Montréal-Nord, Quebec H1G 1A2
Telephone: 514-328-2550 poste 22
Toll-Free: 1-800-522-1226
Fax: 514-327-0604
Email: chevrierd1@chevrierinstruments.com
- All interested parties
- Pressure
- Thermometry
- Standards Council of Canada (SCC)
- SCC file number: 151017
- First issued
- Issue 3.1
This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body.
The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS.
See supplementary notes.
Pressure
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type of Service | Remarks |
---|---|---|---|
0 psi to 1000 psi |
0.4 psi |
II |
For the calibration of Analog gauges, Digital gauges, Pressure transducers, Pressure calibrators. |
1000 psi to 3000 psi |
1 psi |
II |
For the calibration of Analog gauges, Digital gauges, Pressure transducers, Pressure calibrators. |
200 psi to 16000 psi |
6 psi |
II |
For the calibration of Analog gauges, Digital gauges, Pressure transducers, Pressure calibrators. |
0 kPa to 18 kPa |
0.003 kPa |
II |
For the calibration of digital pressure calibrator. |
18 kPa to 200 kPa |
0.03 kPa |
II |
For the calibration of digital pressure calibrator. |
Thermometry
Resistance temperature devices
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type of Service | Remarks |
---|---|---|---|
0.01 °C, triple point of water |
0.01 °C |
II |
For the calibration of Resistance Temperature Devices (PRT, RTD) using: Triple point of water cell. |
29.7646 °C, melting point of gallium |
0.006 °C |
II |
For the calibration of Resistance Temperature Devices (PRT, RTD) using: Gallium melting point cell. |