CLAS Certificate Number 2004-01

Company name
Measurements International Ltd.
Company address

118 Commerce Drive
PO Box 2359
Prescott ON K0E 1T0

Contact
Ryan Brown
Telephone: 613-925-5934
Fax: 613-925-1195
Emailryanbrown@mintl.com
Clients served
  • All interested parties
Fields of calibration
  • Electrical
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 15611
  • First issued
  • Issue 6.0e

This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body.

The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS.

See supplementary notes.

Electrical

In this section

Resistance, DC

For the calibration of resistors using a current comparator bridge and associated range extenders.

Resistance, DC - table 1
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 μΩ to 10 μΩ500 to 20 μΩ/ΩI
10 μΩ to 100 μΩ20 to 2 μΩ/ΩI
100 μΩ to 1 mΩ2 to 0.9 μΩ/ΩI
1 mΩ to 10 mΩ0.9 to 0.22 μΩ/ΩI
10 mΩ to 100 mΩ0.22 to 0.17 μΩ/ΩI
100 mΩ to 1 Ω0.17 to 0.1 μΩ/ΩI
1 Ω to 10 Ω0.1 to 0.19 μΩ/ΩI
10 Ω to 100 Ω0.19 to 0.22 μΩ/ΩI
100 Ω to 1 kΩ0.22 to 0.25 μΩ/ΩI
1 kΩ to 10 kΩ0.25 to 0.1 μΩ/ΩI
13 kΩ0.2 μΩ/ΩI

For the calibration of resistors using a high resistance bridge.

Resistance, DC - table 2
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
10 kΩ to 100 kΩ0.15 to 0.27 μΩ/ΩI
100 kΩ to 1 MΩ0.27 to 0.37 μΩ/ΩI
1 MΩ to 10 MΩ0.37 to 0.64 μΩ/ΩI
10 MΩ to 100 MΩ0.64 to 2 μΩ/ΩI
100 MΩ to 1 GΩ2 to 8.5 μΩ/ΩI

For the calibration of resistors using a Dual Source Bridge. Uncertainty will depend on calibration voltage.

Resistance, DC - table 3
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 GΩ to 10 GΩ8.5 to 35 μΩ/ΩI
10 GΩ to 100 GΩ35 to 45 μΩ/ΩI
100 GΩ to 1 TΩ45 to 60 μΩ/ΩI
1 TΩ to 10 TΩ60 to 210 μΩ/ΩI
10 TΩ to 100 TΩ210 to 1000 μΩ/ΩI
100 TΩ to 1 PΩ1000 to 10000 μΩ/ΩI

Voltage DC

For the calibration of DC voltage using:

  • Binary Voltage Divider (BVD)
  • DVM in series opposition
  • Voltage Divider
Voltage DC
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 V1.2 μV/VI
10 V0.7 μV/VI
100 mV to 10 V2 μV/VI
10 V to 30 V2 μV/VI
30 V to 300 V2 μV/VI
300 V to 1000 V2 μV/VI

Voltage Ratio, DC

Measured from full scale to 10 % of full scale in each range using the linearity specification of the DVM 3458A.
The maximum ratio for each range is 10:1.

Voltage Ratio, DC
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 V1.0 μV/VI
10 V0.5 μV/VI

Current DC

Source: For the calibration of current shunts and current sources using associated range extenders.

Current DC - table 1
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 mA to 10 mA5 to 10 μA/AI
10 mA to 100 A3 to 13 μA/AI
100 A to 400 A13 to 50 μA/AI

Source: for the calibration of current measuring devices using a voltage source and resistors.

Current DC - table 2
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
10 pA to 1 mA560 to 5 μA/AI

Resistance Ratio, DC

Measure resistance ratios up to 150 A using range extenders.

Resistance Ratio, DC
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
100:15×10-5I
1000:15×10-4I

10:1 Resistance Ratios

For the calibration of the ratio errors of DC current comparator bridges using resistors to generate a resistance ratio.

