Company name | SRP Control Systems Ltd. |
---|---|
Company address | 19 – 5155 Spectrum Way Mississauga, Ontario L4W 5A1 |
Contact | Frederic Langlat Telephone: 905-238-2880 Fax: 905-238-9590 Email: flanglat@srpcontrol.com |
Clients served |
|
Fields of calibration |
|
Accreditation body |
|
ISO/IEC 17025 |
|
This CLAS Certificate is published by the CLAS program of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to the CLAS program. See supplementary notes.
Electrical
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type | Remarks |
---|---|---|---|
Voltage, DC | |||
0 mV to 220 mV | 7.5 µV/V + 0.4 μV | II | Source. For the calibration of dc tbAvoltage measuring devices using a multifunction calibrator. |
220 mV to 2.2 V | 5 µV/V + 0.7 μV | ||
2.2 mV to 11 V | 3.5 µV/V + 2.5 μV | ||
11 mV to 22 V | 3.5 µV/V + 4 μV | ||
22 V to 220 V | 5 µV/V + 40 μV | ||
220 V to 1100 V | 6.5 µV/V + 400 μV | ||
0 V to 100 mV | 9 µV/V + 0.3 μV | II | Measure. For the calibration of dc tbAvoltage generating devices using a digital multimeter. |
100 mV to 1 V | 8 µV/V + 0.3 μV | ||
1 V to 10 V | 8 µV/V + 0.5 μV | ||
10 V to 100 V | 10 µV/V + 30 μV | ||
100 V to 1000 V | 10 µV/V + 100 μV | ||
Current, DC | |||
0 μA to 220 μA | 40 µA/A + 6 nA | II | Source. For the calibration of dc tbAcurrent measuring devices using a multifunction calibrator. |
220 μA to 2.2 mA | 35 µA/A + 7 nA | ||
2.2 mA to 22 mA | 35 µA/A + 40 nA | ||
22 mA to 100 mA | 45 µA/A + 0.7 µA | ||
100 mA to 220 mA | 45 µA/A + 0.7 µA + 200·I2 µA/A | ||
220 mA to 1 A | 80 µA/A + 12 µA | ||
1 A to 2.2 A | 80 µA/A + 12 µA + 10·I2 µA/A | ||
2.2 A to 11 A | 360 µA/A + 480 µA | ||
0 to 100 nA | 30 µA/A + 40 pA | II | Measure. For the calibration of dc tbAcurrent generating devices using a digital multimeter. |
100 nA to 1 μA | 20 µA/A + 40 pA | ||
1 μA to 10 μA | 20 µA/A + 0.1 nA | ||
10 μA to 100 μA | 20 µA/A + 0.8 nA | ||
100 μA to 1 mA | 20 µA/A + 5 nA | ||
1 mA to 10 mA | 20 µA/A + 50 nA | ||
10 mA to 100 mA | 35 µA/A + 0.5 μA | ||
100 mA to 1 A | 110 µA/A + 10 μA | ||
Resistance | |||
0 Ω | 40 μΩ | II | Source. For the calibration of tbAresistance measurement devices using a multifunction calibrator. |
1 Ω | 95 μΩ/Ω | ||
1.9 Ω | 95 μΩ/Ω | ||
10 Ω | 23 μΩ/Ω | ||
19 Ω | 23 μΩ/Ω | ||
100 Ω | 10 μΩ/Ω | ||
190 Ω | 10 μΩ/Ω | ||
1 kΩ | 6.5 μΩ/Ω | ||
1.9 kΩ | 6.5 μΩ/Ω | ||
10 kΩ | 6.5 μΩ/Ω | ||
19 kΩ | 6.5 μΩ/Ω | ||
100 kΩ | 8.5 μΩ/Ω | ||
190 kΩ | 8.5 μΩ/Ω | ||
1 MΩ | 13 μΩ/Ω | ||
1.9 MΩ | 18 μΩ/Ω | ||
10 MΩ | 40 μΩ/Ω | ||
19 MΩ | 47 μΩ/Ω | ||
100 kΩ | 100 μΩ/Ω | ||
0.47 Ω to 10 Ω | 15 μΩ/Ω + 50 μΩ | II | Measure capability using a digital multimeter. |
10 Ω to 100 Ω | 12 μΩ/Ω + 0.5 mΩ | ||
100 Ω to 1 kΩ | 10 μΩ/Ω + 0.5 mΩ | ||
1 kΩ to 10 kΩ | 10 μΩ/Ω + 5 mΩ | ||
10 kΩ to 100 kΩ | 10 μΩ/Ω + 50 mΩ | ||
100 kΩ to 1 mΩ | 15 μΩ/Ω + 2 Ω | ||
1 mΩ to 10 mΩ | 50 μΩ/Ω + 100 Ω | ||
10 mΩ to 100 mΩ | 500 μΩ/Ω + 1 kΩ | ||
100 mΩ to 1GΩ | 5000 μΩ/Ω + 10 kΩ |
