.320.380.220.320.22
30 pF7.53.71.90.620.420.310.130.190.220.120.150.084
60 pF2.11.10.540.180.120.0900.0370.0540.0640.0320.0420.053
80 pF0.800.400.200.0690.0460.0350.0180.0230.0290.0310.0420.053
100 pF0.0290.0270.0240.0200.0160.0150.0170.0150.0210.0310.0420.053
200 pF2.21.10.580.210.130.0930.0470.0260.0290.0370.0460.056
300 pF1.30.670.350.130.0810.0540.0320.0250.0290.0360.0460.056
600 pF0.390.210.120.0560.0310.0250.0250.0250.0290.0360.0460.056
800 pF0.160.0950.0610.0380.0260.0250.0250.0250.0290.0360.0460.056
1 nF0.0340.0330.0320.0290.0270.0250.0250.0250.0290.0360.0460.056
2 nF0.140.0640.0400.0320.0260.0220.0200.0220.0340.0500.0680.087
3 nF0.0880.0480.0400.0320.0260.0220.0200.0220.0340.0500.0680.087
6 nF0.0500.0440.0400.0320.0260.0220.0200.0220.0330.0500.0680.087
8 nF0.0460.0440.0400.0320.0260.0220.0200.0220.0330.0500.0680.087
10 nF0.0460.0440.0400.0320.0260.0220.0200.0220.0330.0500.0680.087
20 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
30 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
60 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
80 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
100 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
200 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
300 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
600 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
800 nF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
1 µF0.0560.0540.0490.0390.0310.0250.0230.0250.0410.0620.0840.11
2 µF0.0640.0630.0610.0510.0480.0460.0340.0460.0720.0980.160.20
3 µF0.0640.0630.0610.0510.0480.0460.0340.0460.0730.0980.160.20
6 µF0.0640.0630.0610.0510.0480.0460.0340.0460.0730.100.170.21
8 µF0.0640.0630.0610.0510.0480.0460.0340.0460.0740.100.170.20
10 µF0.0640.0630.0610.0510.0480.0460.0340.0460.0730.0980.160.20
20 µF0.100.100.100.0960.110.120.0720.240.670.991.72.2
30 µF0.100.100.100.0960.110.120.0730.240.670.991.82.2
60 µF0.100.100.100.0960.110.120.0760.250.681.01.82.3
80 µF0.100.100.100.0970.110.120.0740.240.67N/AN/AN/A
100 µF0.100.100.100.0960.110.120.0720.240.66N/AN/AN/A

Annex J

Uncertainty in the Measurement of DC Current Expressed as % of Value

Current Range (A)% of Full Scale
10 %20 %30 %40 %50 %60 %70 %80 %90 %100 %
10 pA to 100 pA0.130.0950.0910.0920.0890.0880.0940.0940.0930.093
0.1 nA to 1 nA0.0600.0610.0600.0600.0600.0600.0600.0600.0600.060
1 nA to 10 nA0.0330.0310.0310.0310.0310.0310.0310.0310.0310.031
10 nA to 100 nA0.0320.0310.0310.0310.0310.0310.0310.0310.0310.031
0.1 µA to 1 µA0.0280.0240.0230.0230.0230.0230.0220.0220.0220.022
1 µA to 10 µA0.0240.0200.0200.0190.0190.0190.0190.0190.0190.019
10 µA to 100 µA0.0230.0190.0190.0190.0180.0180.0180.0180.0180.018

Annex K

Uncertainty in the Measurement of DC Current Expressed as % of Value

Current Range (A)% of Full Scale
10 %20 %30 %40 %50 %60 %70 %80 %90 %100 %
10 pA to 100 pA0.180.130.120.120.120.120.120.120.120.12
0.1 nA to 1 nA0.0880.0930.0930.0920.0920.0920.0920.0920.0920.092
1 nA to 10 nA0.0450.0450.0440.0440.0480.0470.0470.0470.0470.047
10 nA to 100 nA0.0450.0450.0450.0440.0440.0440.0480.0480.0480.048
0.1 µA to 1 µA0.0310.0280.0280.0270.0270.0280.0280.0280.0280.027
1 µA to 10 µA0.0260.0230.0230.0230.0230.0230.0230.0230.0230.022
10 µA to 100 µA0.0250.0230.0230.0220.0220.0220.0220.0220.0220.022

Note: Uncertainties may be linearly interpolated between points.

Annex L

Calibration Measurement Capability in the Calibration of Digital Multimeters and AC Voltage Measuring Devices, expressed as ± {(% of reading) + residual}

Annex L: Table 1
Voltage, ACFrequency
10 Hz to 2 kHz2 kHz to 10 kHz10 kHz to 30 kHz30 kHz to 100 kHz100 kHz to 300 kHz300 kHz to 500 kHz
1 mV to 10 mV0.025 % + 1.1 μV0.033 % + 1.1 μV0.034 % + 1.1 μV0.3 % + 1.1 μV1 % + 4 μV2 % + 4 μV
10 mV to 100 mV0.0068 % + 0.5 μV0.011 % + 0.5 μV0.021 % + 1 μV0.051 % + 5 μV0.2 % + 30 μV1 % + 100 μV
0.1 V to 1 V0.0064 % + 5 μV0.011 % + 5 μV0.021 % + 10 μV0.051 % + 50 μV0.2 % + 300 μV1 % + 1 mV
1 V to 3.3 V0.0064 % + 50 μV0.011 % + 50 μV0.021 % + 100 μV0.051 % + 500 μV0.2 % + 3 mV1 % + 10 mV
3.3 V to 10 V0.0064 % + 50 μV0.011 % + 50 μV0.021 % + 100 μV0.051 % + 500 μVn/an/a
10 V to 33 V0.007 % + 500 μV0.009 % + 500 μV0.021 % + 1 mV0.051 % + 5 mVn/an/a

Calibration Measurement Capability in the Calibration of Digital Multimeters and AC Voltage Measuring Devices, expressed as ± {(% of reading) + residual}

Annex L: Table 2
Voltage, ACFrequency
45 Hz to 2 kHz2 kHz to 10 kHz10 kHz to 30 kHz30 kHz to 100 kHz
33 V to 100 V0.007 % +500 μV0.009 % + 500 μV0.021 % + 1 mV0.051 % + 5 mV
100 V to 330 V0.009 % +25 mV0.009 % + 25 mV0.021 % + 25 mV0.051 % + 100 mV
330 V to 1020 V0.009 % +25 mV0.009 % + 25 mVn/an/a