CLAS Certificate Number 2019-01

Company name SEP Métrologie
Company address 617 rue Simonds Sud
Granby, Quebec, J2J 1C1
Canada
Contact Érick Rainville
Telephone: (450) 776-3286
Email: erainville@sepmetrologie.com
Clients served
  • All interested parties
  • Some calibration services are available on-site. These services are indicated in the "Remarks" column of the following pages.
Fields of calibration
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 151099
  • First issued
  • Issue 3.2e

This CLAS Certificate is published by the CLAS program of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to the CLAS program. See supplementary notes.

Dimensional

Type I Capability

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Remarks
Gauge block, length: Steel, Rectangular and Square
Inch, up to 4 inches (3.19 + 7.34L) µinch
(Note: L in 'inches')
For the calibration of gauge block. SEP Métrologie calibration procedure IE‑004, ANSI/ASME B89.1.9.
Metric, up to 100 millimeters (0.08 + 0.007L) µm
(Note: L in 'millimeters')
Cylindrical ring gauge, diameter
Inch, 0.04 to 1 inches 27 µinch For the calibration of cylindrical ring gauge. SEP Métrologie calibration procedure IE‑008, 33k6-4-2-1 (2017)
Inch, >1 to 4 inches 44 µinch
Metric, 0 to 25 millimeters 0.7 µm
Metric, >25 to 100 millimeters 1.1 µm
Cylindrical plug gauge, diameter
Inch, up to 4 inches (10.71 + 5.82D) µinch
(Note: D in 'inches')
For the calibration of cylindrical plug gauge. SEP Métrologie calibration procedure IE‑004, ANSI/ASME B89.1.5.
Metric, up to 100 millimeters (0.27 + 0.006D) µm
(Note: D in 'millimeters')

Type II Capability

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Remarks
60 degree thread plug gauge and adjustable thread plug gauge: Pitch diameter (measured over thread wires)
Inch, up to 4 inches 180 µinch For the calibration of thread and adjustable plug gauge. SEP Métrologie calibration procedure IE‑009, ANSI/ASME B89.1.17 and ANSI/ASME B1.2.1983
Metric, up to 100 millimeters 4.6 µm
60 degree thread plug gauge and adjustable thread plug gauge: Major diameter
Inch, up to 4 inches (3.19 + 7.34L) µinch
(Note: L in 'inches')
For the calibration of thread and adjustable thread plug gauge. SEP Métrologie calibration procedure IE‑009, ANSI/ASME B89.1.17 and ANSI/ASME B1.2.1983
Metric, up to 100 millimeters (0.08 + 0.007L) µm
(Note: L in 'millimeters')
Indicators, Analog:
Inch, 0.001 inch graduation 120 µinch For the calibration of indicators. SEP Métrologie calibration procedure IE‑006 and IE‑007, ANSI/ASME B89.1.10M.
Inch, 0.0005 inch graduation 61 µinch
Inch, 0.0001 inch graduation 23 µinch
Metric, 0.01 mm graduation 3.1 µm
Metric, 0.002 mm graduation 1.6 µm
Metric, 0.001 mm graduation 0.6 µm
Inch, 0.001 inch graduation 240 µinch

For the calibration of indicators. SEP Métrologie calibration procedure IE‑006 and IE‑007, ANSI/ASME B89.1.10M.

On-site calibration available.

Inch, 0.0005 inch graduation 120 µinch
Inch, 0.0001 inch graduation 40 µinch
Metric, 0.01 mm graduation 6.1 µm
Metric, 0.002 mm graduation 3.1 µm
Metric, 0.001 mm graduation 1.0 µm
Indicators, Digital:
Inch, 0.00005 inch resolution 35 µinch

For the calibration of indicators. SEP Métrologie calibration procedure IE 006 and IE 007, ANSI/ASME B89.1.10M.

Inch, 0.0005 inch resolution 290 µinch
Metric, 0.001 mm resolution 0.9 µm
Metric, 0.01 mm resolution 7.4 µm
Inch, 0.00005 inch resolution 67 µinch

For the calibration of indicators. SEP Métrologie calibration procedure IE‑006 and IE‑007, ANSI/ASME B89.1.10M.

On-site calibration available.

