CLAS Certificate Number 2016-02

 
Company name Transcat Canada - Ottawa
Company address 4043 Carling Avenue
Ottawa, Ontario
Canada K2K 2A4
Contact Francis Kane
Tel: 613-591-8140
Fax: 613-591-6318
Email:Francis.Kane@transcat.ca
Clients served
  • All interested parties
  • Some calibration services are available on-site. These services are indicated in the "Remarks" column of the following pages.
Fields of calibration
SCC accreditation
(ISO/IEC 17025)
  • Accredited Laboratory Nº 827
  • First issued
  • Issue 2.0e

This scope of calibration capabilities is published by the CLAS program of the National Research Council of Canada (NRC) in close co-operation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform the listed calibrations at the given Calibration Measurement Capability (see Supplementary note C and note D) with traceability to the International System of Units (SI) or to standards acceptable to the CLAS program.

 

Type II Capability

Dimensional

Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Micrometers: Outside, Inside, Depth
0.0 to 0.05 inches 7.5 µinch

Comparison to gauge blocks.  Transcat calibration procedure. On-site calibration available.

0.05 to 0.45 inches ( 7 + 7L ) µinch (Note: L in 'inches')
0.45 to 1.0 inch ( 5 + 12L ) µinch (Note: L in 'inches')
1.0 to 4.0 inches ( 2 + 15L ) µinch (Note: L in 'inches')
4.0 to 40 inches ( 9 + 16L ) µinch (Note: L in 'inches')
Anvil Flatness:
0 to 1 inch diameter 6.3 µinch Transcat calibration procedure. On-site calibration available.
Calipers: Outside, Inside, Depth
0.0 to 0.05 inches 7.5 µinch

Comparison to gauge blocks with surface plate and ring gauge.  Transcat calibration procedure. On-site calibration available.

0.05 to 0.45 inches ( 7 + 7L ) µinch (Note: L in 'inches')
0.45 to 1.0 inch ( 5 + 12L ) µinch (Note: L in 'inches')
1.0 to 4.0 inches ( 2 + 15L ) µinch (Note: L in 'inches')
4.0 to 40 inches ( 9 + 16L ) µinch (Note: L in 'inches')
Dial Indicators:
0.0 to 0.45 inches ( 18 + 3L ) µinch (Note: L in 'inches')

Comparison to Dial Indicator Tester.  Transcat calibration procedure

0.45 to 1.0 inch ( 15 + 8L ) µinch (Note: L in 'inches')
1.0 to 4.0 inches ( 9 + 14L ) µinch (Note: L in 'inches')
4.0 to 6 inches ( 9 + 16L ) µinch (Note: L in 'inches')
Height Equipment and Measure:
0.0 to 0.45 inches ( 48 + 1L ) µinch (Note: L in 'inches') Comparison to gauge blocks with surface plate.  Transcat calibration procedure
0.45 to 1.0 inch ( 46 + 4L ) µinch (Note: L in 'inches')
1.0 to 4.0 inches ( 40 + 10L ) µinch (Note: L in 'inches')
4.0 to 24 inches ( 21 + 15L ) µinch (Note: L in 'inches')

Electrical

Measured Quantity & Range or Instrument Frequency Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Current, AC

0 μA to 220 μA

10 Hz to 20 Hz 0.031 % + 16 nA

Fluke 5700A-EP or an equivalent standard. Source. For the calibration of sinewave current measurement devices. On-site calibration available.

20 Hz to 40 Hz 0.019 % + 10 nA
40 Hz to 1 kHz 0.015 % + 8 nA
1 kHz to 5 kHz 0.030 % + 12 nA
5 kHz to 10 kHz 0.11 % + 65 nA

0.22 mA to 2.2 mA

10 Hz to 20 Hz 0.03 % + 40 nA
20 Hz to 40 Hz 0.018 % + 35 nA
40 Hz to 1 kHz 0.014 % + 35 nA
1 kHz to 5 kHz 0.021 % + 0.11 μA
5 kHz to 10 kHz 0.11 % + 0.65 μA

2.2 mA to 22 mA

10 Hz to 20 Hz 0.039 % + 0.40 μA
20 Hz to 40 Hz 0.019 % + 0.35 μA
40 Hz to 1 kHz 0.014 % + 0.35 μA
1 kHz to 5 kHz 0.021 % + 0.55 μA
5 kHz to 10 kHz 0.11 % + 5 μA

22 mA to 220 mA

10 Hz to 20 Hz 0.033 % + 4 μA
20 Hz to 40 Hz 0.018 % + 3.5 μA
40 Hz to 1 kHz 0.014 % + 2.5 μA
1 kHz to 5 kHz 0.021 % + 3.5 μA
5 kHz to 10 kHz 0.11 % + 10 μA
0.22 A to 2.2 A 20 Hz to 1 kHz 0.027 % + 35 μA
1 kHz to 5 kHz 0.046 % + 80 μA
5 kHz to 10 kHz 0.7 % + 0.16 mA
2.2 A to 11 A 40 Hz to 1 kHz 0.048 % + 0.17 mA

Fluke 5700A-EP and 5725A or equivalent standards. Source. For the calibration of sinewave current measurement devices.  On-site calibration available.

