CLAS Certificate Number 2016-04

Company name Mesurepro Inc.
Company address 106-86, Boulevard des Entreprises
Boisbriand, QC
Canada J7G 2T3
Contact Karim Bouyahia, President
Telephone: 514-913-0820
Fax: 514-951-0675
Email: info@mesurepro.ca
Clients served
  • All interested parties
  • Calibration services are available on-site.
Fields of calibration
  • Thermometry
  • Humidity
  • Electrical
  • Time and Frequency
  • Pressure
  • Force
  • Mass
  • Volume
  • Dimensional
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • Accredited Laboratory Nº 832
  • First issued
  • Issue 6.0e

This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body. The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS. See supplementary notes.

Thermometry

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Thermistor and Resistance Temperature Detector (RTD)
-196 ºC 0.02 ºC II For the calibration of temperature measurement systems thermistor and resistance temperature detector (RTD). All probes are calibrated in comparison with a PRT in a thermometry dryblock. Mesurepro procedure PC 5.4.61. On‑site calibrations available.
-95 ºC to 140 ºC 0.03 ºC
>140 ºC to 660 ºC 0.09 ºC
>660 ºC to 1200 ºC 2.1 ºC
Thermocouple
‑196 ºC 1.1 ºC + 0.5 % of reading II For the calibration of temperature measurement systems thermocouple. All probes are calibrated in comparison with a PRT in a thermometry dryblock. Mesurepro procedure PC 5.4.62. On‑site calibrations available.
-95 ºC to 1200 ºC 1.1 ºC + 0.5 % of reading
Liquid in glass thermometers
‑95 ºC to 140 ºC 0.07 ºC II For the calibration of liquid in glass thermometers. All thermometers are calibrated in comparison with a PRT in a thermometry dryblock. Mesurepro procedure PC 5.4.60. On‑site calibrations available.
>140 ºC to 300 ºC 0.12 ºC
Radiation Thermometry
Infrared radiation thermometers
35 ºC to 400 ºC 2.5 ºC II For the calibration of infrared thermometers at emissivity setting of 0.95 with spectral band of 8 to 14 µm. Mesurepro procedure PC 5.4.170. On‑site calibrations available.
Temperature Controlled Chambers
‑55 ºC to ‑40 ºC 4 ºC II For the calibration of temperature controlled chambers. All chambers are calibrated by characterisation in temperature of the volume used. Mesurepro procedure PC 7.2.30. On‑site calibrations available.
>‑40 ºC to 140 ºC 0.6 ºC
>140 ºC to 600 ºC 4 ºC
Temperature and Pressure Controlled Chambers ("Vacuum oven")
0 ºC to 140 ºC 0.6 ºC II For the calibration of temperature controlled chambers ("vacuum oven"). All chambers are calibrated by characterisation in temperature of the volume used. Mesurepro procedure PC 7.2.30‑1. On‑site calibrations available.
‑14.5 psi 1 psi For the calibration of pressure controlled chambers ("vacuum oven"). All chambers are calibrated by characterisation of pressure measurement. Mesurepro procedure PC 7.2.30‑1. On‑site calibrations available.
Thermal bath
‑100 ºC to 660 ºC 0.2 ºC II For the calibration of thermal bath. All baths are calibrated in comparison with a PRT. Mesurepro procedure PC 7.8.63.

Humidity

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Relative Humidity
15 °C to 60 °C 20 % to 90 % RH 1.8 % RH II Source. For the calibration of relative humidity and temperature measuring instruments. Mesurepro procedure PC 5.4.110.

