CLAS Certificate Number 2001-02

Company name
Industrial Technology Centre
Company address
200 – 78 Innovation Drive 
Winnipeg MB  R3T 6C2
Contact
Tom Manson 
Telephone: 204-480-3333 
Toll Free: 800-728-7933 
Email: tmanson@itc.mb.ca
Clients served
  • All interested parties
  • Some calibration services are available on-site. These services are indicated in the "Remarks" column of the following pages.
Fields of calibration
  • Dimensional
  • Pressure
  • Force
  • Mass
  • Hardness
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 15014
  • First issued 1984-06-05 (testing)
  • First issued 2001-09-20 (calibration)
  • Issue 6.5e

This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body.

The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS.

See supplementary notes.

On this page

Dimensional

Type I Capability

Steel gauge block, rectangular and square. length:

ANSI/ASME B89.1.2M, FED SPEC GGG-G-15C, T.O. 33K6-4-1-1, ISO 3650

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 1 in4.6 µin
> 1 in to 4 in(3.5 + 1.1L) µin
(Note: L in 'inches')
> 4 in to 20 in(2.3 + 2.4L) µin
(Note: L in 'inches')
0 mm to 25 mm0.117 µm
> 25 mm to 100 mm(0.089 + 0.0011L) µm
(Note: L in 'millimetres')
> 100 mm to 508 mm(0.058 + 0.0024L) µm
(Note: L in 'millimetres')

Ceramic gauge block, rectangular and square. Length:

ANSI/ASME B89.1.2M, FED SPEC GGG-G-15C, T.O. 33K6-4-1-1, ISO 3650

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 4 in(4.9 + 0.8L) µin
(Note: L in 'inches')
0 mm to 100 mm(0.125 + 0.0008L) µm
(Note: L in 'millimetres')

Chromium Carbide gauge block, rectangular and square. Length:

ANSI/ASME B89.1.2M, FED SPEC GGG-G-15C, T.O. 33K6-4-1-1, ISO 3650

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 4 in(5.3 + 0.6L) µin
(Note: L in 'inches')
0 mm to 100 mm(0.135 + 0.0006L) µm
(Note: L in 'millimetres')

Gauge block, variation in length:

Parallelism of gauge blocks. ANSI/ASME B89.1.2M, FED SPEC GGG-G-15C, T.O. 33K6-4-1-1, ISO 3650

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 12 in1 µin
0 mm to 300 mm0.025 µm

Cylindrical ring gauge, diameter:

ANSI/ASME B89.1.6M

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 21.01 in(12 + 2L) µin
(Note: L in 'inches')
0 mm to 534 mm(0.31 + 0.002L) µm
(Note: L in 'millimetres')

Cylindrical plug gauge, diameter:

ANSI/ASME B89.1.5

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 21.01 in(9 + 2L) µin
(Note: L in 'inches')
0 mm to 534 mm(0.23+ 0.002L) µm
(Note: L in 'millimetres')

Thread wire, diameter:

ANSI/ASME B1.2-App.B (Unified), ANSI/ASME B1.16M-App.B (Metric)

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 1 in12 µin
0 mm to 25 mm0.31 µm

60 degree thread plug gauge and thread setting plug gauge:

Pitch diameter (measured over thread wires)

ANSI/ASME B1.2, ANSI/ASME B1.3, ANSI/ASME B1.15 ANSI/ASME B1.16M, ANSI/ASME B1.22M, ISO 1502
The pitch diameter of a thread plug gauge is determined by measuring the diameter over thread wires inserted in the thread groove on opposite sides of the axis. The preferred term for this measurement is 'thread groove diameter'. Other names for this measurement are 'simple effective diameter' and 'simple pitch diameter'.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 1.5 in100 µin
> 1.5 in to 6 in150 µin
> 6 in to 12 in200 µin
0 mm to 35 mm2.5 µm
> 35 mm to 150 mm3.8 µm
> 150 mm to 300 mm5.1 µm

