CLAS Certificate Number 2023-01

Company name Colour and UV Service
Company address 1410 Dupont Street, Suite 2001
Toronto, ON  M6H 0B6
Canada
Contact Korey Louks
Telephone: (647) 838-4491
Email: koreylouks@colouruvservice.ca
Clients served
  • All interested parties
  • Some calibration services are available on-site. These services are indicated in the "Remarks" column of the following pages.
Fields of calibration
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 151298
  • First issued 2023-03-08
  • Issue 2.0e 2024-09-09

This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body. The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS. See supplementary notes.

Photometry and Radiometry

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Type of service Remarks
Correlated Colour Temperature
2856 K 32 K II For the calibration of colour light booths using a spectroradiometer.
On-site calibration available.
Illuminance
20 lx to 100 000 lx 7.6 % II For the calibration of colour light booths using an illuminance meter.
On-site calibration available.
Spectrophotometer
L*: 0 L*a*b* Units to 100 L*a*b* Units
a*b*: -128 L*a*b* Units to +127 L*a*b* Units
0.45 L*a*b* Units II For the calibration of a Hunterlab, X-Rite or Datacolor D/8° spectrophotometer, as well as any Sphere Spectrophotometer with D/8° optical geometry, using a set of 6 BCRA coloured tiles in the 400 nm – 700 nm wavelength range.
On-site calibration available.
L*: 0 L*a*b* Units to 100 L*a*b* Units
a*b*: -128 L*a*b* Units to +127 L*a*b* Units
0.56 L*a*b* Units II For the calibration of a Hunterlab, X-Rite or Datacolor 0/45° spectrophotometer, as well as any Spectrophotometer with 0/45° optical geometry, using a set of 6 BCRA coloured tiles in the 400 nm – 700 nm wavelength range.
On-site calibration available.
Gloss Meter
0 Gloss Units to 100 Gloss Units 0.63 Gloss Units II

For the calibration of gloss meters using a set of Semi Gloss checking standards. Semi Gloss: 20 degrees, and 60 degrees.
On-site calibration available
.

0 Gloss Units to 100 Gloss Units 0.65 Gloss Units II

For the calibration of gloss meters using a set of Semi Gloss checking standards.
Semi Gloss: 85 degrees.

On-site calibration available.
0 Gloss Units to 100 Gloss Units 0.46 Gloss Units II For the calibration of gloss meters using a set of High Gloss checking standards.
High Gloss: 20 degrees, 60 degrees, and 85 degrees.
On-site calibration available.