Measured Quantity & Range or Instrument | Frequency | Expanded Measurement Uncertainty | Type | Remarks |
---|---|---|---|---|
Voltage, AC | ||||
220 μV to 1100 V | 10 Hz to 100 kHz | 46 uV/V to 1.2 % | II | Source. For the calibration of sinewave voltage measurement devices using a multifunction calibrator. See Annex A for details. |
1 mV to 700 V | 10 Hz to 100 kHz | 90 uV/V to 16 % | II | Measure. For the calibration of voltage sources using a digital multimeter. See Annex C for details. |
Current, AC | ||||
10 μA to 11 A | 10 Hz to 10 kHz | 140 uA/A to 0.78 % | II | Source. For the calibration of sinewave current measurement devices using a multifunction calibrator. See Annex B for details. |
10 μA to 1 A | 10 Hz to 10 kHz | 500 uA/A to 1.4 % | II | Measure. For the calibration of current sources using a digital multimeter. See Annex D for details. |
Time and Frequency
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type | Remarks |
---|---|---|---|
Frequency measure | |||
1 Hz to 10 MHz | 0.02 µHz/Hz + 2 µHz | II | Measure for the calibration of frequency sources devices using a GPS Frequency standard |
Frequency source, fixed output | |||
1 Hz, 2 Hz, 5 Hz 10 Hz, 20 Hz, 50 Hz 100 Hz, 200 Hz, 500 Hz 1 kHz, 2 kHz 10 kHz, 20 kHz, 50 kHz 100 kHz, 200 kHz, 500 kHz 1 MHz, 2 MHz, 5 MHz, 10 MHz |
0.02 µHz/Hz + 2 µHz | II | Source for the calibration of frequency measuring devices using a GPS Frequency standard |
Frequency source, variable output | |||
10 Hz to 2 MHz | 1 µHz/Hz + 1 µHz | II | Source for the calibration of frequency measuring devices using a GPS Frequency standard |
5 MHz to 10 MHz | 0.02 µHz/Hz + 2 µHz |
Pressure
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type | Remarks |
---|---|---|---|
Hydraulic pressure measurement | |||
1.3 MPa to 137.8 MPa | 0.005 % + 80 Pa | II | For the calibration of pressure measuring devices by comparison against primary standard pressure balance DH5304 |
200 psi to 20000 psi | 0.005 % + 0.012 psi | ||
0 MPa to 275.8 MPa | 0.18 MPa | II | For the calibration of pressure measuring devices by comparison against Reference gauge 685 |
0 psi to 40000 psi | 26 psi | ||
Low gauge pressure measurement, pneumatic | |||
0 Pa to 750 Pa | 0.009 % + 0.064 Pa | II | For the calibration of pressure measuring devices by comparison against low pressure controller Ruska 7250 LP |
0 in H2O to 3 in H2O | 0.009 % + 0.00026 in H2O | ||
750 Pa to 3750 Pa | 0.010 % + 0.051 Pa | ||
3 in H2O to 15.1 in H2O | 0.010 % + 0.00020 in H2O | ||
3750 Pa to 7500 Pa | 0.010 % + 0.038 Pa | ||
15.1 in H2O to 30 in H2O | 0.010 % + 0.00015 in H2O | ||
Gauge pressure measurement, pneumatic | |||
0 kPa to 500 kPa | 0.0060 % + 7 Pa | II | For the calibration of pressure measuring devices by comparison against digital piston gauge DH DPG10G |
0 psi to 72.5 psi | 0.0060 % + 0.0010 psi | ||
0 kPa to 2000 kPa | 0.0060 % + 24 Pa | ||
0 psi to 290 psi | 0.0060 % + 0.0035 psi | ||
0 kPa to 5000 kPa | 0.0060 % + 58 Pa | ||