1 kHz to 5 kHz 0.096 % + 0.38 mA
5 kHz to 10 kHz 0.36 % + 0.75 mA
11 A to 20.5 A 45 Hz to 100 Hz 0.095 % + 3.9 mA Fluke 5520A or equivalent standards. Source. For the calibration of sinewave current measurement devices.  On-site calibration available.
100 Hz to 1 kHz 0.12 % + 3.9 mA
1 kHz to 5 kHz 2.3 % + 3.9 mA
29 μA to 330 μA 10 kHz to 30 kHz 1.2 % + 0.31 μA
330 μA to 3.3 mA 10 kHz to 30 kHz 0.78 % + 0.47 μA
3.3 mA to 33 mA 10 kHz to 30 kHz 0.31 % + 3.1 µA
33 mA to 330 mA 10 kHz to 30 kHz 0.31 % + 0.16 mA

0 μA to 100 μA

10 Hz to 20 Hz 0.47 % + 35 nA

Keysight 3458A or equivalent standard.  Measure. For the calibration of current sources using a digital multimeter. On-site calibration available

20 Hz to 45 Hz 0.18 % + 35 nA
45 Hz to 100 Hz 0.072 % + 35 nA
100 Hz to 5 kHz 0.072 % + 35 nA

100 μA to 1 mA

10 Hz to 20 Hz 0.46 % + 0.23 μA
20 Hz to 45 Hz 0.18 % + 0.23 μA
45 Hz to 100 Hz 0.071 % + 0.23 μA
100 Hz to 5 kHz 0.038 % + 0.23 μA

1 mA to 10 mA

10 Hz to 20 Hz 0.46 % + 2.3 μA
20 Hz to 45 Hz 0.18 % + 2.3 μA
45 Hz to 100 Hz 0.071 % + 2.3 μA
100 Hz to 5 kHz 0.038 % + 2.3 μA
10 mA to 100 mA 10 Hz to 20 Hz 0.46 % + 23 μA
20 Hz to 45 Hz 0.18 % + 23 μA
45 Hz to 100 Hz 0.071 % + 23 μA
100 Hz to 5 kHz 0.037 % + 23 μA
100 mA to 1 A 10 Hz to 20 Hz 0.46 % + 0.23 mA
20 Hz to 45 Hz 0.19 % + 0.23 mA
45 Hz to 100 Hz 0.097 % + 0.23 mA
100 Hz to 5 kHz 0.12 % + 0.23 mA
 1 A to 2 A 10 Hz to 45 Hz 0.12 % + 4 mA

Yokogawa WT310EH or equivalent standard. Measure. For the calibration of current sources using a digital multimeter. On­site calibration available

45 Hz to 66 Hz 0.12 % + 1 mA
66 Hz to 1 kHz 0.12 % + 4 mA
2 A to 5 A 10 Hz to 45 Hz 0.13 % + 10 mA
45 Hz to 66 Hz 0.13 % + 2.5 mA
66 Hz to 1 kHz 0.13 % + 10 mA
5 A to 10 A 10 Hz to 45 Hz 0.13 % + 20 mA
45 Hz to 66 Hz 0.13 % + 5 mA
66 Hz to 1 kHz 0.13 % + 20 mA
10 A to 20 A 10 Hz to 45 Hz 0.16 % + 40 mA
45 Hz to 66 Hz 0.16 % + 10 mA
66 Hz to 1 kHz 0.18 % + 40 mA
20 A to 40 A 10 Hz to 45 Hz 0.20 % + 80 mA
45 Hz to 66 Hz 0.20 % + 20 mA
66 Hz to 1 kHz 0.23 % + 80 mA
Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Current, DC
0 µA to 220 µA 0.0041 % + 6 nA

Fluke 5700A-EP with 5725A or equivalent standards.  Source. For the calibration of DC current measuring devices. On­site calibration available

220 µA to 2.2 mA 0.0036 % + 7 nA
2.2 mA to 22 mA 0.0035 % + 40 nA
22 mA to 220 mA 0.0048 % + 0.7 μA
220 mA to 2.2 A 0.02 % + 12 μA
2.2 A to 11 A 0.04 % + 0.48 mA
11 A to 20.5 A 0.084 % + 0.58 mA Fluke 5520A or equivalent standard.  Source. For the calibration of DC current measuring devices. On-site calibration available.
0 µA to 100 µA 0.0026 % + 0.92 nA

Keysight 3458A or equivalent standard.  Measure. For the calibration of DC current generating devices. On-site calibration available.