Major diameter

ANSI/ASME B1.2, ANSI/ASME B1.3, ANSI/ASME B1.15 ANSI/ASME B1.16M, ANSI/ASME B1.22M, ISO 1502
The pitch diameter of a thread plug gauge is determined by measuring the diameter over thread wires inserted in the thread groove on opposite sides of the axis. The preferred term for this measurement is 'thread groove diameter'. Other names for this measurement are 'simple effective diameter' and 'simple pitch diameter'.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 12 in(15 + 4.7L) µin
(Note: L in 'inches')
0 mm to 300 mm(0.38 + 0.0047L) µm
(Note: L in 'millimetres')

Type II Capability

60 degree adjustable thread ring gauge:

ANSI/ASME B1.2, ANSI/ASME B1.16M. ITC Procedure CAL.013.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
Set to thread setting plugThe adjustable thread ring gauge is set to the functional diameter of the thread setting plug

NPT/NPTF Thread plug and ring Standoff

ANSI/ASME B1.2, ANSI/ASME B1.16M. ITC. Procedure CAL.013.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
1/16 in to 2 in200 µin

Micrometer setting standard:

CAN/CGSB-39.18, T.O. 33K6-4-369-1, ASME B89.1.13 ANSI/ASME B89.1.10M.
R is the discernable resolution (for example 1/2, 1/4 or 1/5 of the graduation), Up to 4 inch travel Up to 100 mm travel

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 40 in(9 + 2L) µin
(Note: L in 'inches')
>40 in to 80 in(35 + 3.2L) µin
(Note: L in 'inches')
0 mm to 1016 mm(0.23 + 0.002L) µm
(Note: L in 'millimetres')
>1016 mm to 2032 mm(0.89 + 0.0032L) µm
(Note: L in 'millimetres')

Indicators, Analog:

CAN/CGSB-39.18, T.O. 33K6-4-369-1, ASME B89.1.13 ANSI/ASME B89.1.10M
R is the discernable resolution (for example 1/2, 1/4 or 1/5 of the graduation). Up to 4 inch travel. Up to 100 mm travel

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
Inch, 0.00005 in graduation
(Range: 0 in to 4 in)
(10 + 1.2x10-6R) µin
(Note: R in 'inches')
Metric, 0.001 mm graduation
(Range: 0 mm to 100 mm)
(0.25 + 1200R) µm
(Note: R in 'millimetres')

Indicators, Digital:

ANSI/ASME B89.1.10M, R is the resolution. Up to 4 inch travel.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
Inch, 0.00001 in resolution
(Range: 0 in to 4 in)
(10 + 6.0x105R) µin
(Note: R in 'inches')

Indicators, Digital:

ANSI/ASME B89.1.10M, R is the resolution. Up to 100 mm travel.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
Metric, 0.0002 mm resolution
(Range: 0 mm to 100 mm)
(0.25 + 600R) µm
(Note: R in 'millimetres')

Micrometer: outside

CAN/CGSB-39.18, T.O. 33K6-4-661-1, T.O. 33K6-4-15-1

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 6 in(30 + 3L) µin
(Note: L in 'inches')
> 6 in to 60 in(60 + 5L) µin
(Note: L in 'inches')
0 mm to 150 mm(0.76 + 0.003L) µm
(Note: L in 'millimetres')
> 150 mm to 1500 mm(1.5 + 0.005L) µm
(Note: L in 'millimetres')

Micrometer: inside

CAN/CGSB-39.18, T.O. 33K6-4-661-1, T.O. 33K6-4-15-1

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 24 in(160 + 8L) µin
(Note: L in 'inches')
> 24 in to 60 in(150 + 6L) µin
(Note: L in 'inches')
0 mm to 600 mm(4.1 + 0.008L) µm
(Note: L in 'millimetres')
> 600 mm to 1500 mm(3.8+ 0.006L) µm
(Note: L in 'millimetres')