Electrical

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an uncertainty (±) (See supplementary notes) Type of Service Remarks
Voltage, DC
0.0000 mV to 10.0000 mV 0.050 % + 7 μV II Source. For the calibration of dc voltage measuring devices. On‑site calibration is available.
10.000 mV to 100.000 mV 0.010 % + 7 μV
0.10000 V to 1.00000 V 0.006 % + 10 μV
1.0000 V to 10.0000 V 0.006 % + 50 μV
10.000 V to 100.000 V 0.006 % + 1000 μV
100.00 V to 1000.00 V 0.010 % + 20000 μV
Current, DC
0.000 μA to 200.00 μA 0.050 % + 0.02 μA II Source. For the calibration of dc current measuring devices. On‑site calibration is available.
0.20000 mA to 2.00000 mA 0.025 % + 0.1 μA
2.0000 mA to 20.0000 mA 0.015 % + 0.6 μA
20.000  mA to 200.000 mA 0.015 % + 6 μA
0.2000 A to 2.0000 A 0.015 % + 100 μA
2.000 A to 20.0000 A 0.1 % + 2 mA
Resistance
10 Ω 0.05 % + 50 mΩ II Source. For the calibration of resistance measuring devices. On‑site calibration is available.
100 Ω 0.05 %
1 kΩ 0.02 %
10 kΩ 0.02 %
100 kΩ 0.02 %
1 MΩ 0.05 %
10 MΩ 0.05 %
100 MΩ 0.5 %
Measured Quantity & Range or Instrument Frequency Calibration Measurement Capability expressed as an uncertainty (±) (See supplementary notes) Type of Service Remarks
Voltage, AC
1.0000 mV to 10.0000 mV 20.000 Hz to 400.000 Hz 0.20 % + 0.025 mV II Source. For the calibration of sinusoidal wave voltage measuring devices. Note: Voltage ranges 100 V and 1000 V from 40 Hz to 1kHz. On‑site calibration is available.
400.000 Hz to 10.00 kHz 0.2 % + 0.03 mV
10.000 mV to 100.000 mV 20.000 Hz to 400.000 Hz 0.10 % + 0.05 mV
400.000 Hz to 10.00 kHz 0.15 % + 0.07 mV
0.10000 V to 1.00000 V 20.000 Hz to 400.000 Hz 0.05 % + 0.05 mV
400.000 Hz to 10.00 kHz 0.07 % + 0.1 mV
1.0000 V to 10.0000 V 20.000 Hz to 400.000 Hz 0.05 % + 0.5 mV
400.000 Hz to 10.00 kHz 0.07 % + 3 mV
10.000 V to 100.000 V 20.000 Hz to 400.000 Hz 0.05 % + 10 mV
400.000 Hz to 10.00 kHz 0.07 % + 30 mV
10.00 V to 1000.00 V 20.000 Hz to 400.000 Hz 0.07 % + 0.2 V
400.000 Hz to 10.00 kHz 0.10 % + 0.3 V
Current, AC
1.000 μA to 200.000 μA 20.000 Hz to 200.000 Hz 0.25 % + 0.02 μA II Source. For the calibration of sinusoidal wave current measuring devices. On‑site calibration is available.
200.000 Hz to 1000.00 Hz 0.20 % + 0.2 μA
0.20000 mA to 2.00000 mA 20.000 Hz to 200.000 Hz 0.10 % + 0.2 μA
200.000 Hz to 1000.00 Hz 0.10 % + 0.4 μA
2.0000 mA to 20.0000 mA 20.000 Hz to 200.000 Hz 0.07 % + 1 μA
200.000 Hz to 1000.00 Hz 0.10 % + 4 μA
20.000 mA to 200.000 mA 20.000 Hz to 200.000 Hz 0.07 % + 10 μA
200.000 Hz to 1000.00 Hz 0.10 % + 40 μA
0.2000 A to 2.0000 A 20.000 Hz to 200.000 Hz 0.10 % + 0.1 mA
200.000 Hz to 1000.00 Hz 0.15 % + 1 mA
2.0000 A to 20.000 A 20.000 Hz to 200.000 Hz 0.20 % + 3 mA
200.000 Hz to 1000.00 Hz 0.25 % + 10 mA
Current clamp calibration
Effective AC current output
1 A turns to 1000 A Turns 20.000 Hz to 200.000 Hz 0.42 % + 4.2 mA II Source using a multifunction calibrator and a 50 turn coil. For the calibration of clamp meters. On‑site calibration is available.
Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Type of Service Remarks
Current clamp calibration
Effective DC current output
1 A Turns to 1000 A Turns 0.37 % + 3.3 mA II Source using a multifunction calibrator and a 50 turn coil. For the calibration of clamp meters. On-site calibration is available.
Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Electrical Calibration of Temperature Indicators and Simulators
Thermocouple simulation
Type C
0 ºC to 2316 ºC 1.6 ºC II Source for type C. For the calibration of temperature indicators and process calibrators by electrical simulation of temperature. Mesurepro procedure PC 5.4.70. On‑site calibration available.
Type J
‑210 ºC to 1200 ºC 0.5 ºC II Source for type J. For the calibration of temperature indicators and process calibrators by electrical simulation of temperature. Mesurepro procedure PC 5.4.70. On‑site calibration available.
Type K
‑210 ºC to 1200 ºC 0.6 ºC II Source for type K. For the calibration of temperature indicators and process calibrators by electrical simulation of temperature. Mesurepro procedure PC 5.4.70. On‑site calibration available.
Type N
‑200 ºC to 1300 ºC 0.8 ºC II Source for type N. For the calibration of temperature indicators and process calibrators by electrical simulation of temperature. Mesurepro procedure PC 5.4.70. On‑site calibration available.

Time and Frequency

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an uncertainty (±) (See supplementary notes) Type of Service Remarks
Frequency
0.1000 Hz to 2.00000 MHz 0.005 % II Source. For the calibration of frequency measuring devices. On‑site calibration is available.

Pressure

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Pneumatic
‑12 psi to 0 psi 0.1 % FS II Vacuum gauge. Mesurepro procedure PC 5.4.50. Mobile and On‑site calibration available.
0 psi to 300 psi 0.1 % FS Gauge. Mesurepro procedure PC 5.4.50. Mobile and On‑site calibration available.
Hydraulic
‑300 psi to 16000 psi 0.1 % FS II Gauge. Mesurepro procedure PC 5.4.50.