100 µA to 1 mA 0.0026 % + 5.8 nA
1 mA to 10 mA 0.0026 % + 58 nA
10 mA to 100 mA 0.0043 % + 0.58 μA
100 mA to 1 A 0.0130 % + 12 μA
1 A to 50 A 0.03 % Yokogawa 2792 with Keysight 3458A or equivalent standards.  Measure. For the calibration of DC current generating devices. On-site calibration available.
50 A to 100 A 0.30 % Empro HA-100-50 with Keysight 3458A or equivalent standards.  Measure. For the calibration of DC current generating devices. On-site calibration available.
100 A to 1000 A 0.20 % Empro LAB-1000-100 with Keysight 3458A or equivalent standards.  Measure. For the calibration of DC current generating devices. On-site calibration available.
Voltage, DC
0 V to 220 mV 0.00085 % + 0.44 μV Fluke 5700A-EP or equivalent standard.  Source. For the calibration of DC voltage measuring devices. On-site calibration available.
220 mV to 2.2 V 0.00051 % + 0.7 μV
2.2 V to 11 V 0.00040 % + 2.5 μV
11 V to 22 V 0.00040 % + 4 μV
22 V to 220 V 0.00062 % + 40 μV
220 V to 1100 V 0.00076 % + 0.40 mV
0 V to 100 mV 0.00071 % + 0.58 μV Keysight 3458A or equivalent standard.  Measure. For the calibration of DC voltage generating devices. On-site calibration available.
100 mV to 1 V 0.00050 % + 0.58 μV
1 V to 10 V 0.00051 % + 0.58 μV
10 V to 100 V 0.00076 % + 35 μV
100 V to 500 V 0.0011 % + 0.12 mV
500 V to 800 V 0.0016 % + 0.12 mV
800 V to 1000 V 0.0021 % + 0.12 mV
1 kV to 10 kV 0.0374 % + 0.035  V Vitrek 4700 or/with HVL-100 probe or equal standard.  Measure. For the calibration of DC voltage generating devices.
10 kV to 100 kV 0.0625 % + 0.35  V
Measured Quantity & Range or Instrument Frequency Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Voltage, AC
0 mV to 2.2 mV 10 Hz to 20 Hz 0.16 % + 4 μV

Fluke 5700A-EP or equivalent standard.  Source. For the calibration of sinewave voltage measurement devices. On-site calibration available.

20 Hz to 40 Hz 0.1 % + 4 μV
40 Hz to 20 kHz 0.078 % + 4 μV
20 kHz to 50 kHz 0.13 % + 4 μV
50 kHz to 100 kHz 0.17 % + 5 μV
100 kHz to 300 kHz 0.33 % + 10 μV
300 kHz to 500 kHz 0.47 % + 20 μV
500 kHz to 1 MHz 0.58 % + 20 μV
2.2 mV to 22 mV 10 Hz to 20 Hz 0.042 % + 4 μV
20 Hz to 40 Hz 0.03 % + 4 μV
40 Hz to 20 kHz 0.014 % + 4 μV
20 kHz to 50 kHz 0.03 % + 4 μV
50 kHz to 100 kHz 0.058 % + 5 μV
100 kHz to 300 kHz 0.12 % + 10 μV
300 kHz to 500 kHz 0.16 % + 20 μV
500 kHz to 1 MHz 0.27 % + 20 μV
22 mV to 220 mV 10 Hz to 20 Hz 0.028 % + 12 μV
20 Hz to 40 Hz 0.011 % + 7 μV
40 Hz to 20 kHz 0.0085 % + 7 μV
20 kHz to 50 kHz 0.021 % + 7 μV
50 kHz to 100 kHz 0.047 % + 17 μV
100  kHz to 300 kHz 0.091 % + 20 μV
300 kHz to 500 kHz 0.14 % + 25 μV
500 kHz to 1 MHz 0.28 % + 45 μV
220 mV to 2.2 V 10 Hz to 20 Hz 0.027 % + 40 μV
20 Hz to 40 Hz 0.010 % + 15 μV
40 Hz to 20 kHz 0.0048 % + 8 μV
20 kHz to 50 kHz 0.008 % + 10 μV
50 kHz to 100 kHz 0.012 % + 30 μV
100 kHz to 300 kHz 0.043 % + 80 μV
300 kHz to 500 kHz 0.01 % + 200 μV
500 kHz to 1 MHz 0.18 % + 300 μV
2.2 V to 22 V 10 Hz to 20 Hz 0.028 % + 0.4 mV
20 Hz to 40 Hz 0.010 % + 0.15 mV
40 Hz to 20 kHz 0.005 % + 50 μV
20 kHz to 50 kHz 0.0083 % + 0.1 mV
50 kHz to 100 kHz 0.012 % + 0.2 mV
100 kHz to 300 kHz 0.030 % + 0.6 mV
300 kHz to 500 kHz 0.11 % + 2 mV
500 kHz to 1 MHz 0.17 % + 3.2 mV
22 V to 220 V 10 Hz to 20 Hz 0.028 % + 4 mV
20 Hz to 40 Hz 0.01 % + 1.5 mV
40 Hz to 20 kHz 0.0057 % +  0.6 mV
20 kHz to 50 kHz 0.0093 % + 1 mV
50 kHz to 100 kHz 0.017 % + 2.5 mV
100 kHz to 300 kHz 0.091 % + 16 mV
300 kHz to 500 kHz 0.44 % + 40 mV
500 kHz to 1 MHz 0.8 % + 80 mV
220 V to 1100 V 40 Hz to 1 kHz 0.011 % + 4 mV Fluke 5700A-EP with Fluke 5725A or an equivalent standard. Source. For the calibration of sinewave voltage measurement devices. On­site calibration available.
1 kHz to 20 kHz 0.017 % + 6 mV
20 kHz to 30 kHz 0.061 % + 11 mV
220 V to 750 V 30 kHz to 50 kHz 0.061 % + 11 mV
50 kHz to 100 kHz 0.23 % + 45 mV