0 psi to 725 psi | 0.0060 % + 0.0084 psi | ||
0 kPa to 17236 kPa | 0.0060 % + 220 Pa | ||
0 psi to 2500 psi | 0.0060 % + 0.032 psi | ||
Gauge pressure measurement, pneumatic | |||
1.4 kPa to 172 kPa | 0.0025 % + 0.04 Pa | II | For the calibration of pressure measuring devices by comparison against primary piston gauge Ruska 2465 |
0.2 psi to 25 psi | 0.0025 % + 0.000006 psi | ||
12 kPa to 700 kPa | 0.0025 % + 1.5 Pa | ||
1.7 psi to 100 psi | 0.0025 % + 0.00022 psi | ||
14 kPa to 3500 kPa | 0.0030 % + 1.5 Pa | ||
2 psi to 500 psi | 0.0030 % + 0.00022 psi | ||
14 kPa to 7000 kPa | 0.0030 % + 5 Pa | ||
2 psi to 1000 psi | 0.0030 % + 0.00073 psi | ||
Absolute pressure measurement, pneumatic | |||
1.4 kPa to 172 kPa | 0.0025 % + 0.7 Pa | II | For the calibration of pressure measuring devices by comparison Direct comparison against primary piston gauge (Ruska 2465). |
0.2 psi to 25 psi | 0.0025 % + 0.0001 psi | ||
12 kPa to 700 kPa | 0.0025 % + 1.5 Pa | ||
1.7 psi to 100 psi | 0.0025 % + 0.00022 psi | ||
14 kPa to 3500 kPa | 0.0030 % + 1.5 Pa | ||
2 psi to 500 psi | 0.0030 % + 0.00022 psi | ||
14 kPa to 7000 kPa | 0.0030 % + 5 Pa | ||
2 psi to 1000 psi | 0.0030 % + 0.00073 psi |
Electrical Calibration of Temperature Indicators and Simulators
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type | Remarks |
---|---|---|---|
Thermocouple simulation with cold junction compensation | |||
Type B | |||
600 °C to 800 °C | 0.21 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
800 °C to 900 °C | 0.16 °C | ||
900 °C to 1550 °C | 0.15 °C | ||
1550 °C to 1820 °C | 0.11 °C | ||
Type E | |||
-250 °C to -100 °C | 0.58 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -25 °C | 0.12 °C | ||
-25 °C to 350 °C | 0.10 °C | ||
350 °C to 1000 °C | 0.08 °C | ||
Type J | |||
-210 °C to -100 °C | 0.25 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -30 °C | 0.12 °C | ||
-30 °C to 150 °C | 0.10 °C | ||
150 °C to 1200 °C | 0.09 °C | ||
Type K | |||
-200 °C to -100 °C | 0.25 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -25 °C | 0.13 °C | ||
-25 °C to 1000 °C | 0.10 °C | ||
1000 °C to 1372 °C | 0.12 °C | ||
Type N | |||
-200 °C to -100 °C | 0.27 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -25 °C | 0.13 °C | ||
-25 °C to 410 °C | 0.11 °C | ||
410 °C to 1300 °C | 0.08 °C | ||
Type R | |||
-50 °C to 0 °C | 0.36 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
0 °C to 250 °C | 0.25 °C | ||
250 °C to 900 °C | 0.14 °C | ||
900 °C to 1768 °C | 0.11 °C | ||
Type S | |||
-50 °C to 0 °C | 0.34 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
0 °C to 250 °C | 0.25 °C | ||
250 °C to 900 °C | 0.15 °C | ||
900 °C to 1768 °C | 0.13 °C | ||
Type T | |||
-250 °C to -150 °C | 0.61 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-150 °C to 0 °C | 0.17 °C | ||
0 °C to 120 °C | 0.10 °C | ||
120 °C to 400 °C | 0.08 °C | ||
Thermocouple simulation without cold junction compensation | |||
Type B | |||
600 °C to 800 °C | 0.12 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
