CLAS Certificate Number 2010-02

Company name Instrumentation Saint-Laurent Inc.
Company address 80 rue de la Montagne
St-Joseph du Lac, Québec J0N 1M0
Contact Martin Langlais
Telephone: (450) 473-6169
Fax: (450) 473-0570
EmailInfo@instrumentationsaintlaurent.com
Clients served
  • All interested parties
  • Some calibration services are available on-site. See Remarks for details.
Fields of calibration
Accreditation body
  • Standards Council of Canada (SCC)
ISO/IEC 17025
  • SCC file number: 15807
  • First issued
  • Issue 5.0e

This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body. The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS. See supplementary notes.

Electrical

Expanded Measurement

Measured Quantity & Range or Instrument Expanded Measurement
Uncertainty
Type of Service Remarks
Voltage, DC
0 mV to 110 mV ± 0.01 % rdg + 0.005 % fs II Source. For the calibration of dc voltage measuring devices.
110 mV to 1.1 V ± 0.01 % rdg + 0.005 % fs
1.1 V to 15 V ± 0.01 % rdg + 0.005 % fs
0 mV to 110 mV ± 0.025 % rdg + 0.015 % fs II Measure. For the calibration of dc voltage generating devices.
110 mV to 1,1 V ± 0.025 % rdg + 0.015 % fs
1.1 V to 11 V ± 0.025 % rdg + 0.015 % fs
11 V to 110 V ± 0.05 % rdg + 0.015 % fs
110 V to 300 V ± 0.05 % rdg + 0.015 % fs
Current, DC
0 mA to 22 mA ± 0.01 % rdg + 0.015 % fs II Source. For the calibration of dc current measuring devices.
0 mA to 30 mA ± 0.01 % rdg + 0.15 % fs II Measure. For the calibration of dc current generating devices.
30 mA to 110 mA ± 0.01 % rdg + 0.15 % fs
Resistance
0 Ω to 11 Ω ± 0.021 Ω II Source. For the calibration of resistance measuring devices.
11 Ω to 110 Ω ± 0.051 Ω
110 Ω to 1.1 kΩ ± 0.72 Ω
1.1 kΩ to 11 kΩ ± 8.3 Ω
0 Ω to 11 Ω ± 0.055 Ω II Measure. Measure capability using a calibrator.
11 Ω to 110 Ω ± 0.11 Ω
110 Ω to 1,1 kΩ ± 1.1 Ω
1,1 kΩ to 11 kΩ ± 21 Ω
Voltage, AC
0.11 V to 1.1 V
20 Hz to 40 Hz ±2 % rdg + 10 counts II Measure. For the calibration of ac voltage generating devices.
40 Hz to 500 Hz ±0.5 % rdg + 5 counts
500 Hz to 1 kHz ±2 % rdg + 10 counts
1 kHz to 5 kHz ±10 % rdg + 20 counts
1.1 V to 11 V
20 Hz to 40 Hz ±2 % rdg + 10 counts II Measure. For the calibration of ac voltage generating devices.
40 Hz to 500 Hz ±0.5 % rdg + 5 counts
500 Hz to 1 kHz ±2 % rdg + 10 counts
1 kHz to 5 kHz ±10 % rdg + 20 counts
11 V to 110 V
20 Hz to 40 Hz ±2 % rdg + 10 counts II Measure. For the calibration of ac voltage generating devices.
40 Hz to 500 Hz ±0.5 % rdg + 5 counts
500 Hz to 1 kHz ±2 % rdg + 10 counts
1 kHz to 5 kHz ±10 % rdg + 20 counts
110 V to 300 V
20 Hz to 40 Hz ±2 % rdg + 10 counts II Measure. For the calibration of ac voltage generating devices.
40 Hz to 500 Hz ±0.5 % rdg + 5 counts
500 Hz to 1 kHz ±2 % rdg + 10 counts
1 kHz to 5 kHz ±10 % rdg + 20 counts