0 mV to 10 mV

1 Hz to 40 Hz 0.039 % + 3.5 μV

Keysight 3458A or equivalent standard.  Measure. For the calibration of voltage sources using a digital multimeter. On-site calibration available.

40 Hz to 1 kHz 0.029 % + 1.2 μV
1 kHz to 20 kHz 0.039 % + 1.2 μV
20 kHz to 50 kHz 0.15 % + 1.2 μV
50 kHz to 100 kHz 0.59 % + 1.2 μV
100 kHz to 300 kHz 4.7 % + 2.3 μV
300 kHz to 1 MH 1.5 % + 5.8 μV
10 mV to 100 mV 1 Hz to 40 Hz 0.013 % + 4.6 μV
40 Hz to 1 kHz 0.0095 % + 2.3 μV
1 kHz to 20 kHz 0.017 % + 2.3 μV
20 kHz to 50 kHz 0.037 % + 2.3 μV
50 kHz to 100 kHz 0.093 % + 2.3 μV
100 kHz to 300 kHz 0.36 % + 12 μV
300 kHz to 1 MHz 1.2 % + 12 μV
1 MHz to 2 MHz 1.9 % + 12 μV
2 MHz to 4 MHz 4.7 % + 81 μV
4 MHz to 8 MHz 4.7 % + 92 μV
100 mV to 1 V 1 Hz to 40 Hz 0.0098 % + 46 μV
40 Hz to 1 kHz 0.0094 % + 23 μV
1 kHz to 20 kHz 0.017 % + 23 μV
20 kHz to 50 kHz 0.036 % + 23 μV
50 kHz to 100 kHz 0.093 % + 23 μV
100 kHz to 300 kHz 0.35 % + 0.12 mV
300 kHz to 1 MHz 1.2 % + 0.12 mV
1 MHz to 2 MHz 1.9 % + 0.12 mV
2 MHz to 4 MHz 4.7 % + 0.81 mV
4 MHz to 8 MHz 4.7 % + 0.92 mV
1V to 10 V 1 Hz to 40 Hz 0.0095 % + 0.46 mV
40 Hz to 1 kHz 0.0095 % + 0.23 mV
1 kHz to 20 kHz 0.017 % + 0.23 mV
20 kHz to 50 kHz 0.036 % + 0.23 mV
50 kHz to 100 kHz 0.093 % + 0.23 mV
100 kHz to 300 kHz 0.35 % + 1.2 mV
300 kHz to 1 MHz 1.7 % + 1.2 mV
1 MHz to 2 MHz 1.8 % + 1.2 mV
2 MHz to 4 MHz 4.7 % + 8.1 mV
4 MHz to 8 MHz 4.7 % + 9.3 mV
10 V to 100 V 1 Hz to 40 Hz 0.024 % + 4.6 mV
40 Hz to 1 kHz 0.024 % + 2.3 mV
1 kHz to 20 kHz 0.024 % + 2.3 mV
20 kHz to 50 kHz 0.042 % + 2.3 mV
50 kHz to 100 kHz 0.14 % + 2.3 mV
100 kHz to 300 kHz 0.46 % + 12 mV
300 kHz to 1 MHz 1.7 % + 12 mV
100 V to 700 V 1 Hz to 40 Hz 0.047 % + 46 mV
40 Hz to 1 kHz 0.047 % + 23 mV
1 kHz to 20 kHz 0.071 % + 23 mV
20 kHz to 50 kHz 0.14 % + 23 mV
50 kHz to 100 kHz 0.35 % + 23 mV
700 V to 10 kV 10 Hz to 200 Hz 0.14 % + 0.12 V Vitrek 4700 or/with HVL-100 probe or equal standard. Measure. For the calibration of voltage sources. On­site calibration available.
200 Hz to 450 Hz 0.46 % + 0.12 V
10 kV to 70 kV 30 Hz to 70 Hz 0.14 % + 0.7 V
70 Hz to 200 Hz 1.2 % + 0.7 V