Micrometer: depth

CAN/CGSB-39.18, T.O. 33K6-4-661-1, T.O. 33K6-4-15-1

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 24 in(240 + 9L) µin
(Note: L in 'inches')
0 mm to 600 mm(6.1 + 0.009L) µm
(Note: L in 'millimetres')

Caliper: outside

CAN/CGSB-39-19

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 60 in(290 + 5L) µin
(Note: L in 'inches')
0 mm to 1500 mm(7.4 + 0.005L) µm
(Note: L in 'millimetres')

Caliper: inside

CAN/CGSB-39-19

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 60 in(300 + 5L) µin
(Note: L in 'inches')
0 mm to 1500 mm(7.6 + 0.005L) µm
(Note: L in 'millimetres')

Caliper: depth

CAN/CGSB-39-19

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 24 in(300 + 6L) µin
(Note: L in 'inches')
0 mm to 600 mm(7.6 + 0.006L) µm
(Note: L in 'millimetres')

Height gauge 0 in to 40 in:

NAVAIR 17-20MD-17, FED SPEC GGG-G111C

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.0001 in resolution(86 + 7L) µin
(Note: L in 'inches')
0.0005 in resolution(295 + 7L) µin
(Note: L in 'inches')
0.001 in resolution(580 + 7L) µin
(Note: L in 'inches')

Height gauge 0 mm to 1000 mm:

NAVAIR 17-20MD-17, FED SPEC GGG-G111C

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.001 mm resolution(1.7 + 0.007L) µm
(Note: L in 'millimetres')
0.01 mm resolution(6.0 + 0.007L) µm
(Note: L in 'millimetres')
0.02 mm resolution(11.7 + 0.007L) µm
(Note: L in 'millimetres')

Extensometer Displacement

ASTM E83. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.5 in to 8 in (up to 25 % strain)32L µin
(Note: L in 'inches')

Actuator Travel Displacement

ASTM E2309. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0 in to 24 in0.002 in

Actuator Travel Speed

ASTM E2658. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.01 in/min to 10 in/min0.25 % of reading

Type III Capability

Surface plate

Federal Specification GGG-P-463. On-site calibration available. D is the diagonal length

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
17 in to 161 in diagonalN/A
Flatness(30 + 0.8D) µin
(Note: D in 'inches')
Repeat reading25 µin

Pressure

Type II Capability

Vacuum

Gauge pressure, comparison to a reference pressure transducer. ITC Procedure CAL.038. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
-15 psi to 0 psi0.05 % of reading or ± 0.05 psi, whichever is greater

Pressure

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyRemarks
0 psi to 300 psi0.05 % of reading or ± 0.05 psi, whichever is greaterGauge pressure, comparison to a reference pressure transducer or dead weight tester.
ITC Procedure CAL.038.
On-site calibration available. Calibrations onsite above 100 psi require client pressure source.
300 psi to 3000 psi0.06 % of reading or ± 0.6 psi, whichever is greaterGauge pressure, comparison to a reference pressure transducer or dead weight tester.
ITC Procedure CAL.038.
On-site calibration available. Calibrations onsite require client pressure source.
3000 psi to 10000 psi0.1 % of readingGauge pressure, comparison to a reference pressure transducer or dead weight tester.
ITC Procedure CAL.038.
On-site calibration available. Calibrations onsite require client pressure source.

Force

Type II Capability

Calibration of load cells – Tension

Comparison to a reference load cell per ITC Procedure CAL.040. In place of the pound-force (lbf) unit, calibration or verification may be performed using the SI unit, the Newton (N). 1 lbf is equivalent to 4.448222N.
Calibration in house to 270000 lbf. Calibration onsite in client load frame to 600000 lbf.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.06 lbf to 10 lbf0.01 lbf
1 lbf to 600000 lbf0.1 % of reading

Calibration of load cells – Compression

Comparison to a reference load cell per ITC Procedure CAL.040. In place of the pound-force (lbf) unit, calibration or verification may be performed using the SI unit, the Newton (N). 1 lbf is equivalent to 4.448222N.
Calibration in house to 270000 lbf. Calibration onsite in client load frame to 600000 lbf.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.06 lbf to 10 lbf0.01 lbf
1 lbf to 600000 lbf0.1 % of reading