Force

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Torque
Clockwise and Counterclockwise
0.73 ft•lb to 750 ft•lb 1 % of reading II For the calibration of torque wrenches and screwdrivers. Mesurepro procedure PC 7.2.140.
0.73 ft•lb to 750 ft•lb 3 % of reading For the calibration of torque wrenches and screwdrivers. Mesurepro procedure PC 7.2.140. Mobile and On‑site calibration available.

Mass

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Balances (Electronic, Non Automatic)
80 g 0.13 mg II For the calibration of electronic and non automatic balance. The calibration measurement capability is based upon the use of OIML R111 Class E2 reference weights. On‑site calibration available.
220 g 0.22 mg
1 kg 6 mg
6 kg 27 mg
10 kg 174 mg
30 kg 1459 mg

Volume

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Piston Pipette
0.2 µl 0.2 µl II For the calibration of single channel and multi‑channel pipettes as per ISO 8655‑7. On‑site calibration available.
1 µl 0.2 µl
2 µl 0.2 µl
5 µl 0.2 µl
10 µl 0.46 µl
20 µl 0.46 µl
50 µl 0.46 µl
100 µl 4.7 µl
200 µl 4.7 µl
250 µl 4.7 µl
500 µl 11 µl
1000 µl 23 µl
1250 µl 23 µl
2500 µl 23 µl
5000 µl 23 µl

Dimensional

Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) Type of Service Remarks
Indicators, Analog:
Inch, 0.00005 inch graduation 39 µinch II For the calibration of indicators. Mesurepro calibration procedure PC 7.2.20‑5. On‑site calibration available.
Inch, 0.0001 inch graduation 63 µinch II For the calibration of indicators. Mesurepro calibration procedure PC 7.2.20‑5. On‑site calibration available.
Inch, 0.0005 inch graduation 290 µinch II For the calibration of indicators. Mesurepro calibration procedure PC 7.2.20‑5. On‑site calibration available.
Inch, 0.001 inch graduation 578 µinch II For the calibration of indicators. Mesurepro calibration procedure PC 7.2.20‑5. On‑site calibration available.
Metric, 0.001 mm graduation 0.8 µm II For the calibration of indicators. Mesurepro calibration procedure PC 7.2.20‑5. On‑site calibration available.
Metric, 0.002 mm graduation 1.3 µm II For the calibration of indicators. Mesurepro calibration procedure PC 7.2.20‑5. On‑site calibration available.
Metric, 0.01 mm graduation 5.8 µm II For the calibration of indicators. Mesurepro calibration procedure PC 7.2.20‑5. On‑site calibration available.
Indicators, Digital:
Inch, 0.00005 inch resolution 39 µinch II For the calibration of indicators. Mesurepro
calibration procedure PC 7.2.20-5.
On-site calibration available.
Inch, 0.0001 inch resolution 63 µinch II For the calibration of indicators. Mesurepro
calibration procedure PC 7.2.20-5.
On-site calibration available.
Inch, 0.0005 inch resolution 290 µinch II For the calibration of indicators. Mesurepro
calibration procedure PC 7.2.20-5.
On-site calibration available.
Inch, 0.001 inch resolution 578 µinch II For the calibration of indicators. Mesurepro
calibration procedure PC 7.2.20-5.
On-site calibration available.
Metric, 0.001 mm resolution 0.8 µm II For the calibration of indicators. Mesurepro
calibration procedure PC 7.2.20-5.
On-site calibration available.
Metric, 0.002 mm resolution 1.3 µm II For the calibration of indicators. Mesurepro
calibration procedure PC 7.2.20-5.
On-site calibration available.
Metric, 0.01 mm resolution 5.8 µm II For the calibration of indicators. Mesurepro
calibration procedure PC 7.2.20-5.
On-site calibration available.
Bore Gages
Inch, up to 8 inches 109 µinch II For the calibration of bore gages. Mesurepro calibration procedure PC 7.2.20‑3. On‑site calibration available.
Metric, up to 300 mm 3.5 µm II For the calibration of bore gages. Mesurepro calibration procedure PC 7.2.20‑3. On‑site calibration available.
Calipers: Outside
Inch, up to 12 inches 324 µinch II For the calibration of outside calipers. Mesurepro calibration procedure PC 7.2.20‑1. On‑site calibration available.
>12 inches to 24 inches 405 µinch
Metric, up to 300 mm 7 µm II For the calibration of outside calipers. Mesurepro calibration procedure PC 7.2.20‑1. On‑site calibration available.
>300 mm to 600 mm 9 µm
Micrometers: Outside
Inch, up to 8 inches 97 µinch II For the calibration of outside micrometers. Mesurepro calibration procedure PC 7.2.20‑2. On‑site calibration available.
Metric, up to 300 mm 3.5 µm II For the calibration of outside micrometers. Mesurepro calibration procedure PC 7.2.20‑2. On‑site calibration available.
Height Gauges
Inch, up to 24 inches 530 µinch II For the calibration of height gauges. Mesurepro calibration procedure PC 7.2.20‑4.
Metric, up to 600 mm 13 µm II For the calibration of height gauges. Mesurepro calibration procedure PC 7.2.20‑4.