Current clamp calibration

Measured Quantity & Range or Instrument Frequency Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Clamp-on Ammeter Toroidal Type: Effective AC current output
20.5 A to 150 A 45 Hz to 65 Hz 0.42 % + 29 mA Fluke 5520A with 5500A Current Coil or equivalent standard.  Source using a multifunction calibrator and a 50 turn coil. For the calibration of clamp meters. On-site calibration available.
65 Hz to 440 Hz 1.0 % + 31 mA
150 A to 500 A 45 Hz to 65 Hz 0.42 % + 130 mA
65 Hz to 440 Hz 1.0 % + 140 mA
500 A to 1025 A 45 Hz to 65 Hz 0.43 % + 220 mA
65 Hz to 440 Hz 1.0 % + 230 mA
Clamp-on Ammeter Non-Toroidal Type: Effective AC current output
20.5 A to 150 A 45 Hz to 65 Hz 0.70 % + 290 mA Fluke 5520A with 5500A Current Coil or equivalent standard.  Source using a multifunction calibrator and a 50 turn coil. For the calibration of clamp meters. On-site calibration available.
65 Hz to 440 Hz 1.3 % + 290 mA
150 A to 500 A 45 Hz to 65 Hz 0.70 % + 300 mA
65 Hz to 440 Hz 1.3 % + 1.0 A
500 A to 1025 A 45 Hz to 65 Hz 0.70 % + 350 mA
65 Hz to 440 Hz 1.3 % + 1.1 A
Clamp-on Ammeter Non-Toroidal Type: Effective DC current output
20 A to 150 A   0.58 % + 0.16 A Fluke 5520A with 5500A Current Coil or equivalent standard.  Source using a multifunction calibrator and a 50 turn coil. For the calibration of clamp meters. On-site calibration available.
150 A to 1000 A   0.58 % + 0.58 A

Power, DC

Measured Quantity & Range or Instrument Frequency Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
0.33 mA to 330 mA
11 μW to 1.1 mW DC 0.024 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators
1.1 mW to 110 mW DC 0.027 %
0.11 W to 110 W DC 0.024 %
110 W to 330 W DC 0.018 %
0.33 A to 3 A
11 W to 110 mW DC 0.044 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators
0.11 W to 990 W DC 0.053 %
1 W to 3 kW DC 0.0096 %
3 A to 20.5 A
0.099 W to 0.99 W DC 0.088 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators
0.99 W to 6.8 kW DC 0.07 %
6.8 kW to 20.5 kW DC 0.04 %

Power, AC

Measured Quantity & Range or Instrument Frequency Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
3.3 mA to 9 mA
0.11 mW to 3 mW 10 Hz to 65 Hz 0.13 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
3 mW to 9 W 10 Hz to 65 Hz 0.077 %
9 mA to 33 mA
0.3 mW to 10 mW 10 Hz to 65 Hz 0.089 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
10 mW to 33 W 10 Hz to 65 Hz 0.077 %
33 mA to 90 mA
1 mW to 30 mW 10 Hz to 65 Hz 0.071 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
30 mW to 90 W 10 Hz to 65 Hz 0.057 %
90 mA to 330 mA
3.0 mW to 100 mW 10 Hz to 65 Hz 0.089 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
100 mW to 300 W 10 Hz to 65 Hz 0.078 %
0.33 A to 0.9 A
11 mW to 300 mW 10 Hz to 65 Hz 0.071 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
300 mW to 900 W 10 Hz to 65 Hz 0.081 %
0.9 A to 2.2 A
30 mW to 720 mW 10 Hz to 65 Hz 0.089 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
720 mW to 2 kW 10 Hz to 65 Hz 0.079 %
2.2 A to 4.5 A
80 mW to 1.4 W 10 Hz to 65 Hz 0.088 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
1.4 W to 4.5 kW 10 Hz to 65 Hz 0.18 %
4.5 A to 20.5 A
150 mW to 6.7 W 10 Hz to 65 Hz 0.17 % Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. (PF = 1).  See Note Footnote 1.
6.7 W to 20 kW 10 Hz to 65 Hz 0.17 %
Phase
0 ° to 179.99 ° 10 Hz to 65 Hz 0.11 °

Fluke 5520A or equivalent standard.  Source. For the calibration of measurement devices using multifunction calibrators. On-site calibration available.

