Company name | Transcat Canada – Montreal |
---|---|
Company address | 9900 Côte-de-Liesse Montréal, Quebec H8T 1A1 |
Contact | Nuccio Mercuri Telephone: 514-631-6653 x7249 Fax: 514-631-6122 Email: Nuccio.Mercuri@transcat.ca |
Clients served |
|
Fields of calibration | |
SCC accreditation (ISO/IEC 17025) |
|
This CLAS Certificate is published by the CLAS program of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to the CLAS program. See supplementary notes.
Measured Quantity and Range or Instrument | Calibration and Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) |
Type of Service | Remarks |
---|---|---|---|
Mass | |||
25 kg | 30 mg | I |
One to one comparison. For the calibration of masses with traceability to the International System of Units (SI) through the primary standards of NRC and of Measurement Canada (MR-1). Calibration results are reported in terms of conventional mass as defined by the Organisation internationale de métrologie légale (OIML): "The conventional value of the value of the result of weighing a body in air is equal to mass of a standard, of conventionally chosen density (8000 kg/m3), at a conventionally chosen temperature (20 °C), which balances this body at this reference temperature in air of conventionally chosen density (1.2 kg/m3)." The Calibration Measurement Capability listed relates to measurements on integral conventional standard weights. It can be achieved only if the weights being calibrated are suitable for such a measurement. Calibrations can be given in other units as required. This service is not available on-site. |
20 kg | 20 mg | ||
10 kg | 10 mg | ||
5 kg | 2.5 mg | ||
3 kg | 2.4 mg | ||
2 kg | 2.0 mg | ||
1 kg | 0.5 mg | ||
500 g | 0.24 mg | ||
300 g | 0.15 mg | ||
200 g | 0.10 mg | ||
100 g | 0.080 mg | ||
50 g | 0.040 mg | ||
30 g | 0.050 mg | ||
20 g | 0.033 mg | ||
10 g | 0.017 mg | ||
5 g | 0.010 mg | ||
3 g | 0.008 mg | ||
2 g | 0.007 mg | ||
1 g | 0.007 mg | ||
500 mg | 0.004 mg | ||
300 mg | 0.003 mg | ||
200 mg | 0.002 mg | ||
100 mg | 0.002 mg | ||
50 mg | 0.002 mg | ||
30 mg | 0.002 mg | ||
20 mg | 0.002 mg | ||
10 mg | 0.002 mg | ||
5 mg | 0.002 mg | ||
3 mg | 0.002 mg | ||
2 mg | 0.002 mg | ||
1 mg | 0.002 mg | ||
10 kg | 40 mg | l |
One to one comparison. On-site calibration available. Procedure : PDL-09-MG-030 |
20 kg | 60 mg | ||
25 kg | 70 mg | ||
Balances (Electronic, non-automatic) - Specific values | |||
1 mg, 2 mg and 5 mg | 0.003 mg | II |
For the calibration of balances. The Calibration Measurement Capability listed can be achieved only if the balances being calibrated are suitable for such a measurement. The Calibration Measurement Capability is based upon the use of ASTM Class 1 reference weights. The laboratory corrects for known offsets assigned to its reference weights so that it can calibrate with uncertainties that are better than the ASTM tolerances of the weights used, in some cases. The Calibration Measurement Capability reflects the uncertainty contribution of the weights including co-variances. It also includes the repeatability and readability of the best balances that the laboratory calibrates routinely in the given measurement ranges. The Calibration Measurement Capabilities for measurements between 500 g and 30 kg are dominated by the performances of the best balances calibrated by the laboratory, rather than by the laboratory's reference standards and measurement processes. The uncertainty stated on the calibration report will include the uncertainty contribution of the balances that were calibrated. All values are expressed in conventional mass, as defined above. This service is available on-site. |
10 mg, 20 mg and 50 mg | 0.003 mg | ||
100 mg and 200 mg | 0.003 mg | ||
500 mg | 0.005 mg | ||
1 g and 2 g | 0.007 mg | ||
5 g | 0.010 mg | ||
10 g | 0.019 mg | ||
Balances (Electronic, non-automatic) - Other values | |||
>10 g to 20 g | 0.035 mg | II |
For the calibration of balances. The Calibration Measurement Capability listed can be achieved only if the balances being calibrated are suitable for such a measurement. The Calibration Measurement Capability is based upon the use of ASTM Class 1 reference weights. The laboratory corrects for known offsets assigned to its reference weights so that it can calibrate with uncertainties that are better than the ASTM tolerances of the weights used, in some cases. The Calibration Measurement Capability reflects the uncertainty contribution of the weights including co-variances. It also includes the repeatability and readability of the best balances that the laboratory calibrates routinely in the given measurement ranges. The Calibration Measurement Capabilities for measurements between 500 g and 30 kg are dominated by the performances of the best balances calibrated by the laboratory, rather than by the laboratory's reference standards and measurement processes. The uncertainty stated on the calibration report will include the uncertainty contribution of the balances that were calibrated. All values are expressed in conventional mass, as defined above. This service is available on-site. |
>20 g to 50 g | 0.10 mg | ||
>50 g to 100 g | 0.10 mg | ||
>100 g to 200 g | 0.15 mg | ||
>200 g to 300 g | 0.25 mg | ||
>300 g to 500 g | 0.35 mg | ||
>500 g to 1.2 kg | 2.0 mg | ||
>1.2 kg to 8 kg | 40 mg | ||
>8 kg to 30 kg | 80 mg | ||
0 kg to 60 kg | 300 mg | II |
On-site calibration available. Procedure : PDL-09-MG-010 |
60 kg to 150 kg | 3000 mg | ||
150 kg to 500 kg | 30000 mg | ||
500 kg to 1000 kg | 60000 mg | ||
Piston Pipette | |||
0.2 µL to 2 µL | 0.017 µL | II |
For the calibration of single channel piston pipettes in accordance with the laboratory's calibration procedure PDL-09-MG-050. This calibration service is available on-site. |
> 2 µL to 10 µL | 0.021 µL | ||
> 10 µL to 20 µL | 0.025 µL | ||
> 20 µL to 100 µL | 0.10 µL | ||
> 100 µL to 200 µL | 0.14 µL | ||
> 200 µL to 1000 µL | 0.76 µL | ||
> 1000 µL to 5000 µL | 4.4 µL | ||
> 5000 µL to 10000 µL | 5.9 µL | ||
> 10000 µL to 50000 µL | 7.6 µL | ||
2 µL to 20 µL | 0.31 µL | II |
For the calibration of multi-channel piston pipettes in accordance with the laboratory's calibration procedure PDL-09-MG-050. This calibration service is only available in the laboratory. |
> 20 µL to 200 µL | 0.36 µL | ||
> 200 µL to 1000 µL | 0.76 µL | ||
Thermometry | |||
Infrared radiation thermometers | |||
-15 °C to 0 °C | 0.80 °C | II | For the calibration of Infrared Temperature measuring devices at emissivity setting of 0.9 to 1 with spectral band of 8 to 14 µm using a Blackbody Source. On‑site calibrations available. |
0 °C to 50 °C | 0.65 °C | ||
50 °C to 100 °C | 0.70 °C | ||
100 °C to 120 °C | 0.76 °C | ||
120 °C to 200 °C | 0.95 °C | ||
200 °C to 350 °C | 1.6 °C | ||
350 °C to 500 °C | 2.1 °C |