Company name | DegreeTek Services Inc. |
---|---|
Company address | 210 Davison Crescent Pakenham, Ontario K0A 2X0 Canada |
Contact | David Smith Telephone: (613) 979-5389 Emails: |
Clients served |
|
Fields of calibration | |
Accreditation body |
|
ISO/IEC 17025 |
|
This CLAS Certificate is published by the CLAS program of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to the CLAS program. See supplementary notes.
Electrical
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type of service | Remarks |
---|---|---|---|
Voltage | |||
0 mV to 110 mV | 0.017 mV | II |
Source. For the calibration of dc voltage measuring devices using a process calibrator. On-site calibration available. |
110 mV to 1.1 V | 0.17 mV | ||
1.1 V to 15 V | 2.3 mV | ||
0 V to 110 mV | 0.045 mV |
Measure. For the calibration of dc voltage generating devices using a process calibrator. On site calibration available. . |
|
110 mV to 1.1 V | 0.34 mV | ||
1.1 V to 11 V | 3.4 mV | ||
11 V to 110 V | 61 mV | ||
110 V to 300 V | 0.17 V |
Electrical calibration of temperature indicators
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type of service | Remarks |
---|---|---|---|
Thermocouple simulation | |||
Type J | |||
-210 °C to -100 °C | 0.51 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
-100 °C to 1200 °C | 0.41 °C | ||
Type K | |||
-200 °C to -100 °C | 0.61 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
-100 °C to 1372 °C | 0.51 °C | ||
Type N | |||
-200 °C to -100 °C | 0.81 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
-100 °C to 900 °C | 0.71 °C | ||
900 °C to 1300 °C | 0.51 °C | ||
Type R | |||
-20 °C to 0 °C | 1.4 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 100 °C | 1.3 °C | ||
100 °C to 1767 °C | 1.1 °C | ||
Type S | |||
-20 °C to 0 °C | 1.4 °C |
II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 200 °C | 1.3 °C | ||
200 °C to 1400 °C | 1.1 °C | ||
1400 °C to 1767 °C | 1.2 °C | ||
Type T | |||
-250 °C to -200 °C | 1.1 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
-200 °C to 0 °C | 0.61 °C | ||
0 °C to 400 °C | 0.51 °C | ||
RTD simulation | |||
Pt 385, 100 Ω | |||
-200 °C to 0 °C | 0.11 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 400 °C | 0.21 °C | ||
400 °C to 800 °C | 0.41 °C | ||
Pt 385, 200 Ω | |||
-200 °C to 0 °C | 0.11 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 400 °C | 0.21 °C | ||
400 °C to 630 °C | 0.41 °C | ||
Pt 385, 500 Ω | |||
-200 °C to 0 °C | 0.11 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 400 °C | 0.21 °C | ||
400 °C to 630 °C | 0.41 °C | ||
Pt 385, 1000 Ω | |||
-200 °C to 0 °C | 0.11 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 400 °C | 0.21 °C | ||
400 °C to 630 °C | 0.41 °C | ||
Pt 3916, 100 Ω | |||
-200 °C to -190 °C | 0.31 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
-190 °C to 0 °C | 0.37 °C | ||
0 °C to 630 °C | 0.44 °C | ||
Pt 3926, 100 Ω | |||
-200 °C to 0 °C | 0.11 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 630 °C | 0.21 °C | ||
Cu 427, 10 Ω | |||
-100 °C to 0 °C | 1.0 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
0 °C to 260 °C | 1.1 °C | ||
Ni 672, 120 Ω | |||
-200 °C to 260 °C | 0.11 °C | II |
Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator. On-site calibration available. |
Pressure
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type of service | Remarks |
---|---|---|---|
0 psi to 5 psi | 0.0025 psi | II
|
Pneumatic. Gauge pressure. For the calibration of pressure measuring devices by comparison to a reference pressure transducer. On-site calibration available. |
5 psi to 200 psi | 0.15 psi | ||
200 psi to 5000 psi | 4.0 psi | II |
Hydraulic. Gauge pressure. For the calibration of pressure measuring devices by comparison to a reference pressure transducer. On-site calibration available. |
Thermometry
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type of service | Remarks |
---|---|---|---|
Chambers and Ovens | |||
-85 °C to 0 °C | 1.5 °C | II |
Measure Temperature using type T thermocouple. On-site calibration available. |
0 °C to 200 °C | 1.2 °C | ||
200 °C to 600 °C | 1.7 °C |
Measure Temperature using type N thermocouple. On-site calibration available. |
|
600 °C to 1100 °C | 7.6 °C | ||
-200 °C to 0 °C | 1.2 °C |
Measure Temperature using 100 Ω 385 RTD. On-site calibration available. |
|
0 °C to 400 °C | 1.0 °C |