65 Hz to 500 Hz 0.20 °
500 Hz to 1 kHz 0.39 °
1 kHz to 5 kHz 1.9 °
5 kHz to 10 kHz 3.9 °
10 kHz to 30 kHz 7.8 °
Capacitance
0.19 nF to < 1.1 nF 10 Hz to 10 kHz 0.39 % + 7.8 pF Fluke 5520A or equivalent standard.  Source synthesized capacitance using a multiproduct calibrator. For the calibration of capacitance measuring devices. On-site calibration available.
1.1 nF to < 3.3 nF 10 Hz to 3 kHz 0.39 % + 7.8 pF
3.3 nF to < 11 nF 10 Hz to 1 kHz 0.21 % + 7.8 pF
11 nF to < 33 nF 10 Hz to 1 kHz 0.21 % + 78 pF
33 nF to < 110 nF 10 Hz to 1 kHz 0.21 % + 78 pF
110 nF to < 330 nF 10 Hz to 1 kHz 0.21 % + 0.23 nF
0.33 µF to < 1.1 µF 10 Hz to 600 Hz 0.20 % + 0.78 nF
1.1 µF to < 3.3 µF 10 Hz to 300 Hz 0.20 % + 2.3 nF
3.3 µF to < 11 µF 10 Hz to 150 Hz 0.20 % + 7.8 nF
11 µF to < 33 µF 10 Hz to 120 Hz 0.32 % + 23 nF
33 µF to < 110 µF 10 Hz to 80 Hz 0.35 % + 78 nF
110 µF to < 330 µF DC to 50 Hz 0.35 % + 0.23 µF
0.33 mF to < 1.1 mF DC to 20 Hz 0.35 % + 0.78 µF
1.1 mF to < 3.3 mF DC to 6 Hz 0.35 % + 2.3 µF
3.3 mF to < 11 mF DC to 2 Hz 0.35 % + 7.8 µF
11 mF to < 33 mF DC to 0.6 Hz 0.58 % + 23 µF
33 mF to < 110 mF DC to 0.2 Hz 0.86 % + 78 µF
0.001 µF 1 kHz 0.065 % Standard Capacitors. For the calibration of capacitance measuring devices. On­site calibration available.
0.01 µF 1 kHz 0.063 %
0.05 µF 1 kHz 0.063 %
0.1 µF 1 kHz 0.063 %
1 µF 1 kHz 0.063 %
Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Resistance
10 MΩ to 100 MΩ 0.08 % IET HRRS-B-7-100K-5KV or equivalent standard.  Source. For the calibration of resistance measurement devices On-site calibration available.
100 MΩ to 1 GΩ 0.24 %
1 GΩ to 10 GΩ 0.41 %
10 GΩ to 100 GΩ 0.84 %
100 GΩ to 1 TΩ 2.50 %
1 mΩ, 230 nΩ Fixed Resistors. Source. For the calibration of resistance measurement devices On­site calibration available.
10 mΩ 1.16 µΩ
100 mΩ 48 µΩ
0 Ω to 10 Ω 0.0018 % + 58 µΩ Keysight 3458A with decade resistance or equivalent standards.  Measure capability using a digital multimeter. On-site calibration available.
10 Ω to 100 Ω 0.0015 % + 0.58 mΩ
100 Ω to 1 kΩ 0.0012 % + 0.58 mΩ
1 kΩ to 10 kΩ 0.0012 % + 5.8 mΩ
10 kΩ to 100 kΩ 0.0012 % + 58 mΩ
100 kΩ to 1 MΩ 0.0019 % + 2.3 Ω
1 MΩ to 10 MΩ 0.0062 % + 0.12 kΩ
10 MΩ to 100 MΩ 0.059 % + 1.2 kΩ
100 MΩ to 1 GΩ 0.58 % + 12 kΩ

Electrical Calibration of Temperature Indicators and Simulators

Thermocouple simulation
Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Type B:
250 °C to 350 °C 1.0 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
350 °C to 445 °C 0.77 °C
445 °C to 580 °C 0.61 °C
580 °C to 750 °C 0.47 °C
750 °C to 1000 °C 0.39 °C
1000 °C to 1820 °C 0.31 °C
Type C:
0 °C to 250 °C 0.21 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
250 °C to 1000 °C 0.17 °C
1000 °C to 1500 °C 0.19 °C
1500 °C to 1800 °C 0.22 °C
1800 °C to 2000 °C 0.24 °C
2000 °C to 2250 °C 0.30 °C
2250 °C to 2315 °C

0.33 °C

Type E:
-270 °C to -245°C

2.1 °C

Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
-245 °C to -195 °C

0.20 °C

-195 °C to -155 °C

0.11 °C

-155 °C to -90 °C

0.09 °C

-90 °C to 0 °C

0.08 °C

0 °C to 15 °C

0.08 °C

15 °C to 890 °C

0.07 °C

890 °C to 1000 °C

0.08 °C

Type J:
-210 °C to -180 °C 0.13 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
-180 °C to -120 °C 0.11 °C
-120 °C to -50 °C 0.09 °C
-50 °C to 990 °C 0.08 °C
990 °C to 1200 °C 0.08 °C
Type K:
-270 °C to -255 °C 2.3 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
-255 °C to -195 °C

0.73 °C

-195 °C to -115 °C

0.14 °C

-115 °C to -55 °C

0.10 °C

-55 °C to 1000 °C

0.08 °C

1000 °C to 1372 °C

0.09 °C

Type N:
-270 °C to -260 °C 5.1 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
-260 °C to -200 °C