Electrical calibration of temperature indicators

Measured Quantity & Range or Instrument Expanded Measurement
Uncertainty
Type of Service Remarks
Thermocouple simulation
Type E
-250 °C to -200 °C ± 0.6 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-200 °C to -100 °C ± 0.3 °C
-100 °C to 600 °C ± 0.3 °C
600 °C to 1000 °C ± 0.2 °C
-250 °C to -200 °C ± 1.3 °C II Measure. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-200 °C to -100 °C ± 0.5 °C
-100 °C to 600 °C ± 0.5 °C
600 °C to 1000 °C ± 0.4 °C
Type N
-200 °C to -100 °C ± 0.6 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-100 °C to 900 °C ± 0.5 °C
900 °C to 1300 °C ± 0.3 °C
-200 °C to -100 °C ± 1.0 °C II Measure. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-100 °C to 900 °C ± 0.5 °C
900 °C to 1300 °C ± 0.6 °C
Type J
-210 °C to -100 °C ± 0.3 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-100 °C to 800 °C ± 0.2 °C
800 °C to 1200 °C ± 0.2 °C
-210 °C to -100 °C ± 0.6 °C II Measure. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-100 °C to 800 °C ± 0.3 °C
800 °C to 1200 °C ± 0.5 °C
Type K
-200 °C to -100 °C ± 0.4 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-100 °C to 400 °C ± 0.3 °C
400 °C to 1200 °C ± 0.3 °C
1200 °C to 1372 °C ± 0.4 °C
-200 °C to -100 °C ± 0.7 °C II Measure. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-100 °C to 400 °C ± 0.3 °C
400 °C to 1200 °C ± 0.5 °C
1200 °C to 1372 °C ± 0.7 °C
Type T
-250 °C to -200 °C ± 0.9 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-200 °C to 0 °C ± 0.4 °C
0 °C to 400 °C ± 0.3 °C
-250 °C to -200 °C ± 1.7 °C II Measure. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
-200 °C to 0 °C ± 0.6 °C
0 °C to 400 °C ± 0.3 °C
Type R
-20 °C to 0 °C ± 1.2 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
0 °C to 100 °C ± 1.1 °C
100 °C to 1767 °C ± 0.9 °C
-20 °C to 0 °C ± 2.3 °C II Measure. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
0 °C to 100 °C ± 1.5 °C
100 °C to 1767 °C ± 1.0 °C
Type S
-20 °C to 0 °C ± 1.2 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
0 °C to 200 °C ± 1.1 °C
200 °C to 1400 °C ± 0.9 °C
1400 °C to 1767 °C ± 1.0 °C
-20 °C to 0 °C ± 2.3 °C II Measure. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
0 °C to 200 °C ± 1.5 °C
200 °C to 1400 °C ± 0.9 °C
1400 °C to 1767 °C ± 1.1 °C

Resistance temperature detectors (RTD) simulation

Measured Quantity & Range or Instrument Expanded Measurement
Uncertainty
Type of Service Remarks
PT 385, 100 Ω
-200 °C to 0 °C ± 0.1 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
0 °C to 400 °C ± 0.2 °C
400 °C to 800 °C ± 0.4 °C
-200 °C to 0 °C ± 0.3 °C II Measure. Suitable for the calibration of process calibrators by electrical simulation of temperature.
0 °C to 400 °C ± 0.5 °C
400 °C to 800 °C ± 0.8 °C
PT 385, 1000 Ω
-200 °C to 0 °C ± 0.7 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
0 °C to 200 °C ± 0.7 °C
200 °C to 400 °C ± 0.8 °C
400 °C to 630 °C ± 0.8 °C
-200 °C to 0 °C ± 0.8 °C II Measure. For the calibration of AC frequency devices. Measure capability.
0 °C to 200 °C ± 0.8 °C
200 °C to 400 °C ± 1.1 °C
400 °C to 630 °C ± 1.1 °C
PT 3926, 100 Ω
-200 °C to 0 °C ± 0.1 °C II Source. Suitable for the calibration of temperature indicators by electrical simulation of temperature.
0 °C to 630 °C ± 0.2 °C
-200 °C to 0 °C ± 0.3 °C II Measure. Suitable for the calibration of process calibrators by electrical simulation of temperature.
0 °C to 630 °C ± 0.5 °C

