Company name | Etaltech |
---|---|
Company address | 587 Adanac Street Québec, Québec, G1C 7G6 |
Contact | Christian Beaulieu, Director Telephone: 418-688-5774 ext. 2235 Fax: 418-688-2129 Email: christian.beaulieu@etaltech.com |
Clients served |
|
Fields of calibration | |
Accreditation body |
|
ISO/IEC 17025 |
|
This CLAS Certificate is published by the CLAS service of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the accreditation body. The accreditation body recognizes the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to CLAS. See supplementary notes.
Electrical
Measured Quantity & Range or Instrument | Calibration Measurement Capability expressed as an uncertainty (±) (See supplementary notes) |
Type of Service | Remarks |
---|---|---|---|
Voltage, DC | |||
1 kV to 60 kV | 0.1 % | II | Measure. High voltage divider. For the calibration of Hipot testers (dielectric strength testers) and other high voltage sources. |
Current, DC | |||
0 μA to 329.999 μA | 0.015 % + 0.02 μA | II | Source. For the calibration of dc current measuring devices. |
0.33 mA to 3.29999 mA | 0.01 % + 0.05 μA | ||
3.3 mA to 32.9999 mA | 0.01 % + 0.25 μA | ||
33 mA to 329.999 mA | 0.01 % + 2.5 μA | ||
0 A to 1.09999 A | 0.02 % + 40 μA | ||
1.1 A to 2.99999 A | 0.038 % + 40 μA | ||
0 A to 10.9999 A | 0.05 % + 500 μA | ||
11 A to 20.5 A | 0.1 % + 750 μA | ||
0 nA to 100 nA | 0.003 % + 0.04 nA | Measure. For the calibration of dc current generating devices. | |
100 nA to 1 μA | 0.002 % + 0.04 nA | ||
1 μA to 10 μA | 0.002 % + 0.1 nA | ||
10 μA to 100 μA | 0.002 % + 0.8 nA | ||
100 μA to 1 mA | 0.002 % + 5 nA | ||
1 mA to 10 mA | 0.003 % + 50 nA | ||
10 mA to 100 mA | 0.0035 % + 500 nA | ||
100 mA to 1 A | 0.011 % + 10 μA | ||
1 A to 3 A | 0.10 % + 2 mA | Measure. For the calibration of dc current generating devices. | |
3 A to 10 A | 0.15 % + 800 μA |
Measured Quantity & Range or Instrument | Frequency | Calibration Measurement Capability expressed as an uncertainty (±) (See supplementary notes) |
Type of Service | Remarks |
---|---|---|---|---|
Voltage, AC | ||||
1 kV to 40 kV rms | 50 Hz / 60 Hz | 0.5 % | II | Measure. High voltage divider. For the calibration of high voltage source devices. |
Current, AC | ||||
0.029 mA to 0.32999 mA | 10 Hz to 20 Hz | 0.2 % + 0.1 μA | II | Source. For the calibration of sinusoidal wave current measuring devices. |
20 Hz to 45 Hz | 0.15 % + 0.1 μA | |||
45 Hz to 1 kHz | 0.125 % + 0.1 μA | |||
1 kHz to 5 kHz | 0.3 % + 0.15 μA | |||
5 kHz to 10 kHz | 0.8 % + 0.2 μA | |||
10 kHz to 30 kHz | 1.6 % + 0.4 μA | |||
0.33 mA to 3.2999 mA | 10 Hz to 20 Hz | 0.2 % + 0.15 μA | ||
20 Hz to 45 Hz | 0.125 % + 0.15 μA | |||
45 Hz to 1 kHz | 0.1 % + 0.15 μA | |||
1 kHz to 5 kHz | 0.2 % + 0.2 μA | |||
5 kHz to 10 kHz | 0.5 % + 0.3 μA | |||
10 kHz to 30 kHz | 1.0 % + 0.6 μA | |||
3.3 mA to 32.999 mA | 10 Hz to 20 Hz | 0.18 % + 2 μA | ||
20 Hz to 45 Hz | 0.09 % + 2 μA | |||
45 Hz to 1 kHz | 0.04 % + 2 μA | |||
1 kHz to 5 kHz | 0.08 % + 2 μA | |||
5 kHz to 10 kHz | 0.2 % + 3 μA | |||
10 kHz to 30 kHz | 0.4 % + 4 μA | |||
33 mA to 329.999 mA | 10 Hz to 20 Hz | 0.18 % + 20 μA | ||
20 Hz to 45 Hz | 0.09 % + 20 μA | |||
45 Hz to 1 kHz | 0.04 % + 20 μA | |||
1 kHz to 5 kHz | 0.1 % + 50 μA | |||
5 kHz to 10 kHz | 0.2 % + 100 μA | |||
10 kHz to 30 kHz | 0.4 % + 200 μA | |||
0.33 A to 1.0999 A | 10 Hz to 45 Hz | 0.18 % + 100 μA | ||
45 Hz to 1 kHz | 0.05 % + 100 μA | |||
1 kHz to 5 kHz | 0.6 % + 1 mA | |||
5 kHz to 10 kHz | 2.5 % + 5 mA | |||
1.1 A to 2.99999 A | 10 Hz to 45 Hz | 0.18 % + 100 μA | ||
45 Hz to 1 kHz | 0.06 % + 100 μA | |||
1 kHz to 5 kHz | 0.6 % + 1 mA | |||