Calibration of force testing machines – Tension

ASTM E4. 300 lbf, 2 klbf, 10 klbf, 60 klbf, 270 klbf, and 600 klbf standards.
In place of the pound-force (lbf) unit, calibration or verification may be performed using the SI unit, the Newton (N). 1 lbf is equivalent to 4.448222 N. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.06 lbf to 10 lbf0.025 lbf
1 lbf to 600000 lbf0.25 % of reading over the Class A loading range of the standard

Calibration of force testing machines – Compression

ASTM E4. 300 lbf, 2 klbf, 10 klbf, 60 klbf, 270 klbf, and 600 klbf standards.
In place of the pound-force (lbf) unit, calibration or verification may be performed using the SI unit, the Newton (N). 1 lbf is equivalent to 4.448222 N. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.06 lbf to 10 lbf0.025 lbf
1 lbf to 600000 lbf0.25 % of reading over the Class A loading range of the standard

Mass

Type II Capability

Balances

Class of calibration weights: ANSI/ASTM Class 1. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
1 mg, 2 mg, 5 mg, 10 mg, 20 mg, 50 mg, 100 mg, 200 mg, 500 mg0.01 mg
1 g, 2 g, 5 g0.034 mg
10 g0.05 mg
20 g0.074 mg
50 g0.12 mg
100 g0.25 mg
200 g0.5 mg
500 g1.2 mg
1 kg2.2 mg
2 kg5 mg
5 kg12 mg
10 kg25 mg

NIST Class F. On-site calibration available.

Measured Quantity & Range or InstrumentExpanded Measurement Uncertainty
0.05 oz1 mg
0.1 oz1.3 mg
0.2 oz1.6 mg
0.5 oz2.8 mg
1 oz5.4 mg
2 oz11 mg
4 oz23 mg
8 oz45 mg
1 lb70 mg
2 lb91 mg
5 lb227 mg
10 lb450 mg
50 lb2300 mg
100 lb4500 mg
500 lb23000 mg

Hardness

Type III Capability

Measured Quantity & Range or InstrumentExpanded Measurement UncertaintyRemarks
Indirect verification of Brinell hardness testersGoverned by the uncertainty of the standardized test block used to perform the indirect verificationASTM E10, ITC Procedure CAL.049, Standardized Test Blocks per ASTM E10
On-site calibration available.
Indirect verification of Rockwell hardness testersGoverned by the uncertainty of the standardized test block used to perform the indirect verificationASTM E18, ITC Procedure CAL.050, Standardized Test Blocks per ASTM E18
On-site calibration available.
Indirect verification of Rockwell Superficial Hardness TestersGoverned by the uncertainty of the standardized test block used to perform the indirect verificationASTM E18, ITC Procedure CAL.050, Standardized Test Blocks per ASTM E18
On-site calibration available.
Indirect verification of Knoop hardness testersGoverned by the uncertainty of the standardized test block used to perform the indirect verificationASTM E384, ITC Procedure CAL.051, Standardized Test Blocks per ASTM E384
On-site calibration available.
Indirect verification of Vickers Macro hardness testersGoverned by the uncertainty of the standardized test block used to perform the indirect verificationASTM E92, ITC Procedure CAL.051, Standardized Test Blocks per ASTM E384
On-site calibration available.
Indirect verification of Vickers Micro hardness testersGoverned by the uncertainty of the standardized test block used to perform the indirect verificationASTM E384, ITC Procedure CAL.051, Standardized Test Blocks per ASTM E384
On-site calibration available.
Indirect verification of Portable hardness testersGoverned by the uncertainty of the standardized test block used to perform the indirect verificationASTM E110, ITC Procedure CAL.052, Standardized Test Blocks per ASTM E110
On-site calibration available.