1.1 °C

-200 °C to -140 °C

0.25 °C

-140 °C to -70 °C

0.16 °C

-70 °C to 25 °C

0.13 °C

25 °C to 160 °C

0.11 °C

160 °C to 1300 °C

0.10 °C

Type R:
-50 °C to -30 °C 0.68 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
-30 °C to 45 °C

0.58 °C

45 °C to 160 °C

0.42 °C

160 °C to 380 °C

0.31 °C

380 °C to 775 °C

0.28 °C

775 °C to 1768 °C

0.23 °C

Type S:
-50 °C to -30 °C 0.65 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
-30 °C to 45 °C

0.59 °C

45 °C to 105 °C

0.42 °C

105 °C to 310 °C

0.35 °C

310 °C to 615 °C

0.31 °C

615 °C to 1768 °C 0.27 °C
Type T:
-270 °C to -255 °C 1.8 °C Ectron 1140A or equivalent standard. Source and Measure. For the calibration of temperature indicators and process calibrators by electrical stimulation of temperature. On­site calibration available.
-255 °C to -240 °C 0.52 °C
-240 °C to -210 °C 0.32 °C
-210 °C to -150 °C 0.19 °C
-150 °C to -40 °C 0.13 °C
-40 °C to 100 °C 0.09 °C
100 °C to 400 °C 0.08 °C

Oscilloscope

Parameters Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Amplitude DC
into 50 Ω Load (-6.6 to 6.6) V 0.20 % + 31 µV Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
into 1 MΩ Load (-130 to 130) V 0.039 % + 31 µV
Amplitude Square Wave
into 50 Ω Load
Rate: 10 Hz to 10 kHz
1 mV (pk-pk)  to 6.6 V (pk-pk) 0.19 % + 31 µV Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
into 1 MΩ Load
Rate: 10 Hz to 1 kHz
1 mV (pk-pk)  to 6.6 V (pk-pk) 0.078 % + 31 µV
into 1 MΩ Load
Rate: 1 kHz to 10 kHz
1 mV (pk-pk)  to 6.6 V (pk-pk) 0.19 % + 31 µV
Timing - Generate
50 Ω Load 5 s 0.30 % Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
2 s 0.12 %
1 s 0.062 %
500 ms 0.032 %
200 ms 0.014 %
100 ms 0.0076 %
50 ms 0.0046 %
20 ms  to 1 ns 0.00022 %
Rise Time
5.0 mV (pk-pk) to 2.5 V (pk-pk)
Rate: 1kHz to 10 MHz
250 ps (nominal) 50 ps Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.  See Note Footnote 2.
Leveled Sine Wave
into 50 Ω Load
5.0 mV (pk-pk) to 5.5 V (pk-pk)
50 kHz 1.8 % + 230 µV Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
50 kHz to 100 MHz 2.8 % + 230 µV
100 MHz to 300 MHz 3.2 % + 230 µV
300 MHz to 600 MHz 4.0 % + 230 µV
into 50Ω Load
5.0 mV (pk-pk) to 3.5 V (pk-pk)
600 MHz to 1 GHz 5.5 % + 230 µV
Bandwidth /Flatness
into 50 Ω Load (50 kHz Reference) 5.0 mV (pk-pk) to 5.5 V (pk-pk) 50 kHz to 100 MHz 1.4 % + 78 µV Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
100 MHz to 300 MHz 1.8 % + 78 µV
300 MHz to 600 MHz 3.2 % + 78 µV
into 50Ω Load (50 kHz Reference) 5.0 mV (pk-pk) to 3.5 V (pk-pk) 600 MHz to 1 GHz 4.0 % + 78 µV
Input Impedance
50 Ω Load 40 Ω to 60 Ω 0.082 % Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
1 MΩ Load 500 kΩ to 1.5 MΩ 0.081 %
Input Capacitance
  5.0 pF to 50 pF 3.9 % + 0.39 pF Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
Wave Generator – Amplitude
Sine, Square, Triangle
50 Ω Load 1.8 mV (pk-pk)  to 2.5 V (pk-pk) 2.3 % + 78 µV(pk-pk) Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.
1 MΩ Load 1.8 mV (pk-pk)  to 55 V (pk-pk) 2.3 % + 78 µV(pk-pk)
Wave Generator – Frequency
Sine, Square, Triangle
  10 Hz to 10 kHz 0.0019 % + 0.012 Hz Fluke 5520A / SC1100 or equivalent standard. Source. On­site calibration available.

Time and Frequency

Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Frequency
10 MHz 5.9 x 10-10 Rubidium Oscillator or equivalent standard.  Source.  Uncertainty values of derivatives of 10 MHz will differ due to resolution, noise, and gating errors.
10 MHz 2.3 x 10-9 Keysight 53132A or equivalent standard. Measure capability. On-site calibration available.  Uncertainty values of derivatives of 10 MHz will differ due to resolution, noise, and gating errors.