10:1 Resistance Ratios
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
0.1 Ω to 1 Ω0.05 parts per 106I
1 Ω to 10 Ω0.05 parts per 106I
10 Ω to 100 Ω0.05 parts per 106I
100 Ω to 1 kΩ0.05 parts per 106I
1 kΩ to 10 kΩ0.05 parts per 106I
1 kΩ to 13 kΩ0.05 parts per 106I
10 kΩ to 100 kΩ0.08 parts per 106I
100 kΩ to 1 MΩ0.1 parts per 106I
1 MΩ to 10 MΩ0.15 parts per 106I
10 MΩ to 100 MΩ1.5 parts per 106I

1:1 Resistance Ratios

For the calibration of the ratio errors of DC current comparator bridges using resistors to generate a resistance ratio.

1:1 Resistance Ratios - table 1
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 Ω to 1 Ω0.03 parts per 106I
10 Ω to 10 Ω0.03 parts per 106I
100 Ω to 100 Ω0.03 parts per 106I
1 kΩ to 1 kΩ0.03 parts per 106I
10 kΩ to 10 kΩ0.03 parts per 106I
100 kΩ to 100 kΩ0.10 parts per 106I
1 MΩ to 1 MΩ0.12 parts per 106I
10 MΩ to 10 MΩ0.15 parts per 106I
100 MΩ to 100 MΩ1.5 parts per 106I

For the calibration of direct current comparator range extenders using resistors to generate a resistance ratio.

1:1 Resistance Ratios - table 2
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 Ω to 0.1 Ω0.2 parts per 106I
10 Ω to 1 Ω0.2 parts per 106I
10 Ω to 0.1 Ω0.2 parts per 106I
10 Ω to 0.01 Ω0.3 parts per 106I
10 Ω to 0.001 Ω0.3 parts per 106I
100 Ω to 0.01 Ω2 parts per 106I
100 Ω to 0.001 Ω10 parts per 106I

High Resistance Meters - Direct Reading

For the calibration of high resistance meters using standard resistors.

High Resistance Meters - Direct Reading
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
1 MΩ15 μΩ/ΩI
10 MΩ15 μΩ/ΩI
100 MΩ15 μΩ/ΩI
1 GΩ45 μΩ/ΩI
10 GΩ150 μΩ/ΩI
100 GΩ200 μΩ/ΩI
1 TΩ200 μΩ/ΩI
10 TΩ500 μΩ/ΩI
100 TΩ1800 μΩ/ΩI

Measurement Ratio 10:1

For the calibration of high resistance meters using standard resistors to generate a resistance ratio.

Measurement Ratio 10:1
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
10 MΩ15 μΩ/ΩI
100 MΩ15 μΩ/ΩI
1 GΩ50 μΩ/ΩI
10 GΩ150 μΩ/ΩI
100 GΩ220 μΩ/ΩI
1 TΩ300 μΩ/ΩI
10 TΩ600 μΩ/ΩI
100 TΩ1400 μΩ/ΩI

Measurement Ratio 100:1

For the calibration of high resistance meters using standard resistors to generate a resistance ratio.

Measurement Ratio 100:1
Measured Quantity and Range or InstrumentExpanded Measurement UncertaintyType of Service
100 MΩ20 μΩ/ΩI
1 GΩ50 μΩ/ΩI
10 GΩ140 μΩ/ΩI
100 GΩ200 μΩ/ΩI
1 TΩ220 μΩ/ΩI
10 TΩ400 μΩ/ΩI
100 TΩ1400 μΩ/ΩI

Resistance, AC

For the calibration of resistors using a high voltage capacitance bridge. On-site calibration available.

Resistance, AC
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
12 kΩ, 24 kΩ50 and 60 Hz25 μΩ/ΩI

Current, AC

For the calibration of AC current measuring devices using a power calibration system.

Current, AC
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
5 mA to 5 A50 and 60 Hz30 μA/AI

Voltage, AC

For the calibration of AC voltage measuring devices using a power calibration system.

Voltage, AC
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
6 V to 300 V50 and 60 Hz30 μV/VI

Capacitance

For the calibration of high voltage capacitors using a high voltage capacitance bridge. Both capacitance and dissipation factor are reported. On-site calibration available.