800 °C to 900 °C | 0.09 °C | ||
900 °C to 1550 °C | 0.09 °C | ||
1550 °C to 1820 °C | 0.07 °C | ||
Type E | |||
-250 °C to -100 °C | 0.08 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -25 °C | 0.02 °C | ||
-25 °C to 350 °C | 0.02 °C | ||
350 °C to 1000 °C | 0.02 °C | ||
Type J | |||
-210 °C to -100 °C | 0.04 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -30 °C | 0.02 °C | ||
-30 °C to 150 °C | 0.02 °C | ||
150 °C to 1200 °C | 0.02 °C | ||
Type K | |||
-200 °C to -100 °C | 0.05 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -25 °C | 0.03 °C | ||
-25 °C to 1000 °C | 0.03 °C | ||
1000 °C to 1372 °C | 0.03 °C | ||
Type N | |||
-200 °C to -100 °C | 0.07 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-100 °C to -25 °C | 0.04 °C | ||
-25 °C to 410 °C | 0.03 °C | ||
410 °C to 1300 °C | 0.03 °C | ||
Type R | |||
-50 °C to 0 °C | 0.19 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
0 °C to 250 °C | 0.13 °C | ||
250 °C to 900 °C | 0.08 °C | ||
900 °C to 1768 °C | 0.07 °C | ||
Type S | |||
-50 °C to 0 °C | 0.18 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
0 °C to 250 °C | 0.13 °C | ||
250 °C to 900 °C | 0.08 °C | ||
900 °C to 1768 °C | 0.08 °C | ||
Type T | |||
-250 °C to -150 °C | 0.12 °C | II | Source/Measure. For the calibration of thermocouple indicators and simulators by electrical simulation and measurement using reference calibrator or multimeter. |
-150 °C to 0 °C | 0.03 °C | ||
0 °C to 120 °C | 0.02 °C | ||
120 °C to 400 °C | 0.02 °C | ||
RTD simulation | |||
−200 °C to 850 °C | 0.0034 °C to 0.026 °C | II | For the calibration of Pt100 (375), Pt100 (385), Pt100 (389), Pt100 (391), Pt100 (3926) RTD indicators and simulators by electrical simulation of temperature. |
Thermometry
Measured Quantity & Range or Instrument | Expanded Measurement Uncertainty | Type | Remarks |
---|---|---|---|
-40 °C to -25 °C | 0.032 °C | II | For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
-25 °C to 70 °C | 0.015 °C | II | For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a liquid bath. |
70 °C to 155 °C | 0.032 °C | II | For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
155 °C to 250 °C | 0.087 °C | II | For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
250 °C to 650 °C | 0.2 °C | II | For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
650 °C to 1100 °C | 2 °C | II | For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
-40 °C to 250 °C | 0.6 °C | II | For the calibration of type K and type N thermocouples with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
250 °C to 650 °C | 1.3 °C | ||
650 °C to 1200 °C | 3.1 °C | ||
-40 °C to 650 °C | 0.5 °C | II | For the calibration of type R and type S thermocouples with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
650 °C to 1200 °C | 2.0 °C | ||
-40 °C to 150 °C | 0.8 °C | II | For the calibration of type J, type E, and type T thermocouples with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