Inch, 0.0005 inch resolution 580 µinch
Metric, 0.001 mm resolution 1.7 µm
Metric, 0.01 mm resolution 14.7 µm
Height Gauges
Inch, up to 24 inches (52.87 + 6.13L) µinch
(Note: L in 'inches')
For the calibration of height gauges. SEP Métrologie calibration procedure IE‑013.
Metric, up to 600 millimeters (1.34 + 0.006L) µm
(Note: L in 'millimeters')
Micrometers: Outside
Inch, up to 12 inches (48.04 + 4.73L) µinch
(Note: L in 'inches')
For the calibration of outside micrometers. SEP Métrologie calibration procedure IE‑001, ANSI/ASME B89.1.13
Metric, up to 300 millimeters (1.22+ 0.005L) µm
(Note: L in 'millimeters')
Inch, up to 12 inches (48.04 + 4.73L) µinch
(Note: L in 'inches')

For the calibration of outside micrometers.  SEP Métrologie calibration procedure IE‑001, ANSI/ASME B89.1.13

On-site calibration available.

Metric, up to 300 millimeters (1.22+ 0.005L) µm
(Note: L in 'millimeters')
Micrometer: Depth
Inch, up to 12 inches 590 µinch For the calibration of depth micrometers. SEP Métrologie calibration procedure IE‑003, ANSI/ASME B89.1.13.
Metric, up to 300 millimeters 15 µm
Inch, up to 12 inches 590 µinch

For the calibration of depth micrometers, SEP Métrologie calibration procedure IE‑003, ANSI/ASME B89.1.13.

On-site calibration available.

Metric, up to 300 millimeters 15 µm
Micrometer setting standards
Inch, up to 12 inches (2.08 + 8.12L) µinch
(Note: L in 'inches')
For the calibration of micrometer setting standard. SEP Métrologie calibration procedure IE‑014, 33k6-4-369-1
Metric, up to 300 millimeters (0.05 + 0.008L) µm
(Note: L in 'millimeters')
Caliper: Outside
Inch, up to 12 inches 300 µinch For the calibration of outside calipers. SEP Métrologie calibration procedure IE‑012, ANSI/ASME B89.1.14
Metric, up to 300 millimeters 7.6 µm
Inch, up to 12 inches 300 µinch

For the calibration of outside calipers. SEP Métrologie calibration procedure IE‑012, ANSI/ASME B89.1.14.

On-site calibration available.

Metric, up to 300 millimeters 7.6 µm
Caliper: Inside
Inch, up to 12 inches 290 µinch For the calibration of inside calipers. SEP Métrologie calibration procedure IE 012, ANSI/ASME B89.1.14
Metric, up to 300 millimeters 7.4 µm
Inch, up to 12 inches 290 µinch

For the calibration of inside calipers. SEP Métrologie calibration procedure IE‑012, ANSI/ASME B89.1.14

On-site calibration available.

Metric, up to 300 millimeters 7.4 µm
Caliper: Depth
Inch, up to 12 inches 290 µinch For the calibration of depth calipers. SEP Métrologie calibration procedure IE‑012, ANSI/ASME B89.1.14
Metric, up to 300 millimeters 7.4 µm
Inch, up to 12 inches 290 µinch

For the calibration of depth calipers. SEP Métrologie calibration procedure IE‑012, ANSI/ASME B89.1.14.

On-site calibration available.

Metric, up to 300 millimeters 7.4 µm
Caliper: Step
Inch, up to 12 inches 290 µinch For the calibration of step calipers. SEP Métrologie calibration procedure IE‑012, ANSI/ASME B89.1.14
Metric, up to 300 millimeters 7.4 µm
Inch, up to 12 inches 290 µinch

For the calibration of step calipers. SEP Métrologie calibration procedure IE‑012, ANSI/ASME B89.1.14.

On-site calibration available.

Metric, up to 300 millimeters 7.4 µm
Roundness
Metric, up to 50 millimeters (0.083 + 0.001R) µm
(Note: R is the roundness in µm)
For the calibration of roundness. SEP Métrologie calibration procedure IE‑011.
Diameter
Metric, up to 50 millimeters 0.91 µm For the calibration of diameter. SEP Métrologie calibration procedure IE‑010, 33k6-4-1181-1