Company name | Labmed Solutions |
---|---|
Company address | 9452 du Golf Blvd. Anjou, Quebec H1J 3A1 |
Contact | Guido Di Ridolfo Telephone: 514-395-1090 Toll Free: 1-877-584-7474 Fax: 514-684-5352 Email: guido@labmedsolutions.com |
Clients served |
|
Fields of calibration | |
SCC accreditation (ISO/IEC 17025) |
|
This CLAS Certificate is published by the CLAS program of the National Research Council of Canada (NRC) in cooperation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform calibrations as per the listed Calibration and Measurement Capabilities with metrological traceability to the International System of Units (SI) or to standards acceptable to the CLAS program. See supplementary notes.
Volume
Measured Quantity & Range or Instrument | Calibration and Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) |
Type of Service | Remarks |
---|---|---|---|
Piston pipette | |||
0.2 µL to 10 µL | 0.020 µL | II | For the calibration of single channel and multi-channel piston pipettes. The pipettes are calibrated gravimetrically in accordance with the laboratory's calibration procedure PROT 03 and the ISO 8655-6:2002 standard. |
> 10 µL to 20 µL | 0.12 µL | ||
> 20 µL to 100 µL | 0.15 µL | ||
> 100 µL to 200 µL | 0.76 µL | ||
> 200 µL to 1000 µL | 1.7 µL | ||
> 1000 µL to 5000 µL | 15 µL | ||
> 5000 µL to 10000 µL | 20 µL |
Time and frequency
Measured Quantity & Range or Instrument | Calibration and Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) |
Type of Service | Remarks |
---|---|---|---|
Centrifuge (refrigerated or not) | |||
Rotational speed: 100 to 15000 rpm | 10 rpm | II |
The centrifuges are calibrated in accordance with the laboratory's calibration procedure PROT 01. The calibration of rotational speed is only possible on centrifuges equipped with digital display. The calibration is done using an optical tachometer. This calibration service is only available on-site. |
2 to 20000 Hz | 3 Hz |
The centrifuges are calibrated in accordance with the laboratory's calibration procedure PROT 13. The calibration of rotational speed is only possible on centrifuges equipped with digital display. The calibration is done using multimeter. This calibration service is only available on-site. |
|
Timer | |||
5 minutes | 9 sec | II | The timer of the centrifuges is calibrated in accordance with the laboratory's calibration procedure PROT 01 and PROT 13. |
Thermometry
Measured Quantity & Range or Instrument | Calibration and Measurement Capability expressed as an Uncertainty (±) (See supplementary notes) |
Type of Service | Remarks |
---|---|---|---|
Temperature | |||
4 ºC | 1.4 ºC | II | The temperature within the centrifuges is calibrated in accordance with the laboratory's calibration procedure PROT 01. |
20 ºC | 1 ºC | ||
4 ºC | 1.3 ºC | The temperature within the centrifuges is calibrated in accordance with the laboratory's calibration procedure PROT 13. | |
20 ºC | 0.9 ºC |