Time and frequency

Measured Quantity & Range or Instrument Expanded Measurement
Uncertainty
Type of Service Remarks
Frequency, AC
1 Hz to 10.99 Hz ± 0.1 Hz II Source. For the calibration of AC frequency devices. Measurement capability generated with a calibrator.
11 Hz to 109.9 Hz ± 0.1 Hz
110 Hz to 1099.9 Hz ± 0.1 Hz
1.1 kHz to 21.99 kHz ± 0.002 kHz
22 kHz to 50 kHz ± 0.005 kHz
1 Hz to 109.99 Hz ± 0.05 Hz II Measure. For the calibration of AC frequency devices. Measure capability.
110 Hz to 1099.9 Hz ± 0.5 Hz
1.1 kHz to 10.999 kHz ± 0.005 kHz
11 kHz to 50 kHz ± 0.05 kHz
60 s to 24 h 1 s II For the calibration of stopwatches. Instrumentation St-Laurent Inc. procedure 4IN9110.
On Site Calibration Service.
60 s to 1 h 1 s II For the calibration of internal clocks. Instrumentation St-Laurent Inc. procedure 4IN9110.
On Site Calibration Service.

Pressure

Measured Quantity & Range or Instrument Expanded Measurement
Uncertainty
Type of Service Remarks
Pneumatic
0 psig to 300 psig ± 1 psig II For the calibration of pneumatic devices. Instrumentation St-Laurent procedure 4IN9106. On Site Calibration Service.
0.0 inch H2O to 200.0 inch H2O ± 2.0 inch H2O
0.00 inch H2O to 5.00 inch H2O ± 0.25 inch H2O
0 psi to 5000 psi ± 13 psi
0.00 inch H2O to 1.00 inch H2O ± 0.10 inch H2O
0 psi to 10 000 psi ± 25 psi
0.0 inch Hg to 28.0 inch Hg ± 0.5 inch Hg
0 psig to 300 psig ± 1 psig II For the calibration of pneumatic devices. Instrumentation St-Laurent procedure 4IN9106. Calibration Service in laboratory.
0.0 inch H2O to 200.0 inch H2O ± 0.5 inch H2O
0.00 inch H2O to 2.00 inch H2O ± 0.01 inch H2O
0 psig to 10 000 psig ± 20 psig

Thermometry

Measured Quantity & Range or Instrument Expanded Measurement
Uncertainty
Type of Service Remarks
Thermocouple and probe RTD
-20.0 °C to 800.0 °C ± 2.0 °C II For the calibration of thermocouples and probe RTD. Instrumentation St-Laurent procedure 4IN9101 and 4IN9102. Source and measure. On Site Calibration Service.
800.0 °C to 1100.0 °C ± 4.0 °C
-15.0 °C to 650.0 °C ± 1.0 °C II For the calibration of thermocouples and probe RTD. Instrumentation St-Laurent procedure 4IN9101 and 4IN9102. Source and measure. Calibration Service in laboratory.
Direct Read, RTD
-20.0 °C to 800.0 °C ± 2.0 °C II For the calibration of direct read devices and RTD's. Instrumentation St-Laurent procedure 4IN9101 and 4IN9102. Source and measure. On Site Calibration Service.
800.0 °C to 1100.0 °C ± 4.0 °C
-15.0 °C to 650.0 °C ± 1.0 °C II For the calibration of direct read devices and RTD's. Instrumentation St-Laurent procedure 4IN9101 and 4IN9102. Source and measure. Calibration Service in laboratory.
Liquid in glass thermometers
-20.0 °C to 250.0 °C ± 2.0 °C II For the calibration of liquid in glass thermometers. Instrumentation St-Laurent procedure 4IN9102. Source and measure. On Site Calibration Service.
-15.0 °C to 250.0 °C ±1.0 °C For the calibration of liquid in glass thermometers. Instrumentation St-Laurent procedure 4IN9102. Source and measure. Calibration Service in laboratory.