150 °C to 250 °C | 1.3 °C | ||
250 °C to 400 °C | 2.0 °C | ||
400 °C to 650 °C | 3.3 °C | ||
0 °C to 30 °C | 0.075 °C | II | For the calibration of type J, type K, and type T thermocouples with or without temperature indicators by comparison against the laboratory temperature standard in a dry block. |
Annex A
Best Uncertainty, ± {(μV/V of sourced ac voltage) + residual}, of AC voltage generation for the calibration of voltage measuring devices
Voltage, AC | Frequency | |||||||
---|---|---|---|---|---|---|---|---|
10 Hz to 20 Hz |
20 Hz to 40 Hz |
40 Hz to 20 kHz |
20 kHz to 50 kHz |
50 kHz to 100 kHz |
100 kHz to 300 kHz |
300 kHz to 500 kHz |
500 kHz to 1 MHz |
|
220 μV to 2.2 mV | 240 μV/V + 4 μV | 90 μV/V + 4 μV | 80 μV/V + 4 μV | 200 μV/V + 4 μV | 500 μV/V + 5 μV | 1050 μV/V + 10 μV | 1400 μV/V + 20 μV | 2700 μV/V + 20 μV |
2.2 mV to 22 mV | 240 μV/V + 4 μV | 90 μV/V + 4 μV | 80 μV/V + 4 μV | 200 μV/V + 4 μV | 500 μV/V + 5 μV | 1050 μV/V + 10 μV | 1400 μV/V + 20 μV | 2700 μV/V + 20 μV |
22 mV to 220 mV | 240 μV/V + 12 μV | 90 μV/V + 7 μV | 57 μV/V + 7 μV | 120 μV/V + 7 μV | 310 μV/V + 17 μV | 655 μV/V + 20 μV | 1400 μV/V + 25 μV | 2700 μV/V + 45 μV |
220 mV to 2.2 V | 240 μV/V + 40 μV | 90 μV/V + 15 μV | 42 μV/V + 8 μV | 67 μV/V + 10 μV | 85 μV/V + 30 μV | 336 μV/V + 80 μV | 1000 μV/V + 200 μV | 1700 μV/V + 300 μV |
2.2 V to 22 V | 240 μV/V + 400 μV | 90 μV/V + 150 μV | 42 μV/V + 50 μV | 67 μV/V + 100 μV | 83 μV/V + 200 μV | 254 μV/V + 600 μV | 1000 μV/V + 2000μV | 1500 μV/V + 3200 μV |
22 V to 220 V | 240 μV/V + 4 mV | 90 μV/V + 1.5 mV | 52 μV/V + 0.6 mV | 80 μV/V + 1 mV | 150 μV/V + 2.5 mV | 900 μV/V + 16 mV | 4400 μV/V + 40 mV | 8000 μV/V + 80 mV |
Best Uncertainty, ± {(μV/V of sourced ac voltage) + residual}, of AC voltage generation for the calibration of voltage measuring devices
Voltage, AC | Frequency | ||||
---|---|---|---|---|---|
15 Hz to 50 Hz | 50 Hz to 1 kHz | ||||
220 V to 1100 V | 300 μV/V + 16 mV | 70 μV/V + 3.5 mV |
Best Uncertainty, ± {(μV/V of sourced ac voltage) + residual}, of AC voltage generation for the calibration of voltage measuring devices
Voltage, AC | Frequency | ||||
---|---|---|---|---|---|
40 Hz to 1 kHz |
1 kHz to 20 kHz |
20 kHz to 30 kHz |
30 kHz to 50 kHz |
50 kHz to 100 kHz |
|
220 V to 1100 V | 90 + 4 mV | 165 + 6 mV | 600 + 11 mV | ||
220 V to 750 V | 600 + 11 mV | 2300 + 45 mV |
Annex B
Best Uncertainty, ± {(μA/A of sourced ac current) + residual}, of AC current generation for the calibration of voltage measuring devices
Current, AC | Frequency | ||||
---|---|---|---|---|---|
10 Hz to 20 Hz |
20 Hz to 40 Hz |
40 Hz to 1 kHz |
1 kHz to 5 kHz |
5 kHz to 10 kHz |
|
10 μA to 220 μA | 250 μA + 16 nA | 160 μA + 10 nA | 103 μA + 8 nA | 280 μA + 12 nA | 1100 μA + 65 nA |
220 μA to 2.2 mA | 250 μA + 40 nA | 160 μA + 35 nA | 103 μA + 35 nA | 200 μA + 110 nA | 1100 μA + 650 nA |
2.2 mA to 22 mA | 250 μA + 400 nA | 160 μA + 350 nA | 103 μA + 350 nA | 200 μA + 550 nA | 1100 μA + 5000 nA |
