CLAS Certificate Number 2022-01

Company name DegreeTek Services Inc.
Company address 210 Davison Crescent
Pakenham, Ontario K0A 2X0
Canada
Contact David Smith
Telephone: (613) 979-5389
Emails:
Clients served
  • All interested parties
  • Some calibration services are available on-site. These services are indicated in the "Remarks" column of the following pages.
Fields of calibration
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 151286
  • First issued
  • Issue 1.0

This CLAS Certificate is published by the CLAS program of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to the CLAS program. See supplementary notes.

Electrical

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Type of service Remarks
Voltage
0 mV to 110 mV 0.017 mV II

Source. For the calibration of dc voltage measuring devices using a process calibrator.

On-site calibration available.

110 mV to 1.1 V 0.17 mV
1.1 V to 15 V 2.3 mV
0 V to 110 mV 0.045 mV

Measure. For the calibration of dc voltage generating devices using a process calibrator.

On site calibration available.

.
110 mV to 1.1 V 0.34 mV
1.1 V to  11 V 3.4 mV
11 V to  110 V 61 mV
110 V to  300 V 0.17 V

Electrical calibration of temperature indicators

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Type of service Remarks
Thermocouple simulation
Type J
-210 °C to -100 °C 0.51 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

-100 °C to 1200 °C 0.41 °C
Type K
-200 °C to -100 °C 0.61 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

-100 °C to 1372 °C 0.51 °C
Type N
-200 °C to -100 °C 0.81 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

-100 °C to 900 °C 0.71 °C
900 °C to 1300 °C 0.51 °C
Type R
-20 °C to 0 °C 1.4 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

0 °C to 100 °C 1.3 °C
100 °C to 1767 °C 1.1 °C
Type S
-20 °C to 0 °C 1.4 °C

II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

0 °C to 200 °C 1.3 °C
200 °C to 1400 °C 1.1 °C
1400 °C to 1767 °C 1.2 °C
Type T
-250 °C to -200 °C 1.1 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

-200 °C to 0 °C 0.61 °C
0 °C to 400 °C 0.51 °C
RTD simulation
Pt 385, 100 Ω
-200 °C to 0 °C 0.11 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

0 °C to 400 °C 0.21 °C
400 °C to 800 °C 0.41 °C
Pt 385, 200 Ω
-200 °C to 0 °C 0.11 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

0 °C to 400 °C 0.21 °C
400 °C to 630 °C 0.41 °C
Pt 385, 500 Ω
-200 °C to 0 °C 0.11 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

0 °C to 400 °C 0.21 °C
400 °C to 630 °C 0.41 °C
Pt 385, 1000 Ω
-200 °C to 0 °C 0.11 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

0 °C to 400 °C 0.21 °C
400 °C to 630 °C 0.41 °C
Pt 3916, 100 Ω
-200 °C to -190 °C 0.31 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

-190 °C to 0 °C 0.37 °C
0 °C to 630 °C 0.44 °C
Pt 3926, 100 Ω
-200 °C to 0 °C 0.11 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

0 °C to 630 °C 0.21 °C
Cu 427, 10 Ω
-100 °C to 0 °C 1.0 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.
0 °C to 260 °C 1.1 °C
Ni 672, 120 Ω
-200 °C to 260 °C 0.11 °C II

Source. For the calibration of temperature indicators by electrical simulation of temperature using a process calibrator.

On-site calibration available.

Pressure

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Type of service Remarks
0 psi to 5 psi 0.0025 psi II

 

Pneumatic.

Gauge pressure. For the calibration of pressure measuring devices by comparison to a reference pressure transducer.

On-site calibration available.

5 psi to 200 psi 0.15 psi
200 psi to 5000 psi 4.0 psi II

Hydraulic.

Gauge pressure. For the calibration of pressure measuring devices by comparison to a reference pressure transducer.

On-site calibration available.

Thermometry

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Type of service Remarks
Chambers and Ovens
-85 °C to 0 °C 1.5 °C II

Measure Temperature using type T thermocouple.

On-site calibration available.

0 °C to 200 °C 1.2 °C
200 °C to 600 °C 1.7 °C

Measure Temperature using type N thermocouple.

On-site calibration available.

600 °C to 1100 °C 7.6 °C
-200 °C to 0 °C 1.2 °C

Measure Temperature using 100 Ω 385 RTD.

On-site calibration available.

0 °C to 400 °C 1.0 °C