Capacitance
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
10 pF, 50 pF, 100 pF50 and 60 Hz20 μF/FI
1000 pF50 and 60 Hz15 μF/FI
1200 pF to 1 µF50 and 60 Hz22 μF/FI

Capacitance Bridge Ratio Error

For the calibration of high voltage capacitance bridge ratio errors for voltages up to 100 kV using capacitors to generate a capacitance ratio. Both in-phase and dissipation factor errors are reported.

Capacitance Bridge Ratio Error
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
Available ratios
1:1, 2:1, 5:1,10:1, 20:1, 50:1, 100:1, 200:1, 500:1 and 1000:1
50 and 60 Hz20 parts per 106I

High Voltage Divider Calibration

For the calibration of high voltage dividers using capacitance bridges and standard capacitors. Both in-phase and quadrature errors are reported. On-site calibration available

High Voltage Divider Calibration
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
Available ratios
1:1, 2:1, 5:1, 10:1, 20:1, 50:1 and 100:1
50 and 60 Hz20 parts per 106I

Current Transformers

Comparison to standard current transformer. Both in-phase and quadrature errors are reported. On-site calibration available

Current Transformers
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
Available ratios
1000:1 and 2000:1
50 and 60 Hz15 parts per 106I

Current Ratio

For the calibration of current ratio bridges used for the comparison of current transformers. Using current transformers to generate a current ratio. Both in-phase and quadrature errors are reported.

Current Ratio
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
Available ratios
1:1, 2:1, 3:1, 5:1 and 10:1
50 and 60 Hz25 parts per 106I

Voltage Ratio

For the calibration of voltage ratio bridges used for the comparison of voltage transformers. Using voltage transformers to generate a voltage ratio. Both in-phase and quadrature errors are reported.

Voltage Ratio
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
Available ratios
1:1, 2:1, 5:1, 10:1, 20:1, 50:1 and 100:1
50 and 60 Hz25 parts per 106I

AC Power

For the calibration of ac power measurement devices for voltages up to 240 V and currents up to 5 A at power factors of zero to unity using an AC power calibrator. 
See Annex A.

AC Power
Measured Quantity and Range or InstrumentFrequencyExpanded Measurement UncertaintyType of service
0 W to 1200 W50 and 60 Hz30 μW/WI

Annex A

Examples of best uncertainty, ± (% of applied signal)* for the calibration of ac power measurement devices including watt meters

Annex A
Applied Signal,
Vapp x Iapp
Scale,
Vs x Is
Power Factor, pf
10.50.10
120 V x 5 mA120 V x 5 mA0.0030.0030.00330.0033
120 V x 10 mA120 V x 10 mA0.0030.0030.00330.0033
120 V x 20 mA120 V x 20 mA0.0030.0030.00330.0033
120 V x 50 mA120 V x 50 mA0.0030.0030.00330.0033
120 V x 100 mA120 V x 100 mA0.0030.0030.00330.0033
120 V x 200 mA120 V x 200 mA0.0030.0030.00330.0033
120 V x 500 mA120 V x 500 mA0.0030.0030.00330.0033
120 V x 1 A120 V x 1 A0.0030.0030.00330.0033
120 V x 2 A120 V x 2 A0.0030.0030.00330.0033
120 V x 5 A120 V x 5 A0.0030.0030.00330.0033
240 V x 5 mA240 V x 5 mA0.0030.0030.00330.0033
240 V x 10 mA240 V x 10 mA0.0030.0030.00330.0033
240 V x 20 mA240 V x 20 mA0.0030.0030.00330.0033
240 V x 50 mA240 V x 50 mA0.0030.0030.00330.0033
240 V x 100 mA240 V x 100 mA0.0030.0030.00330.0033
240 V x 200 mA240 V x 200 mA0.0030.0030.00330.0033
240 V x 500 mA240 V x 500 mA0.0030.0030.00330.0033
240 V x 1 A240 V x 1 A0.0030.0030.00330.0033
240 V x 2 A240 V x 2 A0.0030.0030.00330.0033
240 V x 5 A240 V x 5 A0.0030.0030.00330.0033