22 mA to 220 mA | 250 μA + 4 uA | 160 μA + 3.5 uA | 103 μA + 2.5 uA | 200 μA + 3.5 uA | 1100 μA + 10 uA |
220 mA to 2.2 A | 244 μA + 35 uA | 244 μA + 35 uA | 450 μA + 80 uA | 7000 μA + 160 uA | |
2.2 V to 11 A | 460 μA + 170 uA | 950 μA + 380 uA | 3600 μA + 750 uA |
Annex C
Best Uncertainty, ± {(% of measured AC voltage) + residual}, of AC voltage measurement for the calibration of voltage generating devices
Voltage, AC | Frequency | |||||
---|---|---|---|---|---|---|
1 Hz to 40 Hz |
40 Hz to 1 kHz |
1 kHz to 20 kHz |
20 kHz to 50 kHz |
50 kHz to 100 kHz |
100 kHz to 300 kHz |
|
1 mV to 10 mV | 0.03 % + 3 μV | 0.02 % + 1.1 μV | 0.03 % + 1.1 μV | 0.1 % + 1.1 μV | 0.5 % + 1.1 μV | 4 % + 2 μV |
10 mV to 100 mV | 0.007 % + 4 μV | 0.007 % + 2 μV | 0.014 % + 2 μV | 0.03 % + 2 μV | 0.08 % + 2 μV | 0.3 % + 10 μV |
100 mV to 1 V | 0.007 % + 40 μV | 0.007 % + 20 μV | 0.014 % + 20 μV | 0.03 % + 20 μV | 0.08 % + 20 μV | 0.3 % + 100 μV |
1 V to 10 V | 0.007 % + 400 μV | 0.007 % + 200 μV | 0.014 % + 200 μV | 0.03 % + 200 μV | 0.08 % + 200 μV | 0.3 % + 1000 μV |
10 V to 100 V | 0.02 % + 4000 μV | 0.02 % + 2000 μV | 0.02 % + 2000 μV | 0.035 % + 2000 μV | 0.12 % + 2000 μV | 0.4 % + 10000 μV |
100 V to 700 V | 0.04 % + 40000 μV | 0.04 % + 20000 μV | 0.06 % + 20000 μV | 0.12 % + 20000 μV | 0.3 % + 20000 μV |
Best Uncertainty, ± {(% of measured AC voltage) + residual}, of AC voltage measurement for the calibration of voltage generating devices
Voltage, AC | Frequency | ||||
---|---|---|---|---|---|
300 kHz to 1 MHz |
1 MHz to 2 MHz |
2 MHz to 4 MHz |
4 MHz to 8 MHz |
8 MHz to 10 MHz |
|
1 mV to 10 mV | 1.2 % + 5 μV | 7 % + 7 μV | 7 % + 7 μV | 20 % + 8 μV | |
10 mV to 100 mV | 1 % + 10 μV | 1.5 % + 10 μV | 4 % + 70 μV | 4 % + 80 μV | 15 % + 100 μV |
100 mV to 1 V | 1 % + 100 μV | 1.5 % + 100 μV | 4 % + 700 μV | 4 % + 800 μV | 15 % + 1000 μV |
1 V to 10 V | 1 % + 1000 μV | 1.5 % + 1000 μV | 4 % + 7000 μV | 4 % + 8000 μV | 15 % + 10000 μV |
10 V to 100 V | 1.5 % + 10000 μV |
Annex D
Best Uncertainty, ± {(% of measured AC current) + residual}, of AC current measurement for the calibration of current generating devices
Current, AC | Frequency | ||||||
---|---|---|---|---|---|---|---|
10 Hz to 20 Hz |
20 Hz to 45 Hz |
45 Hz to 100 Hz |
100 Hz to 5 kHz |
5 kHz to 20 kHz |
20 kHz to 50 kHz |
50 kHz to 100 kHz |
|
10 μA to 100 μA | 0.4 % + 0.03 μA | 0.15 % + 0.03 μA | 0.06 % + 0.03 μA | 0.06 % + 0.03 μA | |||
100 μA to 1 mA | 0.4 % + 0.2 μA | 0.15 % + 0.2 μA | 0.06 % + 0.2 μA | 0.03 % + 0.2 μA | 0.06 % + 0.2 μA | 0.4 % + 0.4 μA | 0.55 % + 1.5 μA |
1 mA to 10 mA | 0.4 % + 2 μA | 0.15 % + 2 μA | 0.06 % + 2 μA | 0.03 % + 2 μA | 0.06 % + 2 μA | 0.4 % + 4 μA | 0.55 % + 15 μA |
10 mA to 100 mA | 0.4 % + 20 μA | 0.15 % + 20 μA | 0.06 % + 20 μA | 0.03 % + 20 μA | 0.06 % + 20 μA | 0.4 % + 40 μA | 0.55 % + 150 μA |
100 mA to 1 A | 0.4 % + 200 μA | 0.16 % + 200 μA | 0.08 % + 200 μA | 0.1 % + 200 μA | 0.3 % + 200 μA | 1 % + 400 μA |