Force

Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Tension
1 lbf to 400 lbf 0.013 % + 0.6R Where R is the resolution of the device under test NIST Class F Weights
Compression
1 lbf to 400 lbf 0.013 % + 0.6R Where R is the resolution of the device under test NIST Class F Weights
Torque
4 lbf-in to 10 lbf-in 1.9 % Transcat calibration procedure. On-site calibration available.
10 lbf-in to 30 lbf-in 1.0 %
30 lbf-in to 80 lbf-ft 0.76 %
80 lbf-ft to 250 lbf-ft 0.60 %
250 lbf-ft to 600 lbf-ft 0.57 %

Mass

Balances

Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Specific values
30 kg 82 mg ASTM Class 1 weights or equal standard. See note Footnote 3.
25 kg 70 mg
20 kg 55 mg
10 kg 28 mg
5 kg 12.3 mg
2 kg 5.3 mg
1 kg 2.6 mg
500 g 1.9 mg
200 g 0.53 mg
100 g 280 μg
50 g 135 μg
20 g 94 μg
10 g 67 μg
5 g 55 μg
3 g 55 μg
2 g 55 μg
1 g 55 μg
500 mg 44 μg
200 mg 44 μg
100 mg 44 μg
50 mg 44 μg
20 mg 44 μg
10 mg 44 μg
5 mg 44 μg
1 mg 44 μg
Other values
1 lb to 400 lb 0.012 % NIST Class F weight or equal standard. See note Footnote 3.

Pressure

Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
0 psia to 7.5 psia 0.00075 psia Transcat calibration procedure.
Pneumatic: Absolute.
On-site calibration available.
7.5 psia to 15 psia 0.01 %
15 psia to 25 psia 0.0019 psia
25 psia to 500 psia 0.0065 % + 0.001 psia
500 psia to 1015 psia 0.01 %
-30 inHg to -2.3 inHg 0.0091 % Transcat calibration procedure.
Pneumatic: Gauge.
On-site calibration available.
-2.3 inHg to 0 inHg 0.0002 inHg
0 in H2O to 7 in H2O 0.0030 in H2O
7 in H2O to 40 in H2O 0.0023 in H2O
40 in H2O to 14.5 psig 0.0060 %
14.5 psig to 25 psig 0.0017 psig
25 psig to 500 psig 0.0065 %
500 psig to 1000 psig 0.01 %
10 psig to 50 psig 0.004 psig Transcat calibration procedure.
Hydraulic: Gauge.
On-site calibration available.
50 psig to 500 psig 0.0079 %
500 psig to 1000 psig 0.08 psig
1000 psig to 10000 psig 0.0082 %

Thermometry

Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Digital Thermometers, thermometers, RTD, PRT and Thermistor probes
-80 °C to 0 °C 0.018 °C Liquid Bath/Dry Block with HART 5628 Black Stack or equivalent standard. Source. For the calibration of temperature measuring equipment. On­site calibration available.
0.01 °C 0.0033 °C
0 °C to 150 °C 0.033 °C
150 °C C to 300°C 0.042 °C
300 °C to 660 °C 0.067 °C
Dry Blocks, Liquid Baths
­195 °C to 0 °C 0.012 °C HART 5628 Black Stack or equivalent standard. For the calibration of temperature sourcing equipment. On­site calibration available.
0 °C to 420 °C 0.026 °C
420 °C to 660 °C 0.037 °C
Thermocouples Wires and Probe – Type J, K, E, T
­80 °C to 0 °C 0.22 °C Liquid Bath/Dry Block with HART 5628 Black Stack and Ectron 1140A or equivalent standard. For the calibration of Thermocouple wires and probes. On­site calibration available.
0 °C to 100 °C 0.22 °C
100 °C to 150 °C 0.26 °C
150 °C to 300 °C 0.41 °C
300 °C to 660 °C 0.71 °C

Humidity

Measured Quantity & Range or Instrument Calibration Measurement Capability
expressed as an Uncertainty (±)
(See supplementary notes)
Remarks
Relative Humidity
0 °C to 15 °C 10% to 75 % RH 0.50 % RH

Thunder 2500 or equivalent standard. Source. For the calibration of relative humidity and temperature measuring instruments.  Transcat calibration procedure.

0 °C to 15 °C 75 % to 95 % RH 0.65 % RH
15 °C to 35 °C 10 % to 95 % RH 0.50 % RH
35 °C to 70°C 10 % to 50 % RH 0.50 % RH
35 °C to 70 °C 50 % to 75 % RH 0.70% RH
35 °C to 70 °C 75 % to 95 % RH 0.85% RH
Relative Humidity Controlled Chambers
­40°C to -20°C 0 % RH to 100 % RH 1.80 % RH Vaisala HM70/HMP77 or equivalent standard. Measure. For the calibration of humidity controlled chambers. Transcat calibration procedure. On­site calibration available.
-20°C to 40°C 0 % RH to 90 % RH 1.26 % RH
-20°C to 40°C 0 % RH to 100 % RH 2.03 % RH
40°C to 180°C 0 % RH to 100 % RH 1.80 % RH