CLAS Certificate Number 2016-01

Company name La Cie J. Chevrier Instruments Inc.
Company address 4850, bd Gouin est
Montréal-Nord, Quebec H1G 1A2
Contact Denis Chevrier
Telephone: 514-328-2550 poste 22
Toll free: 1-800-522-1226
Fax: 514-327-0604
Emailinfo@chevrierinstruments.com
Clients served
  • All interested parties
Fields of calibration
  • Pressure
  • Thermometry
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 151017
  • First issued
  • Issue 3.1

This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body. The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS. See supplementary notes.

Pressure

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Type of Service Remarks
0 psi to 1000 psi 0.4 psi II

For the calibration of Analog gauges, Digital gauges, Pressure transducers, Pressure calibrators

Comparison method with pneumatic digital pressure calibrator

1000 psi to 3000 psi 1 psi
200 psi to 16000 psi 6 psi

For the calibration of Analog gauges, Digital gauges, Pressure transducers, Pressure calibrators

Hydraulic deadweight tester

0 kPa to 18 kPa 0.003 kPa II

For the calibration of digital pressure calibrator

Comparison method with differential or static digital pressure calibrator.

18 kPa to 200 kPa 0.03 kPa

Thermometry

Measured Quantity & Range or Instrument Expanded Measurement Uncertainty Type of Service Remarks
Resistance temperature devices
0.01 °C, triple point of water 0.01 °C II For the calibration of Resistance Temperature Devices (PRT, RTD) using: Triple point of water cell
29.7646 °C, melting point of gallium 0.006 °C For the calibration of Resistance Temperature Devices (PRT, RTD) using: Gallium melting point cell