5 kHz to 10 kHz | 2.5 % + 5 mA | |||
3 A to 10.9999 A | 45 Hz to 100 Hz | 0.06 % + 2 mA | ||
100 Hz to 1 kHz | 0.1 % + 2 mA | |||
1 kHz to 5 kHz | 3.0 % + 2 mA | |||
11 A to 20.5 A | 45 Hz to 100 Hz | 0.12 % + 5 mA | ||
100 Hz to 1 kHz | 0.15 % + 5 mA | |||
1 kHz to 5 kHz | 3.0 % + 5 mA | |||
0 μA to 100 μA | 10 Hz to 20 Hz | 0.4 % + 0.03 μA | II | Measure. For the calibration of current sources with a digital multimeter. |
20 Hz to 45 Hz | 0.15 % + 0.03 μA | |||
45 Hz to 100 Hz | 0.06 % + 0.03 μA | |||
100 Hz to 5 kHz | 0.06 % + 0.03 μA | |||
100 μA to 1 mA | 10 Hz to 20 Hz | 0.4 % + 0.2 μA | ||
20 Hz to 45 Hz | 0.15 % + 0.2 μA | |||
45 Hz to 100 Hz | 0.06 % + 0.2 μA | |||
100 Hz to 5 kHz | 0.03 % + 0.2 μA | |||
5 kHz to 20 kHz | 0.06 % + 0.2 μA | |||
20 kHz to 50 kHz | 0.4 % + 0.4 μA | |||
50 kHz to 100 kHz | 0.55 % + 1.5 μA | |||
1 mA to 10 mA | 10 Hz to 20 Hz | 0.4 % + 2 μA | ||
20 Hz to 45 Hz | 0.15 % + 2 μA | |||
45 Hz to 100 Hz | 0.06 % + 2 μA | |||
100 Hz to 5 kHz | 0.03 % + 2 μA | |||
5 kHz to 20 kHz | 0.06 % + 2 μA | |||
20 kHz to 50 kHz | 0.4 % + 4 μA | |||
50 kHz to 100 kHz | 0.55 % + 15 μA | |||
10 mA to 100 mA | 10 Hz to 20 Hz | 0.4 % + 20 μA | ||
20 Hz to 45 Hz | 0.15 % + 20 μA | |||
45 Hz to 100 Hz | 0.06 % + 20 μA | |||
100 Hz to 5 kHz | 0.03 % + 20 μA | |||
5 kHz to 20 kHz | 0.06 % + 20 μA | |||
20 kHz to 50 kHz | 0.4 % + 40 μA | |||
50 kHz to 100 kHz | 0.55 % + 150 μA | |||
100 mA to 1 A | 10 Hz to 20 Hz | 0.4 % + 0.2 mA | ||
20 Hz to 45 Hz | 0.16 % + 0.2 mA | |||
45 Hz to 100 Hz | 0.08 % + 0.2 mA | |||
100 Hz to 5 kHz | 0.1 % + 0.2 mA | |||
5 kHz to 20 kHz | 0.3 % + 0.2 mA | |||
20 kHz to 50 kHz | 1 % + 0.4 mA | |||
1 A to 3 A | 3 Hz to 5 Hz | 1.1 % + 6 mA | II | Measure. For the calibration of current sources with a digital multimeter. |
5 Hz to 10 Hz | 0.35 % + 6 mA | |||
10 Hz to 5 kHz | 0.15 % + 6 mA | |||
5 kHz to 10 kHz | 0.35 % + 70 mA | |||
3 A to 10 A | 3 Hz to 5 Hz | 2 % + 6 mA | ||
5 Hz to 10 Hz | 1.1 % + 6 mA | |||
10 Hz to 5 kHz | 0.15 % + 6 mA | |||
5 kHz to 10 kHz | 0.35 % + 70 mA |
Force
Measured Quantity & Range or Instrument | Calibration Measurement Capability expressed as an uncertainty (±) (See supplementary notes) |
Type of Service | Remarks |
---|---|---|---|
Torque | |||
Clockwise | |||
5 in•lb to 50 in•lb | 3 % of reading | II |
For the calibration of torque wrenches and screwdrivers. Etaltech procedure: Torque General 5-50inlbs Auto_2000-400-02-1, 2000-610-2. Comparison to a torque standard (Standard CDI, monitor 2000-610-2 + transducer 2000-400-02-1). |
40 in•lb to 400 in•lb | 3 % of reading | II |
For the calibration of torque wrenches and screwdrivers. Etaltech procedure: Torque General 40-400inlb Auto_2000-400-02-2, 2000-610-2. Comparison to a torque standard (Standard CDI, monitor 2000-610-2 + transducer 2000-400-02-2). |
100 in•lb to 1000 in•lb | 3 % of reading | II |
For the calibration of torque wrenches and screwdrivers. Etaltech procedure: Torque General 100-1000inlb Auto_2000-400-02-3. Comparison to a torque standard (Standard CDI, monitor 2000-610-2 + transducer 2000-400-02-3). |
25 ft•lb to 250 ft•lb | 3 % of reading | II |
For the calibration of torque wrenches and screwdrivers. Etaltech procedure: Torque General 25-250ftlb Auto_2000-400-02-4, 2000-610-2. Comparison to a torque standard (Standard CDI, monitor 2000-610-2 + transducer 2000-400-02-4). |
200 ft•lb to 2000 ft•lb | 3 % of reading | II |
For the calibration of torque wrenches and screwdrivers. Etaltech procedure: Torque General 200-2000 Auto_2000-14-02, 2000-610-2. Comparison to a torque standard (Standard CDI, monitor 2000-610-2 + transducer 2000-14-02). |